Three-dimensional ultrashort echo time cones (3D UTE-Cones) magnetic resonance imaging of entheses and tendons

被引:37
|
作者
Chen, Bimin [1 ,2 ]
Zhao, Yinghua [1 ,3 ]
Cheng, Xin [4 ]
Ma, Yajun [1 ]
Chang, Eric Y. [1 ,5 ]
Kavanaugh, Arthur [6 ]
Liu, Sirun [2 ]
Du, Jiang [1 ]
机构
[1] Univ Calif San Diego, Dept Radiol, 200 West Arbor Dr, San Diego, CA 92103 USA
[2] Jinan Univ, Dept Radiol, Guangzhou, Guangdong, Peoples R China
[3] Southern Med Univ, Dept Radiol, Guangzhou, Guangdong, Peoples R China
[4] Jinan Univ, Dept Histol & Embryol, Guangzhou, Guangdong, Peoples R China
[5] VA San Diego Healthcare Syst, Radiol Serv, San Diego, CA USA
[6] Univ Calif San Diego, Dept Med, San Diego, CA 92103 USA
关键词
UTE; Cones; T2*; MTR; Enthesis; Tendon; EARLY PSORIATIC-ARTHRITIS; RHEUMATOID-ARTHRITIS; PORE-WATER; TE UTE; MRI; SEQUENCES; SPONDYLOARTHRITIS; EPIDEMIOLOGY; DEFINITIONS; DACTYLITIS;
D O I
10.1016/j.mri.2017.12.034
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
We present three-dimensional ultrashort echo time Cones (3D-UTE-Cones) imaging as well as quantification of T2* and magnetization transfer ratio (MTR) of Achilles tendon and its enthesis of healthy volunteers and patients with psoriatic arthritis (PsA) using a 3 T scanner. Quantitative T2*, T2 and magnetization transfer ratio (MTR) measurements of Achilles tendon and its enthesis were performed on healthy volunteers (n = 7) and PsA patients (n = 9) with 3D-UTE-Cones and clinical sequences at 3 T. T2* was measured via single-component fitting of UTE images from two interleaved dual echo 3D-UTE-Cones acquisitions. MTR was measured with a dual echo 3D-UTE-Cones-MT sequence. Clinical morphological imaging and quantification of T2 with a Carr-PurcellMeiboom-Gill (CPMG) sequence and MTR with a gradient recalled echo (GRE) sequence were also performed on each subject. T2* and MTR were analyzed for the two groups and the significance was evaluated. The 3D-UTE-Cones sequence provided high resolution imaging of entheses and tendons. Cones-T2* and Cones-MTR values were significantly higher for the PsA patients. GRE-MTR values showed no significant differences between the groups. No reliable T2 measurement could be achieved with the CPMG sequence due to insufficient signal from entheses and tendons. The 3D-UTE-Cones sequences can be used for morphological and quantitative evaluation of entheses and tendons in PsA patients.
引用
下载
收藏
页码:4 / 9
页数:6
相关论文
共 50 条
  • [41] Comparison of T2*mapping between regular echo time and ultrashort echo time with 3D cones at 3 tesla for knee meniscus
    Yi, Jisook
    Lee, Young Han
    Song, Ho-Taek
    Suh, Jin-Suck
    MEDICINE, 2018, 97 (48)
  • [42] Clinical application of ultrashort echo time (UTE) and zero echo time (ZTE) magnetic resonance (MR) imaging in the evaluation of osteoarthritis
    Karen Y. Cheng
    Dina Moazamian
    Yajun Ma
    Hyungseok Jang
    Saeed Jerban
    Jiang Du
    Christine B. Chung
    Skeletal Radiology, 2023, 52 : 2149 - 2157
  • [43] Clinical application of ultrashort echo time (UTE) and zero echo time (ZTE) magnetic resonance (MR) imaging in the evaluation of osteoarthritis
    Cheng, Karen Y. Y.
    Moazamian, Dina
    Ma, Yajun
    Jang, Hyungseok
    Jerban, Saeed
    Du, Jiang
    Chung, Christine B. B.
    SKELETAL RADIOLOGY, 2023, 52 (11) : 2149 - 2157
  • [44] Magic angle effect on adiabatic T1ρ imaging of the Achilles tendon using 3D ultrashort echo time cones trajectory
    Wu, Mei
    Ma, Yajun
    Wan, Lidi
    Jerban, Saeed
    Jang, Hyungseok
    Chang, Eric Y.
    Du, Jiang
    NMR IN BIOMEDICINE, 2020, 33 (08)
  • [45] 3D printed phantoms mimicking cortical bone for the assessment of ultrashort echo time magnetic resonance imaging
    Rai, Robba
    Manton, David
    Jameson, Michael G.
    Josan, Sonal
    Barton, Michael B.
    Holloway, Lois C.
    Liney, Gary P.
    MEDICAL PHYSICS, 2018, 45 (02) : 758 - 766
  • [46] Kidney Stone Imaging with 3D Ultra-Short Echo Time (UTE) Magnetic Resonance Imaging. A Phantom Study
    Ibrahim, El-Sayed H.
    Pooley, Robert A.
    Bridges, Mellena D.
    Cernigliaro, Joseph G.
    Haley, William E.
    2014 36TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2014, : 2356 - 2359
  • [47] Optimized three-dimensional ultrashort echo time: Magnetic resonance fingerprinting for myelin tissue fraction mapping
    Zhou, Zihan
    Li, Qing
    Liao, Congyu
    Cao, Xiaozhi
    Liang, Hui
    Chen, Quan
    Pu, Run
    Ye, Huihui
    Tong, Qiqi
    He, Hongjian
    Zhong, Jianhui
    HUMAN BRAIN MAPPING, 2023, 44 (06) : 2209 - 2223
  • [48] Knee osteochondral junction imaging using a fast 3D T1-weighted ultrashort echo time cones sequence at 3T
    Cai, Zhenyu
    Wei, Zhao
    Wu, Mei
    Jerban, Saeed
    Jang, Hyungseok
    Li, Shaolin
    Yuan, Xuchun
    Ma, Ya-Jun
    MAGNETIC RESONANCE IMAGING, 2020, 73 : 76 - 83
  • [49] Knee menisci segmentation and relaxometry of 3D ultrashort echo time cones MR imaging using attention U-Net with transfer learning
    Byra, Michal
    Wu, Mei
    Zhang, Xiaodong
    Jang, Hyungseok
    Ma, Ya-Jun
    Chang, Eric Y.
    Shah, Sameer
    Du, Jiang
    MAGNETIC RESONANCE IN MEDICINE, 2020, 83 (03) : 1109 - 1122
  • [50] Three-dimensional ultrashort echo-time imaging using a FLORET trajectory
    Robison, Ryan K.
    Anderson, Ashley G., III
    Pipe, James G.
    MAGNETIC RESONANCE IN MEDICINE, 2017, 78 (03) : 1038 - 1049