Adverse reaction to metal debris in a consecutive series of DUROM™ hip resurfacing: pseudotumour incidence and metal ion concentration

被引:4
|
作者
Hartmann, Albrecht [1 ]
Kieback, Jan-Dirk [2 ]
Luetzner, Joerg [1 ]
Guenther, Klaus-Peter [1 ]
Goronzy, Jens [1 ]
机构
[1] Tech Univ Dresden, Univ Med Carl Gustav Carus, Univ Ctr Orthopaed & Traumatol, Dresden, Germany
[2] Univ Bonn, Dept Orthopaed & Trauma Surg, Bonn, Germany
关键词
ARMD; Hip; Metal debris; Metal ion; Pseudotumour; Resurfacing; REDUCTION SEQUENCE-MRI; ON-METAL; SURFACE REPLACEMENT; ARTICULAR SURFACE; EARLY FAILURE; BLOOD COBALT; RISK-FACTORS; ARTHROPLASTY; PREVALENCE; CHROMIUM;
D O I
10.5301/hipint.5000468
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction: The aim of this study was to evaluate the incidence of adverse reactions to metal debris (ARMD) in a consecutive series of DUROM (TM) Hip Resurfacing Arthroplasty (HRA) at mid-term follow-up. Methods: Between October 2003 and March 2007 a total of 134 consecutive DUROM (TM) HRA in 121 patients were performed at our institution. Follow-up could be obtained in 101 unrevised patients (83%) at a mean time of 8.51 +/- 0.97 years postoperatively and included patient-related outcome measurement, plain radiographs, MARS-MRI as well as whole blood metal ion assessment. Results: 17 (16.5%) out of 103 hips revealed pseudotumour occurrence in MRI investigation, 1 (10.6%) with a diameter of >= 2 cm. Higher incidence of pseudotumours was found patients with femoral component size < 50 mm. 38 patients had elevated cobalt levels, 36 of them (35%) in the range of 2-7 mu g/l and 2 patients (1.9%) at a rate of > 7 mu g/l. In contrast to cobalt determination, only elevated chromium values showed a positive association with pseudotumour occurrence and size. Discussion: A significant proportion of patients developed pseudotumours and metal ion elevation in a consecutive cohort of DUROM T HRA after mid-term follow-up. The incidence, however, seems not to differ from results of other well performing resurfacing brands; clinical relevance of our findings is unclear. Regarding potential local as well as systemic effects of metal particle release, close follow-up of patients is essential, even with clinically well-performing implants.
引用
收藏
页码:343 / 348
页数:6
相关论文
共 50 条
  • [1] The development of the Durom™ metal-on-metal hip resurfacing
    Grigoris, P.
    Roberts, P.
    Panousis, K.
    [J]. HIP INTERNATIONAL, 2006, 16 (02) : S65 - S72
  • [2] Adverse reaction to metal debris after Birmingham hip resurfacing arthroplasty
    Junnila, Mika
    Seppanen, Matti
    Mokka, Jari
    Virolainen, Petri
    Polonen, Tuukka
    Vahlberg, Tero
    Mattila, Kimmo
    Tuominen, Esa K. J.
    Rantakokko, Juho
    Aarimaa, Ville
    Itala, Ari
    Makela, Keijo T.
    [J]. ACTA ORTHOPAEDICA, 2015, 86 (03) : 345 - 350
  • [3] A case series of 35 hip revisions for adverse reactions to metal debris following Cormet hip resurfacing
    Norris, Rory J.
    McArthur, John
    Parsons, Helen
    Smith, Nicholas A.
    Sprowson, Andrew P.
    Foguet, Pedro
    [J]. HIP INTERNATIONAL, 2014, 24 (06) : 575 - 581
  • [4] Early Failure of the Durom Prosthesis in Metal-on-Metal Hip Resurfacing in Chinese Patients
    Li, Jia
    He, Chongru
    Li, Dahe
    Zheng, Wei
    Liu, Denghui
    Xu, Weidong
    [J]. JOURNAL OF ARTHROPLASTY, 2013, 28 (10): : 1816 - 1821
  • [5] Adverse reaction to metal debris in a painful hemiarthroplasty of the hip
    Palmer, Jonathan Simon
    Dean, Michael William
    Briant-Evans, Toby
    Stranks, Geoff
    [J]. JOURNAL OF SURGICAL CASE REPORTS, 2016, (09):
  • [6] Adverse reaction to metal debris: metallosis of the resurfaced hip
    Pritchett, James W.
    [J]. CURRENT ORTHOPAEDIC PRACTICE, 2012, 23 (01): : 50 - 58
  • [7] One-Component Revision of Failed Hip Resurfacing from Adverse Reaction to Metal Wear Debris
    Pritchett, James W.
    [J]. JOURNAL OF ARTHROPLASTY, 2014, 29 (01): : 219 - 224
  • [8] Adverse reaction to metal debris following hip resurfacing THE INFLUENCE OF COMPONENT TYPE, ORIENTATION AND VOLUMETRIC WEAR
    Langton, D. J.
    Joyce, T. J.
    Jameson, S. S.
    Lord, J.
    Van Orsouw, M.
    Holland, J. P.
    Nargol, A. V. F.
    De Smet, K. A.
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2011, 93B (02): : 164 - 171
  • [9] Differences in concentration of metal debris in blood, serum, and plasma samples of patients with metal-on-metal hip resurfacing arthroplasty
    Khan, M.
    Kuiper, J. H.
    Sieniawska, Christine
    Richardson, J. B.
    [J]. JOURNAL OF ORTHOPAEDICS, 2016, 13 (04) : 450 - 454
  • [10] Large-Diameter Modular Metal-on-Metal Total Hip Arthroplasty Incidence of Revision for Adverse Reaction to Metallic Debris
    Barrett, William P.
    Kindsfater, Kirk A.
    Lesko, James P.
    [J]. JOURNAL OF ARTHROPLASTY, 2012, 27 (06): : 976 - 983