Evaluation of four point-of-care meters for rapid determination of blood lactate concentrations in dogs

被引:31
|
作者
Acierno, Mark J. [1 ]
Mitchell, Mark A. [1 ]
机构
[1] Louisiana State Univ, Sch Vet Med, Dept Vet Clin Sci, Baton Rouge, LA 70810 USA
关键词
D O I
10.2460/javma.230.9.1315
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Objective-To determine whether blood lactate values determined in dogs with 4 commercially available point-of-care meters were in agreement with values determined with a critical care laboratory blood analyzer. Design-Prospective study. Animals-50 dogs evaluated for emergency treatment. Procedures-Blood samples were collected at initial evaluation and processed on 4 point-of-care meters and on a critical care laboratory blood analyzer. Results-All 4 point-of-care lactate meters generated measurements that were in agreement with the hospital's critical care analyzer. Values for agreement (bias) between the 4 point-of-care meters and the critical care analyzer were -0.652 (limits of agreement [LA], -1.958 to 0.6541), -0.670 (LA, -2.110 to 0.769), -0.096 (LA, -2.071 to 1.879), and -0.498 (LA, -2.616 to 1.620), respectively. Conclusions and Clinical Relevance-Despite its prognostic and therapeutic relevance, blood lactate measurement in dogs has been hampered by the inability to perform the test in a timely fashion. Results of the present study indicated that several handheld point-of-care lactate meters provided results that were in agreement with a laboratory critical care blood analyzer.
引用
收藏
页码:1315 / 1318
页数:4
相关论文
共 50 条
  • [31] Evaluation of point-of-care maternal venous lactate testing in normal pregnancy
    Maguire, Patrick J.
    Finlay, Janna
    Power, Karen A.
    Harley, Ruth
    Mhurchu, Muireann Ni
    Sheehan, Sharon R.
    Fanning, Rebecca A.
    Turner, Michael J.
    JOURNAL OF MATERNAL-FETAL & NEONATAL MEDICINE, 2016, 29 (16): : 2607 - 2610
  • [32] Use of point-of-care glucose meters in the intensive care units
    Roman, A.
    Preiser, J. -C.
    REANIMATION, 2011, 20 (04): : 350 - 353
  • [33] Rapid Optical Heating of Blood for Clinical Point-of-care Diagnostics
    Catanzaro, Brian E.
    Hill, Ted
    Hankins, Steve
    Gandola, Kent
    ADVANCED BIOMEDICAL AND CLINICAL DIAGNOSTIC SYSTEMS VIII, 2010, 7555
  • [34] Evaluation of a "point-of-care" system for the determination of glycated hemoglobin
    Previtali, Giuseppe
    Amadei, Marina
    Donati, Mirella
    Frasanni, Michela
    Zurlini, Corrado
    BIOCHIMICA CLINICA, 2012, 36 (05) : 388 - 389
  • [35] Multicenter survey for blood contamination on hospital point-of-care glucose testing meters.
    Louie, RF
    Lau, MJ
    Tran, NK
    Tang, Z
    Lee, JH
    Kost, GJ
    CLINICAL CHEMISTRY, 2003, 49 (06) : A152 - A153
  • [36] Quality of Glucose Measurement with Blood Glucose Meters at the Point-of-Care: Relevance of Interfering Factors
    Heinemann, Lutz
    DIABETES TECHNOLOGY & THERAPEUTICS, 2010, 12 (11) : 847 - 857
  • [37] Comparison of three point-of-care blood glucose meters for use in adult and juvenile alpacas
    Tennent-Brown, Brett S.
    Koenig, Amie
    Williamson, Lisa H.
    Boston, Raymond C.
    JAVMA-JOURNAL OF THE AMERICAN VETERINARY MEDICAL ASSOCIATION, 2011, 239 (03): : 380 - 386
  • [38] Clinical Utilization of Point-of-Care Blood L-Lactate Concentrations in Naturally Occurring Respiratory Disease in Feedlot Cattle
    Zeineldin, Mohamed
    Ghanem, Mohamed
    Abd El-Raof, Yassein
    Elattar, Hossam
    PAKISTAN VETERINARY JOURNAL, 2017, 37 (02) : 210 - 214
  • [39] Point-of-care blood gas analysers: a performance evaluation
    St-Louis, P
    CLINICA CHIMICA ACTA, 2001, 307 (1-2) : 139 - 144
  • [40] Evaluation of a rapid, point-of-care test for hepatitis C using oral fluid, blood and plasma
    Lee, S. R.
    Kardos, K.
    Kurtz, L.
    Guillon, G.
    Fischl, M.
    Mendez, A.
    Petrillo, L.
    Mokkapati, V.
    Belenky, A.