Field output factors for small fields: A large multicentre study

被引:10
|
作者
Dufreneix, S. [1 ]
Bellec, J. [2 ]
Josset, S. [1 ]
Vieillevigne, L. [3 ,4 ]
机构
[1] Inst Cancerol Ouest, Angers, St Herblain, France
[2] Ctr Eugene Marquis, Rennes, France
[3] Inst Univ Canc Toulouse, Inst Claudius Regaud, Toulouse, France
[4] Univ Toulouse 3, Ctr Rech & Cancerol Toulouse, INSERM, ERL5294,CNRS,Oncopole,UMR1037, Toulouse, France
关键词
Small field dosimetry; Output factors; IAEA TRS-483;
D O I
10.1016/j.ejmp.2021.01.001
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The determination of output factors in small field dosimetry is a crucial point, especially when implementing stereotactic radiotherapy (SRT). Herein, a working group of the French medical physicist society (SFPM) was created to collect small field output factors. The objective was to gather and disseminate information on small field output factors based on different detectors for various clinical SRT equipment and measurement configurations. Method: Participants were surveyed for information about their SRT equipment, including the type of linear particle accelerator (linac), collimator settings, measurement conditions for the output factors and the detectors used. Participants had to report both the ratio of detector readings and the correction factors applied as described in the IAEA TRS-483 code of practice for nominal field sizes smaller or equal to 3 cm. Mean field output factors and their associated standard deviations were calculated when data from at least 3 linacs were available. Results: 23 centres were enrolled in the project. Standard deviations of the mean field output factors were systematically smaller than 1.5% for field sizes larger or equal to 1 cm and reached 5% for the smallest field size (0.5 cm). Deviations with published data were smaller than 2% except for the 0.5 cm circular fixed aperture collimator of the CyberKnife where it reached 3.5%. Conclusion: These field output factor values obtained via a large multicentre study can be considered as an external cross verification for any radiotherapy centre starting a SRT program and should help minimize systematic errors when determining small field output factors.
引用
收藏
页码:191 / 196
页数:6
相关论文
共 50 条
  • [1] Defining field output factors in small fields based on dose area product measurements: A feasibility study
    Jurczak, Julien
    Rapp, Benjamin
    Bordy, Jean-Marc
    Josset, Stephanie
    Dufreneix, Stephane
    [J]. MEDICAL PHYSICS, 2024, 51 (05) : 3677 - 3686
  • [2] The influence of small field output factors simulated uncertainties on the calculated dose in VMAT plans for brain metastases: a multicentre study
    Clemente, Stefania
    Falco, Maria Daniela
    Cagni, Elisabetta
    Talamonti, Cinzia
    Boccia, Mafalda
    Gino, Eva
    Lorenzini, Elena
    Rosica, Federica
    Russo, Serenella
    Alparone, Alessandro
    Zefiro, Daniele
    Fiandra, Christian
    [J]. BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1119):
  • [3] Determination of output factors for small proton therapy fields
    Fontenot, Jonas D.
    Newhauser, Wayne D.
    Bloch, Charles
    White, R. Allen
    Titt, Uwe
    Starkschall, George
    [J]. MEDICAL PHYSICS, 2007, 34 (02) : 489 - 498
  • [4] AN ALTERNATIVE METHOD FOR THE DETERMINATION OF OUTPUT FACTORS FOR SMALL FIELDS
    CapriIe, P.
    Hartmann, G. H.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S34 - S35
  • [5] Field Size Definition and Output Factors in Small Field Dosimetry
    Ding, G.
    Das, I.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E969 - E969
  • [6] On the field size definition and field output factors in small field dosimetry
    Ding, George X. X.
    Das, Indra J. J.
    [J]. MEDICAL PHYSICS, 2023, 50 (06) : 3833 - 3841
  • [7] New Improved Methods to Read Output Factors for Small Fields
    Smith, C.
    Bacala, A.
    Pella, S.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [8] MEASURING OUTPUT FACTORS OF SMALL FIELDS FOR NOVALIS TX UNIT
    Carot, I. Mendez
    Sarvari, A.
    Casar, B.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S474 - S474
  • [9] A novel method for the determination of field output factors and output correction factors for small static fields for six diodes and a microdiamond detector in megavoltage photon beams
    Casar, Bozidar
    Gershkevitsh, Eduard
    Mendez, Ignasi
    Jurkovic, Slaven
    Huq, M. Saiful
    [J]. MEDICAL PHYSICS, 2019, 46 (02) : 944 - 963
  • [10] Small fields output factors and correction factors determination for a linac with circular cones
    Girardi, A.
    Fiandra, C.
    Gallio, E.
    Giglioli, F. R.
    Ragona, R.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2016, 119 : S216 - S216