Revision Total Hip Arthroplasty Using Imageless Navigation With the Concept of Combined Anteversion

被引:11
|
作者
Chang, Jun-Dong [1 ]
Kim, In-Sung [1 ]
Prabhakar, Sharad [1 ]
Mansukhani, Sameer A. [1 ]
Lee, Sang-Soo [2 ]
Yoo, Je-Hyun [3 ]
机构
[1] Hallym Univ, Dept Orthopaed Surg, Dongtan Sacred Heart Hosp, 7 Keunjaebong Gil, Hwaseong Si 445907, Gyeonggi Do, South Korea
[2] Hallym Univ, Chuncheon Sacred Heart Hosp, Dept Orthopaed Surg, Coll Med, Chunchon, South Korea
[3] Hallym Univ, Coll Med, Sacred Heart Hosp, Dept Orthopaed Surg, Anyang, South Korea
来源
JOURNAL OF ARTHROPLASTY | 2017年 / 32卷 / 05期
关键词
imageless navigation; combined anteversion; revision total hip arthroplasty; component position; computer-assisted orthopaedic surgery; ACETABULAR COMPONENT; DISLOCATION; PLACEMENT;
D O I
10.1016/j.arth.2016.12.030
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: The purpose of the study was to evaluate the implant positions and clinical results of revision total hip arthroplasty (THA) using an imageless navigation with the concept of combined anteversion. Methods: A total of 40 cementless revision THAs (24 men and 16 women) using an imageless navigation with the concept of combined anteversion were retrospectively evaluated. The concept of combined anteversion was applied in cup positioning based on Widmer's equation (cup anteversion + 0.7 x stem anteversion). The mean follow-up period was 80.7 months. Postoperatively, the inclination of the cup was evaluated on standard anteroposterior view of the radiograph, and the anteversion of the cup and femoral stem was evaluated using computed tomography scan. A cup inclination of 40 degrees +/- 10 degrees and combined anteversion of the cup and femoral stem of 37 degrees +/- 10 degrees based on Widmer's equation were regarded as the "safe zone." Results: The average anteversion of the revised femoral stems was 15.3 degrees +/- 2.9 degrees (range, 9.5 degrees-21.5 degrees), whereas that of the remained femoral stems was 17.4 degrees +/- 9.7 degrees (range, 4.2 degrees-29.8 degrees). The inclination, anteversion of the cup, and combined anteversion after revision THA were 42.3 degrees +/- 3.1 degrees (range, 32.1 degrees- 48.2 degrees), 25.0 degrees +/- 2.9 degrees (range, 16.9 degrees-29.5 degrees), and 36.1 degrees +/- 3.4 degrees (range, 27.2 degrees-42.9 degrees), respectively. Therefore, the position of the implants, relative to the safe zone, showed no outliers after the revision surgery. Neither dislocation nor osteolysis was observed after the surgery. Conclusion: Favorable results of this study indicate that imageless navigation helps the surgeon in placing the components of revision THA in the safe zone. This study also shows that when this safe zone is consistently obtained, then no postoperative dislocations were observed in these patients over the 6-year follow-up period. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1576 / 1580
页数:5
相关论文
共 50 条
  • [41] Revision total hip arthroplasty
    Lakstein, D.
    MINERVA ORTOPEDICA E TRAUMATOLOGICA, 2010, 61 (01) : 51 - 65
  • [42] Revision total hip arthroplasty
    Sporer, SM
    Paprosky, WG
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2003, (417) : 203 - 209
  • [43] Combined acetabular and femoral navigation for resurfacing total hip arthroplasty
    Sugano, N
    Nishii, T
    Nakahodo, K
    Sasama, T
    Sato, Y
    Tamura, S
    Sakai, KT
    Haraguchi, K
    Nishihara, S
    Ohzono
    Yonenobu, K
    Ochi, T
    CARS 2000: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2000, 1214 : 226 - 230
  • [44] Combined acetabular and femoral surgical navigation in total hip arthroplasty
    Sugano, N
    Sato, Y
    Sasama, T
    Nakahodo, K
    Yoden, S
    Nishii, T
    Sakai, T
    Haraguchi, K
    Tamura, S
    Ohzono, K
    Ochi, T
    CARS '99: COMPUTER ASSISTED RADIOLOGY AND SURGERY, 1999, 1191 : 722 - 725
  • [45] Accuracy of Imageless Computer Assisted Navigation System through in Total Hip Arthroplasty in vitro and in vivo Studies
    Lim, D.
    Lin, F.
    Wixson, R.
    Hendrix, R.
    MacDonald, M.
    Makhsous, M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 3044 - +
  • [46] Measuring leg Length and offset with an imageless navigation system during total hip arthroplasty: is it really accurate?
    Renkawitz, Tobias
    Schuster, Tibor
    Herold, Thomas
    Goessmann, Holger
    Sendtner, Ernst
    Grifka, Joachim
    Kalteis, Thomas
    INTERNATIONAL JOURNAL OF MEDICAL ROBOTICS AND COMPUTER ASSISTED SURGERY, 2009, 5 (02): : 192 - 197
  • [47] Conversion total hip arthroplasty: Primary or revision total hip arthroplasty
    Schwarzkopf, Ran
    Baghoolizadeh, Mahta
    WORLD JOURNAL OF ORTHOPEDICS, 2015, 6 (10): : 750 - 753
  • [48] The learning curve associated with imageless navigation in total knee arthroplasty
    Michelle A. Zabat
    Christian T. Oakley
    Scott E. Marwin
    Morteza Meftah
    Ran Schwarzkopf
    Archives of Orthopaedic and Trauma Surgery, 2023, 143 : 1013 - 1019
  • [49] The learning curve associated with imageless navigation in total knee arthroplasty
    Zabat, Michelle A.
    Oakley, Christian T.
    Marwin, Scott E.
    Meftah, Morteza
    Schwarzkopf, Ran
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2023, 143 (02) : 1013 - 1019
  • [50] Imageless Computer Navigation in Total Knee Arthroplasty-The Pitfalls
    Kwok-Hing, Chiu
    Kin-Wing, Cheung
    Kwong-Yin, Chung
    JOURNAL OF ORTHOPAEDICS TRAUMA AND REHABILITATION, 2011, 15 (02) : 40 - 42