Do Stem Taper Microgrooves Influence Taper Corrosion in Total Hip Arthroplasty? A Matched Cohort Retrieval Study

被引:30
|
作者
Arnholt, Christina M. [1 ]
MacDonald, Daniel W. [1 ]
Underwood, Richard J. [2 ]
Guyer, Eric P. [3 ]
Rimnac, Clare M. [4 ]
Kurtz, Steven M. [1 ,2 ]
Mont, Michael A. [5 ]
Klein, Gregg R. [6 ]
Lee, Gwo-Chin [7 ]
Chen, Antonia F. [8 ]
Hamlin, Brian R. [9 ]
Cates, Harold E. [10 ]
Malkani, Arthur L. [11 ,12 ]
Kraay, Matthew J. [4 ]
机构
[1] Drexel Univ, Implant Res Ctr, 3401 Market St,Suite 345, Philadelphia, PA 19104 USA
[2] Exponent Inc, Philadelphia, PA USA
[3] Exponent Inc, Atlanta, GA USA
[4] Case Western Reserve Univ, Ctr Evaluat Implant Performance, Cleveland, OH 44106 USA
[5] Sinai Hosp Baltimore, Baltimore, MD USA
[6] Hartzband Ctr Hip & Knee Replacement, Paramus, NJ USA
[7] Univ Penn Hlth Syst, Presbyterian Med Ctr, Philadelphia, PA USA
[8] Thomas Jefferson Univ Hosp, Rothman Inst, Philadelphia, PA 19107 USA
[9] Univ Pittsburgh, Med Ctr, Pittsburgh, PA USA
[10] Tennessee Orthopaed Clin, Knoxville, TN USA
[11] Jewish Hosp, Louisville, KY USA
[12] St Marys Hlth Care, Louisville, KY USA
来源
JOURNAL OF ARTHROPLASTY | 2017年 / 32卷 / 04期
关键词
mechanically assisted crevice corrosion; fretting; corrosion; taper; microgrooves; ON-METAL BEARINGS; SURFACE-TOPOGRAPHY; FAILURE; NECK; REPLACEMENTS; MODULARITY; TRUNNIONS; IMPLANTS; JUNCTION; WEAR;
D O I
10.1016/j.arth.2016.11.018
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Previous studies identified imprinting of the stem morphology onto the interior head bore, leading researchers to hypothesize an influence of taper topography on mechanically assisted crevice corrosion. The purpose of this study was to analyze whether microgrooved stem tapers result in greater fretting corrosion damage than smooth stem tapers. Methods: A matched cohort of 120 retrieved head-stem pairs from metal-on-polyethylene bearings was created controlling for implantation time, flexural rigidity, apparent length of engagement, and head size. There were 2 groups of 60 heads each, mated with either smooth or microgrooved stem tapers. A high-precision roundness machine was used to measure and categorize the surface morphology. Fretting corrosion damage at the head-neck junction was characterized using the Higgs-Goldberg scoring method. Fourteen of the most damaged heads were analyzed for the maximum depth of material loss and focused ion beam cross-sectioned to view oxide and base metal. Results: Fretting corrosion damage was not different between the 2 cohorts at the femoral head (P = .14, Mann-Whitney) or stem tapers (P = .35). There was no difference in the maximum depths of material loss between the cohorts (P = .71). Cross-sectioning revealed contact damage, signs of micro-motion, and chromium-rich oxide layers in both cohorts. Microgroove imprinting did not appear to have a different effect on the fretting corrosion behavior. Conclusion: The results of this matched cohort retrieval study do not support the hypothesis that taper surfaces with microgrooved stems exhibit increased in vivo fretting corrosion damage or material release. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1363 / 1373
页数:11
相关论文
共 50 条
  • [1] Taper corrosion: a complication of total hip arthroplasty
    Morlock, Michael M.
    Hube, Robert
    Wassilew, Georgi
    Prange, Felix
    Huber, Gerd
    Perka, Carsten
    EFORT OPEN REVIEWS, 2020, 5 (11) : 776 - 784
  • [2] Do Ceramic Femoral Heads Reduce Taper Fretting Corrosion in Hip Arthroplasty? A Retrieval Study
    Kurtz, Steven M.
    Kocagoez, Sevi B.
    Hanzlik, Josa A.
    Underwood, Richard J.
    Gilbert, Jeremy L.
    MacDonald, Daniel W.
    Lee, Gwo-Chin
    Mont, Michael A.
    Kraay, Matthew J.
    Klein, Gregg R.
    Parvizi, Javad
    Rimnac, Clare M.
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2013, 471 (10) : 3270 - 3282
  • [3] What Do We Know About Taper Corrosion in Total Hip Arthroplasty?
    Jacobs, J. J.
    Cooper, H. J.
    Urban, R. M.
    Wixson, R. L.
    Della Valle, C. J.
    JOURNAL OF ARTHROPLASTY, 2014, 29 (04): : 668 - 669
  • [4] Does Taper Design Affect Taper Fretting Corrosion in Ceramic-on-Polyethylene Total Hip Arthroplasty? A Retrieval Analysis
    Siljander, Matthew P.
    Gehrke, Corinn K.
    Wheeler, Samantha D.
    Sobh, Ali H.
    Moore, Drew D.
    Flierl, Michael A.
    Baker, Erin A.
    JOURNAL OF ARTHROPLASTY, 2019, 34 (07): : S366 - +
  • [5] Head Taper Corrosion Causing Head Bottoming Out and Consecutive Gross Stem Taper Failure in Total Hip Arthroplasty
    Morlock, Michael M.
    Dickinson, Emilie C.
    Guenther, Klaus-Peter
    Buente, Dennis
    Polster, Valerie
    JOURNAL OF ARTHROPLASTY, 2018, 33 (11): : 3581 - 3590
  • [6] Early Outcomes of Revision Surgery for Taper Corrosion of Dual Taper Total Hip Arthroplasty in 187 Patients
    Dimitriou, Dimitris
    Liow, Ming Han Lincoln
    Tsai, Tsung-Yuan
    Leone, William A.
    Li, Guoan
    Kwon, Young-Min
    JOURNAL OF ARTHROPLASTY, 2016, 31 (07): : 1549 - 1554
  • [7] Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty
    Kilian Elia Stockhausen
    Christoph Riedel
    Alex Victoria Belinski
    Dorothea Rothe
    Thorsten Gehrke
    Felix Klebig
    Matthias Gebauer
    Michael Amling
    Mustafa Citak
    Björn Busse
    Scientific Reports, 11
  • [8] Variability in stem taper surface topography affects the degree of corrosion and fretting in total hip arthroplasty
    Stockhausen, Kilian Elia
    Riedel, Christoph
    Belinski, Alex Victoria
    Rothe, Dorothea
    Gehrke, Thorsten
    Klebig, Felix
    Gebauer, Matthias
    Amling, Michael
    Citak, Mustafa
    Busse, Bjorn
    SCIENTIFIC REPORTS, 2021, 11 (01)
  • [9] Head-neck taper corrosion following total hip arthroplasty with Stryker Meridian stem
    Patel, Sonika
    Talmo, Carl T.
    Nandi, Sumon
    HIP INTERNATIONAL, 2016, 26 (06) : E49 - E51
  • [10] Corrosion of a Hip Stem With a Modular Neck Taper Junction A Retrieval Study of 16 Cases
    Kop, Alan M.
    Swarts, Eric