Measurement of output factors for small photon beams

被引:147
|
作者
Sauer, Otto A. [1 ]
Wilbert, Juergen [1 ]
机构
[1] Univ Wurzburg, Klin Strahlentherapie, D-97080 Wurzburg, Germany
关键词
output factors; small photon beams; dosimetry; radiation detectors;
D O I
10.1118/1.2734383
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A variety of detectors and procedures for the measurement of small field output factors are discussed in the current literature. Different detectors with or without corrections are recommended. Correction factors are often derived by Monte Carlo methods, where the bias due to approximations in the model is difficult to judge. Over that, results appear to be contradictory in some cases. In this work, output factors were measured for field sizes from 4 mm up to 180 mm side length with different detectors. A simple linear correction for the energy response of solid state detectors is proposed. This led to identical values down to 8 mm field size, as long as the size of the detector is small against the field size. The correction was of the order of a few percent. For a shielded silicon diode it was well below 1%. A physically meaningful function is proposed in order to calculate output factors for arbitrary field sizes with high accuracy. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:1983 / 1988
页数:6
相关论文
共 50 条
  • [11] Stereotactic Cone Output Factors of Flattening-Filter-Free Mode Photon Beams; a Plastic Scintillation Detector Measurement Study
    Kim, J.
    Kahn, D.
    Qian, X.
    Lu, S.
    Hsia, A. T.
    Xu, Z.
    [J]. MEDICAL PHYSICS, 2020, 47 (06) : E853 - E853
  • [12] Determination of small field output factors in 6 and 10 MV flattening filter free photon beams using various detectors
    Masanga, W.
    Tangboonduangjit, P.
    Khamfongkhruea, C.
    Tannanonta, C.
    [J]. 13TH SOUTH-EAST ASIAN CONGRESS OF MEDICAL PHYSICS 2015 (SEACOMP), 2016, 694
  • [13] Output correction factors for small static fields in megavoltage photon beams for seven ionization chambers in two orientations - perpendicular and parallel
    Casar, Bozidar
    Gershkevitsh, Eduard
    Mendez, Ignasi
    Jurkovic, Slaven
    Huq, M. Saiful
    [J]. MEDICAL PHYSICS, 2020, 47 (01) : 242 - 259
  • [14] An Experimental and Monte Carlo Study of Output Factors for Small Radiosurgery Beams
    Yang, Y.
    Kim, J.
    Li, F.
    Li, X.
    Li, T.
    Yue, N.
    Heron, D.
    Huq, M.
    [J]. MEDICAL PHYSICS, 2008, 35 (06) : 2785 - +
  • [15] Analytical Formalism for the Output Factors of Small MLC-Shaped Beams
    Abu Arida, L.
    Wadi-Ramahi, S.
    Dababneh, S.
    [J]. MEDICAL PHYSICS, 2010, 37 (06) : 3263 - +
  • [16] Measurement of the in-air output ratio for high-energy photon beams used in radiotherapy
    Appasamy, Murugan
    Xavier, Sidonia Valas
    Kuppusamy, Thayalan
    Velayudham, Ramasubramanian
    [J]. TURKISH JOURNAL OF MEDICAL SCIENCES, 2013, 43 (03) : 441 - 447
  • [17] Monte Carlo calculated output correction factors for nine small field detectors in Varian Clinac iX 6 MV photon beams
    Benmakhlouf, H.
    Sempau, J.
    Andreo, P.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S201 - S202
  • [18] Output correction factors for nine small field detectors in 6 MV radiation therapy photon beams: A PENELOPE Monte Carlo study
    Benmakhlouf, Hamza
    Sempau, Josep
    Andreo, Pedro
    [J]. MEDICAL PHYSICS, 2014, 41 (04)
  • [19] Measurement of ionization chamber absorbed dose kQ factors in megavoltage photon beams
    McEwen, Malcolm R.
    [J]. MEDICAL PHYSICS, 2010, 37 (05) : 2179 - 2193