Near-patient testing for coagulopathy after cardiac surgery

被引:7
|
作者
Johi, RR
Cross, MH
Hansbro, SD
机构
[1] Gen Infirm, Dept Anaesthesia, Leeds LS1 3EX, W Yorkshire, England
[2] Gen Infirm, Dept Clin Perfus, Leeds LS1 3EX, W Yorkshire, England
关键词
complications; coagulopathy; heart; cardiopulmonary bypass; measurement techniques; coagulation; surgery; cardiovascular;
D O I
10.1093/bja/aeg032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. We assessed the accuracy and precision of a new near-patient testing system (Hemochron(R) Response) by measuring prothrombin time and activated partial thromboplastin time (PT and APTT) in 50 patients undergoing cardiac surgery using cardiopulmonary bypass and comparing the results with laboratory assays. Methods. Blood samples were taken at the beginning of surgery and the PT and APTT was measured both in the laboratory and by the Hemochron(R) Response. The tests were repeated 30 min after reversal of heparin with protamine. Results. Before bypass, the bias for PT was only +0.34, with small 95% limits of agreement. Making the same measurements after bypass, the Hemochron(R) Response under-read and the bias was -3.27, with an increase of the 95% limits of agreement. With the APTT, the bias and the 95% limits of agreement were greater before bypass, and became even wider after bypass. Conclusions. We found good agreement in the PT and clinically acceptable levels of agreement in the APTT during the pre-bypass period. After bypass, bias became greater for both PT and APTT and the limits of agreement could be clinically unacceptable.
引用
收藏
页码:499 / 501
页数:3
相关论文
共 50 条
  • [31] An evaluation of near-patient testing for Helicobacter pylori in general practice
    Jones, R
    Phillips, I
    Felix, G
    Tait, C
    [J]. ALIMENTARY PHARMACOLOGY & THERAPEUTICS, 1997, 11 (01) : 101 - 105
  • [32] THE IMPACT OF NEAR-PATIENT TESTING ON THE ORGANIZATION AND COSTS OF AN ANTICONVULSANT CLINIC
    ELLIOT, K
    WATSON, ID
    TSINTIS, P
    GRAY, J
    STEWART, MJ
    KADR, H
    LAWSON, DH
    [J]. THERAPEUTIC DRUG MONITORING, 1990, 12 (05) : 434 - 437
  • [33] Near-patient stability studies
    Barnes, AR
    [J]. JOURNAL OF CLINICAL PHARMACY AND THERAPEUTICS, 1996, 21 (01) : 49 - 55
  • [34] A solution to the problem of quality control for near-patient testing in general practice
    Higginson, BM
    White, P
    [J]. CURRENT PERSPECTIVES IN HEALTHCARE COMPUTING, CONFERENCE, 1997, : 281 - 289
  • [35] Comparison of two near-patient testing instruments with two different thromboplastins
    Testa, S
    Morstabilini, P
    Denti, N
    Pigoli, G
    Ferrari, L
    [J]. THROMBOSIS AND HAEMOSTASIS, 1999, : 399 - 400
  • [36] Comparison of two near-patient testing instruments with two different thromboplastins
    Testa, S
    Morstabilini, P
    Denti, N
    Pigoli, G
    Ferrari, L
    [J]. THROMBOSIS AND HAEMOSTASIS, 1999, : 859 - 859
  • [37] Near-patient testing - point-of-care or point of costs and convenience?
    Jahn, UR
    Van Aken, H
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2003, 90 (04) : 425 - 427
  • [38] Application guide: training of professional users of devices for near-patient testing
    Barabas, Nicola
    Bietenbeck, Andreas
    [J]. LABORATORIUMSMEDIZIN-JOURNAL OF LABORATORY MEDICINE, 2017, 41 (05):
  • [39] Near-patient parathyroid hormone immunoassay for intraoperative monitoring in parathyroid surgery
    Kao, PC
    van Heerden, JA
    [J]. JOURNAL OF CLINICAL LIGAND ASSAY, 1999, 22 (02): : 123 - 128
  • [40] Visible/Near-Infrared spectrophotometric blood typing sensor for automated near-patient testing
    Anthony, Steven R.
    Ramasubramanian, M. K.
    [J]. 2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 1980 - 1983