In this paper, the detective quantum efficiency (DQE) of cadmium zinc telluride (CZT) detector samples for digital radiography has been measured. Specifically, this study is aimed at investigating the zero-frequency DQE(0) under different X-ray tube and detector parameters. The experimental results of this study indicate that the DQE(0) of the CZT samples is strongly dependent upon the irradiation geometry. This is attributed to the incomplete charge collection process, which can be further improved by controlling the purity of the samples and the contact type.
引用
收藏
页码:1479 / 1484
页数:6
相关论文
共 50 条
  • [1] Detective quantum efficiency (DQE) of CZT semiconductor detectors for digital radiography
    Giakos, GC
    Suryanarayanan, S
    Guntupalli, R
    Odogba, J
    Shah, N
    Vedantham, S
    Chowdhury, S
    Mehta, K
    Sumrain, S
    Patnekar, N
    Moholkar, A
    Kumar, V
    Endorf, RE
    IMTC/O3: PROCEEDINGS OF THE 20TH IEEE INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE, VOLS 1 AND 2, 2003, : 664 - 668
  • [2] EVALUATION OF AREA PHOTON DETECTORS BY A METHOD BASED ON DETECTIVE QUANTUM EFFICIENCY (DQE)
    GRUNER, SM
    MILCH, JR
    REYNOLDS, GT
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1978, 25 (01) : 562 - 565
  • [3] Effects of scintillator on the detective quantum efficiency (DQE) of a digital imaging system
    Farman, TT
    Vandre, RH
    Pajak, JC
    Miller, SR
    Lempicki, A
    Farman, AG
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2006, 101 (02): : 221 - 225
  • [4] Precision of the Detective Quantum Efficiency in Digital Mammography Detectors
    Kim, Dong Sik
    Lee, Eunae
    TENCON 2017 - 2017 IEEE REGION 10 CONFERENCE, 2017, : 1127 - 1130
  • [5] Semiconductor detectors for digital radiography
    Bisogni, MG
    NUOVO CIMENTO DELLA SOCIETA ITALIANA DI FISICA A-NUCLEI PARTICLES AND FIELDS, 1996, 109 (09): : 1303 - 1308
  • [6] Measurement of the detective quantum efficiency (DQE) of digital X-ray detectors according to the novel standard IEC 62220-1
    Illers, H
    Buhr, E
    Hoeschen, C
    RADIATION PROTECTION DOSIMETRY, 2005, 114 (1-3) : 39 - 44
  • [7] A comparison of digital radiography systems in terms of effective detective quantum efficiency
    Bertolini, Marco
    Nitrosi, Andrea
    Rivetti, Stefano
    Lanconelli, Nico
    Pattacini, Pierpaolo
    Ginocchi, Vladimiro
    Iori, Mauro
    MEDICAL PHYSICS, 2012, 39 (05) : 2617 - 2627
  • [8] Detective quantum efficiency (DQE) in PET scanners: A simulation study
    Karpetas, George E.
    Michail, Christos M.
    Fountos, George P.
    Kalyvas, Nektarios I.
    Valais, Ioannis G.
    Kandarakis, Ioannis S.
    Panayiotakis, George S.
    APPLIED RADIATION AND ISOTOPES, 2017, 125 : 154 - 162
  • [9] Software for Determination of Detective Quantum Efficiency of Digital X-Ray Detectors
    Kruchinin S.A.
    Zelikman M.I.
    Biomedical Engineering, 2015, 48 (05) : 270 - 272
  • [10] Effective detective quantum efficiency of two detectors in a prototype digital breast tomosynthesis
    Choi, Seungyeon
    Choi, Sunghoon
    Lee, Donghoon
    Kim, Ye-seul
    Choi, Young-Wook
    Kim, Hee-Joung
    OPTIK, 2019, 178 : 1137 - 1144