Dosimetric verification of IMRT and 3D conformal treatment delivery using EPID

被引:0
|
作者
Nasseri, S. [1 ]
Bahreyni, M. H. [1 ]
Momennezhad, M. [1 ,2 ]
Gholamhosseinian, H. [1 ]
Shahedi, F. [1 ]
Hashemi, S. M. [1 ]
Mohammadi, M. [3 ,4 ]
机构
[1] Mashhad Univ Med Sci, Med Phys Res Ctr, Mashhad, Razavi Khorasan, Iran
[2] Mashhad Univ Med Sci, Nucl Med Res Ctr, Mashhad, Razavi Khorasan, Iran
[3] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA 5000, Australia
[4] Univ Adelaide, Sch Phys Sci, Adelaide, SA 5005, Australia
关键词
Electronic portal imaging devices (EPIDS); Back-projection; Water equivalent path length; Digital reconstructed radiograph (DRR); ALGORITHM; EXIT;
D O I
10.1016/j.apradiso.2022.110116
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: Electronic portal imaging devices (EPIDs) could potentially be useful for either in-vivo or pre-treatment dosimetric verification of external beam radiation therapy. The accuracy of EPID for dosimetric purposes is highly dependent on the specific method used for the determination of dose-response characteristics. The aim of this study was to develop a simple and time-saving EPID back-projection dosimetry algorithm for 2D dose verification in 3D conformal and intensity-modulated beams.Methods: The procedure of dose reconstruction includes a first calibration step using ionization chamber measurements to convert the Electronic Portal Image (EPI) pixel values into an absorbed dose in water. Subsequently, several corrections were applied to the Portal Dose Images (PDIs) for the effect of field size, attenuator thickness, scattering radiation, beam hardening and EPID off-axis response. Furthermore, to consider tissue inhomogeneity for accurate dose reconstruction, the patient's water equivalent path length (WEPL) was calculated using a range of digitally reconstructed radiographs (DRRs) obtained at various thicknesses by Plastimatch software. The EPIDderived dose maps accuracy was assessed by comparing with the treatment planning system (TPS) calculated dose in the prostate region of Alderson phantom irradiated with 3D conformal and intensity-modulated beams.Results: The gamma analysis for the dose plane showed agreements of 96.95% and 93.5% for 3D conformal and IMRT fields, respectively, with 3%/3 mm acceptance criteria.Conclusion: The presented algorithm can provide accurate absolute 2D dose maps for clinical use in the context of 3DCRT or IMRT Quality Assurance (QA) programs.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Automatic detection of IMRT delivery errors using a quantitative 2D dosimetric verification system
    Childress, N
    Rosen, I
    MEDICAL PHYSICS, 2004, 31 (06) : 1843 - 1843
  • [42] Dosimetric and radiobiologic comparison of 3D conformal, IMRT, VMAT and proton therapy for the treatment of early-stage glottic cancer
    Matthiesen, Chance
    Herman, Tania De La Fuente
    Singh, Hardev
    Mascia, Anthony
    Confer, Michael
    Simpson, Hilarie
    Higby, Christine
    Arain, Abeer
    Keole, Sameer
    Herman, Terence
    Bogardus, Carl
    Zhao, Yan D.
    Ahmad, Salahuddin
    JOURNAL OF MEDICAL IMAGING AND RADIATION ONCOLOGY, 2015, 59 (02) : 221 - 228
  • [43] Pronation vs. supination: a dosimetric comparison of 3D conformal RT and IMRT of rectal cancer
    Wagner, K.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2007, 183 : 76 - 76
  • [44] 3D dose verification for stereotactic lung and spine treatments using the EPID
    McCowan, Peter
    vanBeek, Timothy
    van Uytven, Eric
    McCurdy, Boyd
    MEDICAL PHYSICS, 2017, 44 (08) : 4390 - 4390
  • [45] 3D in vivo dose verification of entire hypo-fractionated IMRT treatments using an EPID and cone-beam CT
    McDermott, Leah N.
    Wendling, Markus
    Nijkamp, Jasper
    Mans, Anton
    Sonke, Jan-Jakob
    Mijnheer, Ben J.
    van Herk, Marcel
    RADIOTHERAPY AND ONCOLOGY, 2008, 86 (01) : 35 - 42
  • [46] A Radiobiological Perspective between IMRT and 3D Conformal Treatment of Prostate Cancer
    Stathakis, S.
    Buckey, C.
    Mavroidis, P.
    Swanson, G.
    Papanikolaou, N.
    Smith, M.
    Giantsoudi, D.
    Dzintars, E.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S623 - S623
  • [47] A comparison of 3D conformal radiotherapy and IMRT treatment plans in prostate cancer
    Helal, A.
    Sundar, S.
    Przeslak, A.
    Perkins, A.
    EJC SUPPLEMENTS, 2005, 3 (02): : 244 - 244
  • [48] EPID dosimetry - configuration and pre-treatment IMRT verification
    Grzadziel, Aleksandra
    Smolinska, Barbara
    Rutkowski, Roman
    Slosarek, Krzysztof
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2007, 12 (06) : 307 - 312
  • [49] Nasopharyngeal carcinoma.: Treatment planning with IMRT and 3D conformal radiotherapy
    Kristensen, Claus A.
    Kjaer-Kristoffersen, Flemming
    Sapru, Wendy
    Berthelsen, Anne K.
    Loft, Annika
    Specht, Lena
    ACTA ONCOLOGICA, 2007, 46 (02) : 214 - 220
  • [50] EPID Dosimetry - Configuration and Pre-Treatment IMRT Verification
    Grzadziel, A.
    Smolinska, B.
    Rutkowski, R.
    Slosarek, K.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1932 - 1936