Validation of a Clinical PET Scanner Using Monte Carlo Simulation Code: MCNP5

被引:0
|
作者
Musarudin, Marianie [1 ]
Saripan, M. Iqbal [1 ]
Mashohor, Syamsiah [1 ]
Saad, Wira Hidayat Mohd [1 ]
Hashim, Suhairul [2 ]
Nordin, Abdul Jalil [3 ]
机构
[1] Univ Putra Malaysia, Fac Engn, Serdang 43400, Malaysia
[2] Univ Teknol Malaysia, Fac Sci, Skudai, Johor, Malaysia
[3] Univ Putra Malaysia, Ctr Diagnost Nucl Imaging, Serdang 43400, Malaysia
关键词
validation; Monte Carlo; PET; WIRE-MESH COLLIMATOR; ATTENUATION CORRECTION; SPATIAL-RESOLUTION; BLOCK DETECTORS; PERFORMANCE; INSTRUMENTATION; DESIGN;
D O I
10.1109/EUROSIM.2013.16
中图分类号
TP301 [理论、方法];
学科分类号
081202 ;
摘要
This paper presents a realistic PET scanner simulation using Monte Carlo code, version MCNP5. The objective of our study was to verify the code used for this simulation by comparing the results obtained from the simulation with those obtained from the measurement did on a real PET scanner. This study will provide a basic benchmark for our further study on PET imaging. We modeled the scanner based on the physical specification of Siemens Biograph TruePoint PET scanner. We recorded the generated list-mode data which contains all the information needed to model the PET processes for instance coincidence photon detection. To account for the statistical fluctuations occur in the detector and photomultiplier tube, a Gaussian energy blurring model was applied to the energy deposited in the detector. The scatter and attenuation correction to correct the effect of scattered and attenuated events also took into account in this study. All of these post-simulation processes were done using a program developed using matlab. To validate the simulation, the simulated and measured energy and spatial resolution were compared. We have successfully modeled a PET system based on MCNP5. We also verified that this simulation result in a good agreement data with the real imaging.
引用
收藏
页码:36 / 41
页数:6
相关论文
共 50 条
  • [1] Uncertainty analysis in environmental radioactivity measurements using the Monte Carlo code MCNP5
    Gallardo, S.
    Querol, A.
    Ortiz, J.
    Rodenas, J.
    Verdu, G.
    Villanueva, J. F.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 214 - 218
  • [2] Validation of the MCNP5 computer code
    Lipilina, E. N.
    Lyutov, V. D.
    Filippova, E. Ya.
    ATOMIC ENERGY, 2006, 101 (02) : 564 - 568
  • [3] Validation of the MCNP5 computer code
    E. N. Lipilina
    V. D. Lyutov
    É. Ya. Filippova
    Atomic Energy, 2006, 101 : 564 - 568
  • [4] Using lattice tools and unfolding methods for hpge detector efficiency simulation with the Monte Carlo code MCNP5
    Querol, A.
    Gallardo, S.
    Rodenas, J.
    Verdu, G.
    RADIATION PHYSICS AND CHEMISTRY, 2015, 116 : 219 - 225
  • [5] Validation of a Monte Carlo simulation of the Inveon PET scanner using GATE
    Lu, Lijun
    Zhang, Houjin
    Bian, Zhaoying
    Ma, Jianhua
    Feng, Qiangjin
    Chen, Wufan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2016, 828 : 170 - 175
  • [6] Evaluation of an extrapolation chamber for dosimetry in computed tomography beams using Monte Carlo code (MCNP5)
    Castro, Maysa C.
    Silva, Natalia F.
    Santos, Lucas R.
    Cintra, Felipe B.
    Caldas, Linda V. E.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167
  • [7] HPGe well detector calibration procedure by MCNP5 Monte Carlo computer code
    Santo, Aline S. E.
    Wasserman, Francis G.
    Conti, Claudio C.
    ANNALS OF NUCLEAR ENERGY, 2012, 46 : 213 - 217
  • [8] Monte Carlo study of an electronic brachytherapy source using MCNP5 and EGSnrc
    Davis, S.
    DeWerd, L.
    MEDICAL PHYSICS, 2007, 34 (06) : 2472 - 2473
  • [9] Validation of the Monte Carlo code MCNP-DSP
    Valentine, TE
    Mihalczo, JT
    ANNALS OF NUCLEAR ENERGY, 1997, 24 (02) : 79 - 98
  • [10] A full characterisation of the GELINA neutron flux distributions and resolution functions with the Monte Carlo code MCNP5
    Ene, D.
    Borcea, C.
    Flaska, M.
    Kopecky, S.
    Negret, A.
    Mondelaers, W.
    Plompen, A. J. M.
    INTERNATIONAL CONFERENCE ON NUCLEAR DATA FOR SCIENCE AND TECHNOLOGY, VOL 1, PROCEEDINGS, 2008, : 397 - +