Dosimetric characterization of a brachytherapy applicator using MCNP5 modelisation and in-phantom measurements

被引:4
|
作者
Gerardy, I. [1 ]
Rodenas, J. [2 ]
van Dycke, M. [3 ]
Gallardo, S. [2 ]
Ceccolini, Elisa [4 ]
机构
[1] Inst Super Ind Bruxelles, B-1000 Brussels, Belgium
[2] Univ Politecn Valencia, Dept Ingn Quim & Nucl, E-46071 Valencia, Spain
[3] Clin St Jean, B-1000 Brussels, Belgium
[4] Alma Mater Studiorum Univ Bologna, Fac Ingn, Bologna, Italy
关键词
Brachytherapy; MCNP5; Dosimetry; Radiochromic films; Pinpoint chamber;
D O I
10.1016/j.apradiso.2009.10.018
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
A gynaecological applicator consisting of a metallic intra-uterine tube with a plastic vaginal applicator and an HDR Ir-192 source have been simulated with MCNP5 (Monte Carlo code). A solid phantom has been designed to perform measurements around the applicator with radiochromic films. The isodose curves obtained are compared with curves calculated with the F4MESH tally of MCNP5 with a good agreement. A pinpoint ionization chamber has been used to evaluate dose at some reference points. (c) 2009 Elsevier Ltd. All rights reserved.
引用
收藏
页码:735 / 737
页数:3
相关论文
共 50 条
  • [31] Modeling and experimental verification of Biograph TruePoint PET/CT using MCNP5
    Waeleh, Nazreen
    Saripan, M. Iqbal
    Musarudin, Marianie
    Mashohor, Syamsiah
    Saad, Fathinul Fikri Ahmad
    2020 IEEE-EMBS CONFERENCE ON BIOMEDICAL ENGINEERING AND SCIENCES (IECBES 2020): LEADING MODERN HEALTHCARE TECHNOLOGY ENHANCING WELLNESS, 2021, : 319 - 323
  • [32] Characteristics of multihole collimator gamma camera simulation modeled using MCNP5
    Saripan, M. I.
    Hashim, S.
    Mashohor, S.
    Adnan, W. A. Wan
    Marhaban, M. H.
    CURRENT ISSUES OF PHYSICS IN MALAYSIA, 2008, 1017 : 205 - +
  • [33] Dose Estimation in Lung Brachytherapy Using Voxelized Phantom 4DCT and MCNP
    Hadad, K.
    Zaker, N.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [34] Application of the MCNP5 code to the Modeling of vaginal and intra-uterine applicators used in intracavitary brachytherapy: a first approach
    Gerardy, I.
    Rodenas, J.
    van Dycke, M.
    Gallardo, S.
    Tondeur, F.
    INTERNATIONAL WORKSHOP ON MONTE CARLO TECHNIQUES IN RADIOTHERAPY DELIVERY AND VERIFICATION - THIRD MCGILL INTERNATIONAL WORKSHOP, 2008, 102
  • [35] The standfast whole body counter and the sliced bomab phantom: Efficiency as a function of number of sources and energy modeled by MCNP5
    Kramer, Gary H.
    Capello, Kevin
    HEALTH PHYSICS, 2007, 92 (02): : 170 - 175
  • [36] Evaluation of the biological shields of the secondary standard dosimetry laboratory of Ghana using MCNP5
    Deatanyah, P.
    Emi-Reynolds, G.
    Amoako, J.K.
    Arwui, C.C.
    Wotorchi-Gordon, S.
    Lawluvi, H.
    Kpeglo, D.O.
    Annkah, J.
    Research Journal of Applied Sciences, Engineering and Technology, 2012, 4 (06) : 544 - 550
  • [37] MODELING AND OPTIMIZATION OF HPGE DETECTOR GC0518 USING MCNP5 CODE
    Stribrnsky, Branislav
    Hinca, Robert
    Farkas, Gabriel
    Petriska, Martin
    Sluge, Vladimir
    RADIATION PROTECTION DOSIMETRY, 2022, 198 (9-11) : 704 - 711
  • [38] Modeling an HPGe detector response to gamma-rays using MCNP5 code
    Prozorova, I. V.
    Sabitova, R. R.
    Ghal-Eh, N.
    Bedenko, S. V.
    INTERNATIONAL JOURNAL OF MODERN PHYSICS C, 2019, 30 (11):
  • [39] Validation of a Clinical PET Scanner Using Monte Carlo Simulation Code: MCNP5
    Musarudin, Marianie
    Saripan, M. Iqbal
    Mashohor, Syamsiah
    Saad, Wira Hidayat Mohd
    Hashim, Suhairul
    Nordin, Abdul Jalil
    2013 8TH EUROSIM CONGRESS ON MODELLING AND SIMULATION (EUROSIM), 2013, : 36 - 41
  • [40] Fricke gel dosimetric catheters in high dose rate brachytherapy. In phantom dose distribution measurements of a 5 catheter implant
    Carrara, M.
    Gambarini, G.
    Borroni, M.
    Cerrotta, A.
    Fallai, C.
    Invernizzi, M.
    Cavatorta, C.
    Zonca, G.
    RADIATION MEASUREMENTS, 2011, 46 (12) : 1924 - 1927