Determination of output factors for small proton therapy fields

被引:24
|
作者
Fontenot, Jonas D.
Newhauser, Wayne D.
Bloch, Charles
White, R. Allen
Titt, Uwe
Starkschall, George
机构
[1] Univ Texas, MD Anderson Canc Ctr, Dept Radiat Phys, Houston, TX 77030 USA
[2] Univ Texas, Grad Sch Biomed Sci, Houston, TX 77030 USA
[3] Univ Texas, MD Anderson Canc Ctr, Dept Biostat & Appl Math, Houston, TX 77030 USA
[4] Baylor Coll Med, Dept Radiol, Houston, TX 77030 USA
关键词
proton therapy; Monte Carlo; patient-specific dose calibration; range compensator;
D O I
10.1118/1.2428406
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Current protocols for the measurement of proton dose focus on measurements under reference conditions; methods for measuring dose under patient-specific conditions have not been standardized. In particular, it is unclear whether dose in patient-specific fields can be determined more reliably with or without the presence of the patient-specific range compensator. The aim of this study was to quantitatively assess the reliability of two methods for measuring dose per monitor unit (DIMU) values for small-field treatment portals: one with the range compensator and one without the range compensator. A Monte Carlo model of the Proton Therapy Center-Houston double-scattering nozzle was created, and estimates of DIMU values were obtained from 14 simulated treatments of a simple geometric patient model. Field-specific DIMU calibration measurements were simulated with a dosimeter in a water phantom with and without the range compensator. DIMU values from the simulated calibration measurements were compared with DIMU values from the corresponding treatment simulation in the patient model. To evaluate the reliability of the calibration measurements, six metrics and four figures of merit were defined to characterize accuracy, uncertainty, the standard deviations of accuracy and uncertainty, worst agreement, and maximum uncertainty. Measuring DIMU without the range compensator provided superior results for five of the six metrics and for all four figures of merit. The two techniques yielded different results primarily because of high-dose gradient regions introduced into the water phantom when the range compensator was present. Estimated uncertainties (similar to 1 mm) in the position of the dosimeter in these regions resulted in large uncertainties and high variability in DIMU values. When the range compensator was absent, these gradients were minimized and DIMU values were less sensitive to dosimeter positioning errors. We conclude that measuring DIMU without the range compensator present provides more reliable results than measuring it with the range compensator in place. (c) 2007 American Association of Physicists in Medicine.
引用
收藏
页码:489 / 498
页数:10
相关论文
共 50 条
  • [1] AN ALTERNATIVE METHOD FOR THE DETERMINATION OF OUTPUT FACTORS FOR SMALL FIELDS
    CapriIe, P.
    Hartmann, G. H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S34 - S35
  • [2] Empirical Determination of Output Factors for Proton Therapy Fields Using a Uniform Scanning Proton Beam and 10cm Snout
    Aaron, E.
    Cheng, C.
    Lamba, M.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [3] Small fields output factors and correction factors determination for a linac with circular cones
    Girardi, A.
    Fiandra, C.
    Gallio, E.
    Giglioli, F. R.
    Ragona, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S216 - S216
  • [4] Validation of a new formulism and the related correction factors on output factor determination for small photon fields
    Wang, Yizhen
    Younge, Kelly
    Nielsen, Michelle
    Mutanga, Theodore
    Cui, Congwu
    Das, Indra J.
    [J]. MEDICAL PHYSICS, 2014, 41 (08) : 23 - 23
  • [5] Small Fields Output Factors Measurement and Correction Factor Determination for Several Active and Passive Detectors
    Sharma, D.
    Sharma, S.
    Chowdhary, R.
    Pilakkal, S.
    Rashal, S.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3595 - 3595
  • [6] Determination of output factors for small photon beams
    Sauer, O
    Wilbert, J
    Laqua, D
    [J]. MEDICAL PHYSICS, 2005, 32 (06) : 2108 - 2108
  • [7] A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams
    Casar, Bozidar
    Gershkevitsh, Eduard
    Mendez, Ignasi
    Jurkovic, Slaven
    Huq, M. Saiful
    [J]. MEDICAL PHYSICS, 2019, 46 (02) : 944 - 963
  • [8] New Improved Methods to Read Output Factors for Small Fields
    Smith, C.
    Bacala, A.
    Pella, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [9] MEASURING OUTPUT FACTORS OF SMALL FIELDS FOR NOVALIS TX UNIT
    Carot, I. Mendez
    Sarvari, A.
    Casar, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S474 - S474
  • [10] Field output factors for small fields: A large multicentre study
    Dufreneix, S.
    Bellec, J.
    Josset, S.
    Vieillevigne, L.
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2021, 81 : 191 - 196