Dosimetric verification of IMRT and 3D conformal treatment delivery using EPID

被引:0
|
作者
Nasseri, S. [1 ]
Bahreyni, M. H. [1 ]
Momennezhad, M. [1 ,2 ]
Gholamhosseinian, H. [1 ]
Shahedi, F. [1 ]
Hashemi, S. M. [1 ]
Mohammadi, M. [3 ,4 ]
机构
[1] Mashhad Univ Med Sci, Med Phys Res Ctr, Mashhad, Razavi Khorasan, Iran
[2] Mashhad Univ Med Sci, Nucl Med Res Ctr, Mashhad, Razavi Khorasan, Iran
[3] Royal Adelaide Hosp, Dept Med Phys, Adelaide, SA 5000, Australia
[4] Univ Adelaide, Sch Phys Sci, Adelaide, SA 5005, Australia
关键词
Electronic portal imaging devices (EPIDS); Back-projection; Water equivalent path length; Digital reconstructed radiograph (DRR); ALGORITHM; EXIT;
D O I
10.1016/j.apradiso.2022.110116
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
Purpose: Electronic portal imaging devices (EPIDs) could potentially be useful for either in-vivo or pre-treatment dosimetric verification of external beam radiation therapy. The accuracy of EPID for dosimetric purposes is highly dependent on the specific method used for the determination of dose-response characteristics. The aim of this study was to develop a simple and time-saving EPID back-projection dosimetry algorithm for 2D dose verification in 3D conformal and intensity-modulated beams.Methods: The procedure of dose reconstruction includes a first calibration step using ionization chamber measurements to convert the Electronic Portal Image (EPI) pixel values into an absorbed dose in water. Subsequently, several corrections were applied to the Portal Dose Images (PDIs) for the effect of field size, attenuator thickness, scattering radiation, beam hardening and EPID off-axis response. Furthermore, to consider tissue inhomogeneity for accurate dose reconstruction, the patient's water equivalent path length (WEPL) was calculated using a range of digitally reconstructed radiographs (DRRs) obtained at various thicknesses by Plastimatch software. The EPIDderived dose maps accuracy was assessed by comparing with the treatment planning system (TPS) calculated dose in the prostate region of Alderson phantom irradiated with 3D conformal and intensity-modulated beams.Results: The gamma analysis for the dose plane showed agreements of 96.95% and 93.5% for 3D conformal and IMRT fields, respectively, with 3%/3 mm acceptance criteria.Conclusion: The presented algorithm can provide accurate absolute 2D dose maps for clinical use in the context of 3DCRT or IMRT Quality Assurance (QA) programs.
引用
收藏
页数:7
相关论文
共 50 条
  • [1] Dosimetric pre-treatment verification of IMRT using an EPID; clinical experience
    van Zijtveld, Mathilda
    Dirkx, Maarten L. P.
    de Boer, Hans C. J.
    Heijmen, Ben J. M.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 (02) : 168 - 175
  • [2] 3D in vivo dose verification of an entire IMRT treatment using an EPID and Cone-beam CT
    McDermott, L.
    Wending, M.
    Mans, A.
    de Jong, R.
    Sonke, J.
    van Herk, M.
    Mijnheer, B.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S98 - S98
  • [3] Dosimetric IMRT verification with a flat-panel EPID
    Warkentin, B
    Steciw, S
    Rathee, S
    Fallone, BG
    MEDICAL PHYSICS, 2003, 30 (12) : 3143 - 3155
  • [4] The dosimetric impact of respiration motion on 3D, IMRT, and tomotherapy treatment delivery
    Byrne, T
    Ramsey, C
    Mahan, S
    Chase, D
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S496 - S496
  • [5] In Vivo 3D Dose Verification for IMRT Using Electronic Portal Imaging Device (EPID)
    Chang, C.
    Anderson, N. G.
    Zhu, T. C.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [6] Dose Verification for IMRT DMLC Delivery Using An EPID Imager
    Zhao, B.
    Yang, Y.
    Li, T.
    Li, X.
    Heron, D.
    Huq, M.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [7] 3D dose reconstruction for clinical evaluation of IMRT pretreatment verification with an EPID
    van Zijtveld, Mathilda
    Dirkx, Maarten L. P.
    de Boer, Hans C. J.
    Heijmen, Ben J. M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 82 (02) : 201 - 207
  • [8] A dosimetric comparison of 3 types of breast treatment plans. 3D conformal, RapidArc and IMRT
    Pella, S.
    Dumitru, N.
    Pudasani, M.
    RADIOTHERAPY AND ONCOLOGY, 2018, 127 : S995 - S995
  • [9] Dosimetric verification needs for conformal therapy and IMRT
    Fraass, B
    Moran, J
    MEDICAL PHYSICS, 2002, 29 (06) : 1358 - 1358
  • [10] Whole IMRT, Hybrid IMRT and 3D Conformal Plan A Dosimetric Comparison for Large Target
    Sharma, S.
    Manigandan, D.
    Gandhi, A.
    Subramani, V.
    Sharma, D.
    Kumar, P.
    Julka, P.
    Rath, G.
    MEDICAL PHYSICS, 2012, 39 (06) : 3840 - 3841