Influence of Implant Impression Methods, Polymer Materials, and Implant Angulation on the Accuracy of Dental Models

被引:1
|
作者
Djurovic Koprivica, Daniela [1 ]
Puskar, Tatjana [1 ]
Budak, Igor [2 ]
Sokac, Mario [2 ]
Jeremic Knezevic, Milica [1 ]
Maletin, Aleksandra [1 ]
Milekic, Bojana [1 ]
Vukelic, Djordje [2 ]
机构
[1] Univ Novi Sad, Fac Med, Novi Sad 21000, Serbia
[2] Univ Novi Sad, Fac Tech Sci, Novi Sad 21000, Serbia
关键词
3D scanning; geometrical accuracy; CAD inspection; implants; polymer materials; impression materials; EDENTULOUS PATIENTS; PROSTHESES; SURVIVAL; ABUTMENT; CASTS; BONE;
D O I
10.3390/polym14142821
中图分类号
O63 [高分子化学(高聚物)];
学科分类号
070305 ; 080501 ; 081704 ;
摘要
The paper presents the influence of impression methods, polymer materials, and implant angulation on the accuracy of the definitive working model for the production of implant-supported dental restorations, based on the analysis of results obtained using different impression methods, materials, and parallel and angulated implants. The study findings indicate that all aforementioned factors impact the accuracy of the definitive working model. Specifically, 20 degrees implant angulation in relation to the vertical plane has a greater impact on the impression accuracy compared to parallel implants. The open and splint method in combination with addition silicone, as well as the splint method and polyether combination yielded more accurate results when using implants under 20 degrees angulation compared to other method and material combinations. The splint method in combination with addition silicone resulted in the smallest mean deviations from the center of the parallel implant base compared to other combinations of methods and materials. Analysis results further revealed statistically significant differences in the measured indicators across impression methods, implants, and polymer materials.
引用
收藏
页数:21
相关论文
共 50 条
  • [21] The Influence of Impression Material on the Accuracy of the Master Cast in Implant Restorations
    Stefos, Spyridon
    Kourtis, Stefanos
    Sarafianou, Aspasia
    Zoidis, Panagiotis
    OPEN DENTISTRY JOURNAL, 2018, 12 : 1123 - 1136
  • [22] Influence of impression technique and material on the accuracy of multiple implant impressions
    Woestmann, Bernd
    Rehmann, Peter
    Balkenhol, Markus
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2008, 21 (04) : 299 - 301
  • [23] ACCURACY OF IMPRESSION TECHNIQUES IN IMPLANT DENTISTRY
    Alrumi, Bader Abdullah
    al Yami, Ibrahim Mohammed
    Alghamdi, Randah Abdulrhman
    Aldajani, Ali Mohammed
    Alhelaly, Hussein Mohammed
    Mohammed, Abdulrahman Thabet
    Abotaleb, Mohannd Talebalrahman
    Aljuaid, Mohammed Mbearak
    Alotaibi, Jamal Mousa
    Alelyani, Mohammed Dakhil
    INDO AMERICAN JOURNAL OF PHARMACEUTICAL SCIENCES, 2019, 6 (01): : 1396 - 1402
  • [24] Implant Impression Techniques using Different Materials and Methods: A Review
    Yasar, Muhammed Necati
    Cetinsahin, Cem
    Bayar, Omer
    Ozer, Hasan Yildrim
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2022, 16 (02) : ZE12 - ZE17
  • [25] Influence of implant angulation and clinical implant scan body height on the accuracy of complete arch intraoral digital scans
    Gomez-Polo, Miguel
    Sallorenzo, Alessandro
    Ortega, Rocio
    Gomez-Polo, Cristina
    Barmak, Abdul B.
    Att, Wael
    Revilla-Leon, Marta
    JOURNAL OF PROSTHETIC DENTISTRY, 2024, 131 (01): : 119 - 127
  • [26] Can transfer type and implant angulation affect impression accuracy? A 3D in vitro evaluation
    Farronato, Davide
    Pasini, Pietro Mario
    Campana, Veronica
    Lops, Diego
    Azzi, Lorenzo
    Manfredini, Mattia
    ODONTOLOGY, 2021, 109 (04) : 884 - 894
  • [27] Influence of the implant scan body bevel location, implant angulation and position on intraoral scanning accuracy: An in vitro study
    Gomez-Polo, Miguel
    Alvarez, Francisco
    Ortega, Rocio
    Gomez-Polo, Cristina
    Barmak, Abdul B.
    Kois, John C.
    Revilla-Leon, Marta
    JOURNAL OF DENTISTRY, 2022, 121
  • [28] Can transfer type and implant angulation affect impression accuracy? A 3D in vitro evaluation
    Davide Farronato
    Pietro Mario Pasini
    Veronica Campana
    Diego Lops
    Lorenzo Azzi
    Mattia Manfredini
    Odontology, 2021, 109 : 884 - 894
  • [29] Comparative Evaluation of the Effects of Implant Position, Impression Material, and Tray Type on Implant Impression Accuracy
    Gokcen-Rohlig, Bilge
    Ongul, Deger
    Sancakli, Erkan
    Sermet, Bulent
    IMPLANT DENTISTRY, 2014, 23 (03) : 283 - 288
  • [30] Accuracy of an autonomous dental implant robotic system in dental implant surgery
    Wu, Yu
    Zou, Shiqi
    Lv, Pin
    Wang, Xiao
    JOURNAL OF PROSTHETIC DENTISTRY, 2025, 133 (03): : 764 - 770