Design and Characterization of a Dedicated Cone-Beam CT Scanner for Detection of Acute Intracranial Hemorrhage

被引:1
|
作者
Xu, J. [1 ]
Sisniega, A. [1 ]
Zbijewski, W. [1 ]
Dang, H. [1 ]
Stayman, J. W. [1 ]
Wang, X. [2 ]
Foos, D. H. [2 ]
Aygun, N. [3 ]
Koliatsos, V. E. [4 ]
Siewerdsen, J. H. [1 ,3 ]
机构
[1] Johns Hopkins Univ, Dept Biomed Engn, Baltimore, MD 21218 USA
[2] Carestream Hlth Inc, Rochester, NY USA
[3] Johns Hopkins Univ, Dept Radiol, Baltimore, MD USA
[4] Johns Hopkins Univ, Dept Neurol, Baltimore, MD 21218 USA
关键词
Cone-Beam CT; intracranial hemorrhage; task based detectability; system design; OPTIMIZATION; PERFORMANCE;
D O I
10.1117/12.2216544
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Purpose: Prompt and reliable detection of intracranial hemorrhage (ICH) has substantial clinical impact in diagnosis and treatment of stroke and traumatic brain injury. This paper describes the design, development, and preliminary performance characterization of a dedicated cone-beam CT (CBCT) head scanner prototype for imaging of acute ICH. Methods: A task-based image quality model was used to analyze the detectability index as a function of system configuration, and hardware design was guided by the results of this model-based optimization A robust artifact correction pipeline was developed using GPU-accelerated Monte Carlo (MC) scatter simulation, beam hardening corrections, detector veiling glare, and lag deconvolution. An iterative penalized weighted least-squares (PWLS) reconstruction framework with weights adjusted for artifact-corrected projections was developed. Various bowtie filters were investigated for potential dose and image quality benefits, with a MC-based tool providing estimates of spatial dose distribution. Results: The initial prototype will feature a source-detector distance of 1000 mm and source-axis distance of 550 mm, a 43x43 cm(2) flat panel detector, and a 15 rotating anode x-ray source with 15 kW power and 0.6 focal spot size. Artifact correction reduced image nonuniformity by similar to 250 HU, and PWLS reconstruction with modified weights improved the contrast to noise ratio by 20%. Inclusion of a bowtie filter can potentially reduce dose by 50% and improve CNR by 25%. Conclusions: A dedicated CBCT system capable of imaging millimeter-scale acute ICH was designed. Preliminary findings support feasibility of point-of-care applications in 1131 and stroke imaging, with clinical studies beginning on a prototype.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Diagnostic Image Quality Evaluation of a Dedicated Extremity Cone-Beam CT Scanner: Pre-Clinical Studies and First Clinical Results
    Muhit, A.
    Zbijewski, W.
    Stayman, J.
    Thawait, G.
    Yorkston, J.
    Foos, D.
    Packard, N.
    Yang, D.
    Senn, R.
    Carrino, J.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3973 - 3973
  • [42] Detection of impurity content in wheat based on cone-beam CT
    Wang, Gangyang
    Zhu, Chunhua
    Wang, Jianhou
    2022 IEEE INTERNATIONAL CONFERENCE ON TRUST, SECURITY AND PRIVACY IN COMPUTING AND COMMUNICATIONS, TRUSTCOM, 2022, : 1286 - 1289
  • [43] Detection of Collaterals from Cone-Beam CT Images in Stroke
    Abd Aziz, Azrina
    Izhar, Lila Iznita
    Asirvadam, Vijanth Sagayan
    Tang, Tong Boon
    Ajam, Azimah
    Omar, Zaid
    Muda, Sobri
    SENSORS, 2021, 21 (23)
  • [44] Cone-beam CT angiography to assess the microvascular anatomy of intracranial arterial dissections
    Diana, Francesco
    Raz, Eytan
    Biraschi, Francesco
    Romano, Daniele G.
    Toccaceli, Giada
    Shapiro, Maksim
    Peschillo, Simone
    NEURORADIOLOGY JOURNAL, 2022, 35 (04): : 527 - 532
  • [45] Task-based modeling and optimization of a cone-beam CT scanner for musculoskeletal imaging
    Prakash, P.
    Zbijewski, W.
    Gang, G. J.
    Ding, Y.
    Stayman, J. W.
    Yorkston, J.
    Carrino, J. A.
    Siewerdsen, J. H.
    MEDICAL PHYSICS, 2011, 38 (10) : 5612 - 5629
  • [46] Design Optimization of Segmented Scintillators for Megavoltage Cone-Beam CT
    Liu, L.
    Antonuk, L.
    El-Mohri, Y.
    Zhao, Q.
    Jiang, H.
    MEDICAL PHYSICS, 2014, 41 (06) : 540 - 540
  • [47] Exact FBP reconstruction for a cone-beam CT scanner with independent rotations of the source and the detector
    Rit, S.
    Clackdoyle, R.
    Keuschnigg, P.
    Steininger, P.
    RADIOTHERAPY AND ONCOLOGY, 2015, 115 : S198 - S199
  • [48] Micro cone-beam CT scanner based on X-ray polycapillary optics
    Zhou La-Zhen
    Xia Wen-Jing
    Xu Qian-Qian
    Chen Zan
    Li Fang-Zuo
    Liu Zhi-Guo
    Sun Tian-Xi
    ACTA PHYSICA SINICA, 2022, 71 (09)
  • [49] DentiiScan: the First Cone-Beam CT Scanner for Dental and Maxillofacial Imaging Developed in Thailand
    Thongvigitmanee, Saowapak S.
    Kasemsarn, Suthasinee
    Sirisalee, Pasu
    Aootaphao, Sorapong
    Rajruangrabin, Jartuwat
    Yampri, Pinyo
    Srivongsa, Tanapon
    Sa-Ing, Vera
    Thajchayapong, Pairash
    2013 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2013,
  • [50] Positional dependence of the CT number with use of a cone-beam CT scanner for an electron density phantom in particle beam therapy
    Kusano Y.
    Uesaka S.
    Yajima K.
    Kumagai M.
    Mizuno H.
    Mori S.
    Radiological Physics and Technology, 2013, 6 (1) : 241 - 247