Near-patient testing for coagulopathy after cardiac surgery

被引:7
|
作者
Johi, RR
Cross, MH
Hansbro, SD
机构
[1] Gen Infirm, Dept Anaesthesia, Leeds LS1 3EX, W Yorkshire, England
[2] Gen Infirm, Dept Clin Perfus, Leeds LS1 3EX, W Yorkshire, England
关键词
complications; coagulopathy; heart; cardiopulmonary bypass; measurement techniques; coagulation; surgery; cardiovascular;
D O I
10.1093/bja/aeg032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Background. We assessed the accuracy and precision of a new near-patient testing system (Hemochron(R) Response) by measuring prothrombin time and activated partial thromboplastin time (PT and APTT) in 50 patients undergoing cardiac surgery using cardiopulmonary bypass and comparing the results with laboratory assays. Methods. Blood samples were taken at the beginning of surgery and the PT and APTT was measured both in the laboratory and by the Hemochron(R) Response. The tests were repeated 30 min after reversal of heparin with protamine. Results. Before bypass, the bias for PT was only +0.34, with small 95% limits of agreement. Making the same measurements after bypass, the Hemochron(R) Response under-read and the bias was -3.27, with an increase of the 95% limits of agreement. With the APTT, the bias and the 95% limits of agreement were greater before bypass, and became even wider after bypass. Conclusions. We found good agreement in the PT and clinically acceptable levels of agreement in the APTT during the pre-bypass period. After bypass, bias became greater for both PT and APTT and the limits of agreement could be clinically unacceptable.
引用
收藏
页码:499 / 501
页数:3
相关论文
共 50 条
  • [1] Near-patient coagulation testing to predict bleeding after cardiac surgery: a cohort study
    Mumford, Andrew D.
    Harris, Jessica
    Plummer, Zoe
    Lee, Kurtis
    Verheyden, Veerle
    Reeves, Barnaby C.
    Rogers, Chris A.
    Angelini, Gianni D.
    Murphy, Gavin J.
    [J]. RESEARCH AND PRACTICE IN THROMBOSIS AND HAEMOSTASIS, 2017, 1 (02) : 242 - 251
  • [2] The Value of Routine Near-patient Haemostasis Testing for Predicting Bleeding After Cardiac Surgery
    Mumford, Andrew
    Harris, Jessica
    Plummer, Zoe
    Verheyden, Veerle
    Lee, Kurtis
    Rogers, Chris
    Reeves, Barney C.
    Murphy, Gavin J.
    [J]. CIRCULATION, 2015, 132
  • [3] Near-patient testing
    Jahn, UR
    Van Aken, H
    [J]. BRITISH JOURNAL OF ANAESTHESIA, 2003, 91 (04) : 608 - 608
  • [4] DEVELOPMENTS IN NEAR-PATIENT TESTING
    HAILEY, DM
    LEA, AR
    [J]. MEDICAL LABORATORY SCIENCES, 1990, 47 (04): : 319 - 325
  • [5] Near-patient testing by proxy
    Johnston, JD
    Jupp, RA
    Mitchell, WD
    [J]. ANNALS OF CLINICAL BIOCHEMISTRY, 1997, 34 : 104 - 105
  • [6] NEAR-PATIENT LABORATORY TESTING
    FARR, AD
    [J]. MEDICAL LABORATORY SCIENCES, 1990, 47 (04): : 249 - 250
  • [7] Near-Patient Testing for the Diagnosis of Vaginitis
    Hall, Geraldine S.
    Sautter, Robert L.
    [J]. POINT OF CARE, 2009, 8 (02): : 87 - 89
  • [8] Near-patient testing in primary care
    Fitzmaurice, D
    [J]. BRITISH JOURNAL OF GENERAL PRACTICE, 2004, 54 (506): : 650 - 651
  • [9] Near-patient testing in primary care
    Khunti, Kamlesh
    [J]. BRITISH JOURNAL OF GENERAL PRACTICE, 2010, 60 (572): : 157 - 158
  • [10] Diagnostic performance of near-patient testing for influenza
    Beckmann, Christiane
    Hirsch, Hans H.
    [J]. JOURNAL OF CLINICAL VIROLOGY, 2015, 67 : 43 - 46