Continuous flow 2H/1H and and18O/16O analysis of water samples with dual inlet precision

被引:223
|
作者
Gehre, M
Geilmann, H
Richter, J
Werner, RA
Brand, WA
机构
[1] Max Planck Inst Biogeochem, D-07701 Jena, Germany
[2] UFZ Helmholtz Ctr Environm Res, D-04318 Leipzig, Germany
关键词
D O I
10.1002/rcm.1672
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A method for isotope ratio analysis of water samples is described comprising an on-line high-temperature reduction technique in a helium carrier gas. Using a gas-tight syringe, injection of 0.5 to 1 muL sample is made through a heated septum into a glassy carbon reactor at temperatures in excess of 1300degreesC. More than 150 injections can be made per day and both isotope ratios of interest delta(2)H and 6180, can be measured with the same setup. The technique has the capability to transfer high-precision stable isotope ratio analysis of water samples from a specialized to a routine laboratory task compatible with other common techniques (automated injection for GC, LC, etc.). Experiments with an emphasis on the reactor design were made in two different laboratories using two different commercially available high-temperature elemental analyser (EA) systems. In the Jena TC/EA unit, sample-to-sample memory (single injection) has been reduced to similar to1% and high precision of about 0.1parts per thousand for delta(18)O and <1parts per thousand for delta(2)H has been achieved by a redesign of the glassy carbon reactor and by redirecting the gas flow of the commercial TC/EA unit. With the modified reactor, the contact of water vapour with surfaces other than glassy carbon is avoided completely. The carrier gas is introduced at the bottom of the reactor thereby flushing the outer tube compartment of the tube-in-tube assembly before entering the active heart of the reactor. With the Leipzig high-temperature reactor (HTP) similar precision was obtained with a minor modification (electropolishing) of the injector metal sleeve. With this system, the temperature dependence of the reaction has been studied between 1100 and 1450degreesC. Complete yield and constant isotope ratio information has been observed only for temperatures above 1325degreesC. For temperatures above 1300degreesC the reactor produces an increasing amount of CO background from reaction of glass carbon with the ceramic tube. This limits the usable temperature to a maximum of 1450degreesC. Relevant gas permeation through the Al2O3 walls has not been detected up to 1600degreesC. Copyright (C) 2004 John Wiley Sons, Ltd.
引用
收藏
页码:2650 / 2660
页数:11
相关论文
共 50 条
  • [21] Chemical and Isotopic (18O/16O and D/1H) Composition of Groundwaters of Central and Southern Primorye
    Chudaeva, V. A.
    Chudaev, O. V.
    Yurchenko, S. G.
    RUSSIAN JOURNAL OF PACIFIC GEOLOGY, 2008, 2 (06) : 512 - 520
  • [22] Chemical and isotopic (18O/16O and D/1H) composition of groundwaters of central and southern Primorye
    V. A. Chudaeva
    O. V. Chudaev
    S. G. Yurchenko
    Russian Journal of Pacific Geology, 2008, 2 : 512 - 520
  • [23] Water isotope ratios D/H, 18O/16O, 17O/16O in and out of clouds map dehydration pathways
    Webster, CR
    Heymsfield, AJ
    SCIENCE, 2003, 302 (5651) : 1742 - 1745
  • [24] Fractionation of D/H and 18O/16O Water Isotopes in a Packed Distillation Column
    Kulov, N. N.
    Polkovnichenko, A. V.
    Lupachev, E. V.
    Rastunova, I. L.
    Magomedbekov, E. P.
    THEORETICAL FOUNDATIONS OF CHEMICAL ENGINEERING, 2020, 54 (03) : 389 - 396
  • [25] Fractionation of D/H and 18O/16O Water Isotopes in a Packed Distillation Column
    N. N. Kulov
    A. V. Polkovnichenko
    E. V. Lupachev
    I. L. Rastunova
    E. P. Magomedbekov
    Theoretical Foundations of Chemical Engineering, 2020, 54 : 389 - 396
  • [26] Measurement of the D/H, 18O/16O, and 17O/16O Isotope Ratios in Water by Laser Absorption Spectroscopy at 2.73 μm
    Wu, Tao
    Chen, Weidong
    Fertein, Eric
    Masselin, Pascal
    Gao, Xiaoming
    Zhang, Weijun
    Wang, Yingjian
    Koeth, Johannes
    Brueckner, Daniela
    He, Xingdao
    SENSORS, 2014, 14 (05): : 9027 - 9045
  • [27] Analytical aspects of measuring 2H/1H and 18O/16O ratios in urine from doubly labelled water studies by high-temperature conversion elemental analyser-isotope-ratio mass spectrometry
    Chery, I.
    Zahariev, A.
    Simon, C.
    Blanc, S.
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2015, 29 (07) : 562 - 572
  • [28] A new continuous flow isotope ratio mass spectrometry system for the analysis of δ2H, δ17O and δ18O of small (120 μg) water samples in atmospheric applications
    Franz, P
    Röckmann, T
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (13) : 1429 - 1435
  • [29] Reactivity Impact of 2H and 16O Elastic Scattering Nuclear Data on Critical Systems with Heavy Water
    Roubtsov, D.
    Kozier, K. S.
    Chow, J. C.
    Plompen, A. J. M.
    Kopecky, S.
    Svenne, J. P.
    Canton, L.
    NUCLEAR DATA SHEETS, 2014, 118 : 414 - 417
  • [30] High-Precision Calculation of the Proportions of Water with δ2H and δ18O, the Cumulative Effect of Evaporation in the Vertical Direction and Depleted δ2H and δ18O of the Shallow Soil Water Caused by Evaporation
    Zeng, Zhenyu
    Zhang, Xiang
    Pan, Guoyan
    Xiao, Yang
    WATER, 2022, 14 (17)