Near-patient testing for coagulopathy after cardiac surgery

被引:7
|
作者
Johi, RR
Cross, MH
Hansbro, SD
机构
[1] Gen Infirm, Dept Anaesthesia, Leeds LS1 3EX, W Yorkshire, England
[2] Gen Infirm, Dept Clin Perfus, Leeds LS1 3EX, W Yorkshire, England
关键词
complications; coagulopathy; heart; cardiopulmonary bypass; measurement techniques; coagulation; surgery; cardiovascular;
D O I
10.1093/bja/aeg032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. We assessed the accuracy and precision of a new near-patient testing system (Hemochron(R) Response) by measuring prothrombin time and activated partial thromboplastin time (PT and APTT) in 50 patients undergoing cardiac surgery using cardiopulmonary bypass and comparing the results with laboratory assays. Methods. Blood samples were taken at the beginning of surgery and the PT and APTT was measured both in the laboratory and by the Hemochron(R) Response. The tests were repeated 30 min after reversal of heparin with protamine. Results. Before bypass, the bias for PT was only +0.34, with small 95% limits of agreement. Making the same measurements after bypass, the Hemochron(R) Response under-read and the bias was -3.27, with an increase of the 95% limits of agreement. With the APTT, the bias and the 95% limits of agreement were greater before bypass, and became even wider after bypass. Conclusions. We found good agreement in the PT and clinically acceptable levels of agreement in the APTT during the pre-bypass period. After bypass, bias became greater for both PT and APTT and the limits of agreement could be clinically unacceptable.
引用
收藏
页码:499 / 501
页数:3
相关论文
共 50 条
  • [21] Near-patient testing will not improve the control of sexually transmitted infections
    Dean, G. L.
    [J]. SEXUALLY TRANSMITTED INFECTIONS, 2006, 82 (06) : 509 - 512
  • [22] Near-patient platelet function testing in patients undergoing coronary artery surgery: a pilot study
    Reece, M. J.
    Klein, A. A.
    Salviz, E. A.
    Hastings, A.
    Ashworth, A.
    Freeman, C.
    Luddington, R. J.
    Nair, S.
    Besser, M. W.
    [J]. ANAESTHESIA, 2011, 66 (02) : 97 - 103
  • [23] Coagulopathy after cardiac surgery - In response
    Duggan, Edel
    Ryan, Thomas
    [J]. ANESTHESIA AND ANALGESIA, 2007, 105 (05): : 1514 - 1515
  • [24] An evaluation of near-patient testing of anticoagulant control in general practice
    Edgeworth, Amanda
    Coles, Edward C.
    [J]. INTERNATIONAL JOURNAL OF HEALTH CARE QUALITY ASSURANCE, 2010, 23 (04) : 410 - 421
  • [25] Near-patient testing devices to monitor oral anticoagulant therapy
    Tripodi, A
    Chantarangkul, V
    Mannucci, PM
    [J]. BRITISH JOURNAL OF HAEMATOLOGY, 2001, 113 (04) : 847 - 852
  • [26] Point-of-care testing, near-patient testing and patient self-testing: warning points
    Banfi, Giuseppe
    Bozic, Borut
    Cihan, Murat
    Pasalic, Daria
    Pennestri, Federico
    Plebani, Mario
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024,
  • [27] Near-patient testing devices to monitor vitamin K antagonists
    Brionne-Francois, Marie
    Le Querrec, Agnes
    Lasne, Dominique
    [J]. ANNALES DE BIOLOGIE CLINIQUE, 2013, 71 : 59 - 69
  • [28] Near-patient testing holds most promise for acute conditions
    Cals, Jochen W. L.
    Geersing, Geert-Jan
    [J]. BRITISH JOURNAL OF GENERAL PRACTICE, 2010, 60 (575): : 450 - 451
  • [29] Comparison of structured use of routine laboratory tests or near-patient assessment with clinical judgement in the management of bleeding after cardiac surgery
    Avidan, MS
    Alcock, EL
    Da Fonseca, J
    Ponte, J
    Desai, JB
    Despotis, GJ
    Hunt, BJ
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2004, 92 (02) : 178 - 186
  • [30] NEAR-PATIENT TESTING - IMPLICATIONS FOR LABORATORY-BASED PROFESSIONS
    MARKS, V