Rotational Fluoroscopy Assists in Detection of Intra-Articular Screw Penetration During Volar Plating of the Distal Radius

被引:14
|
作者
Tweet, Matthew L. [1 ]
Calfee, Ryan P. [1 ]
Stern, Peter J. [1 ]
机构
[1] Univ Cincinnati, Coll Med, Dept Orthopaed Surg, Cincinnati, OH USA
来源
基金
美国国家卫生研究院;
关键词
Complication; distal radius; fracture; fluoroscopy; volar plate; INTERNAL-FIXATION; OPEN REDUCTION; FRACTURES;
D O I
10.1016/j.jhsa.2009.12.033
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Purpose Intra-articular screw penetration is one complication of volar plate fixation of distal radius fractures. This study was designed to determine the most commonly used imaging techniques and views during volar plating of distal radius fractures and to evaluate surgeons' ability to detect intra-articular screw placement on static fluoroscopic images and rotational fluoroscopy. Methods Active members of the American Society for Surgery of the Hand were polled regarding preferred imaging techniques (fluoroscopic vs cassette radiographs) and views (rotational fluoroscopy or static orthogonal/anatomic tilt/semipronated imaging) during volar plating of distal radius fractures. After the survey, volar locking plates were applied to 30 cadaveric distal radiuses. We intentionally penetrated a single screw into the radiocarpal joint in half of the specimens (15 arms) and intentionally did not penetrate the radiocarpal joint in the other half. Imaging (standard posteroanterior [PA] and lateral views, 11 degrees tilt PA and 22 tilt lateral views, and two 360 degrees fluoroscopy movies) was performed using a custom jig. Five surgeons blinded to results reviewed randomized image sets evaluating for intra-articular screw placement. Receiver operating characteristic curves were constructed to compare the reliability of each fluoroscopic projection or movie. Results Among 696 survey respondents, 606 exclusively used fluoroscopic imaging (without cassette radiographs) and over 450 preferred either tilt images or rotational fluoroscopy to detect intra-articular screw penetration. In our cadaveric model, rotational fluoroscopy provided the highest sensitivity (93%) and specificity (96%) for the detection of intra-articular screw penetration. Rotational fluoroscopy was significantly more reliable (p<.01) than most images (standard lateral, 11 degrees PA, 22 degrees lateral, paired PA/lateral) and trended strongly toward better reliability for all remaining images (standard PA [p=.07], paired 11 degrees PA/22 degrees lateral [p=.08], and 22 degrees tilt fluoroscopy movie [p=.11]). Conclusions Rotational fluoroscopy improves the surgeon's ability to detect intra-articular screw penetration during volar plating of the distal radius. No combination of imaging allowed detection of all intra-articular screws. A high level of suspicion for intra-articular screw penetration should be maintained during volar plating of distal radial fractures. (J Hand Surg 2010;35A:619-627. Copyright (C) 2010 by the American Society for Surgery of the Hand. All rights reserved.)
引用
收藏
页码:619 / 627
页数:9
相关论文
共 50 条
  • [41] Dorsal screw penetration following implant removal after volar locked plating of distal radius fracture
    Luzian C. P. Haug
    Christian Deml
    Michael Blauth
    Rohit Arora
    Archives of Orthopaedic and Trauma Surgery, 2011, 131 : 1279 - 1282
  • [42] Three Year Review of Dorsal Plating for Complex Intra-Articular Fractures of the Distal Radius
    Spiteri, Michelle
    Ng, Wayne
    Matthews, Jamie
    Power, Dominic
    Brewster, Mark
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2018, 23 (02): : 221 - 226
  • [43] Combined Plating of Intra-Articular Distal Radius Fractures, a Consecutive Series of 74 Cases
    Sagerfors, Marcus
    Lundqvist, Eva
    Bjorling, Patrik
    JOURNAL OF WRIST SURGERY, 2020, 09 (05) : 388 - 395
  • [44] Volar Plating: Imaging Modalities for the Detection of Screw Penetration
    Bergsma, Minke
    Denk, Katharina
    Doornberg, Job N.
    van den Bekerom, Michel P. J.
    Kerkhoffs, Gino M. M. J.
    Jaarsma, Ruurd L.
    Obdeijn, Miryam C.
    JOURNAL OF WRIST SURGERY, 2019, 8 (06) : 520 - 530
  • [45] The Dorsal Tangential X-Ray View to Determine Dorsal Screw Penetration During Volar Plating of Distal Radius Fractures
    Brunner, Alexander
    Siebert, Christin
    Stieger, Claudia
    Kastius, Alexander
    Link, Bjoern-Christian
    Babst, Reto
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2015, 40 (01): : 27 - 33
  • [46] The Keystone Challenge- Volar Displaced Intra-Articular Distal Radius Fractures with Lunate Fossa Fragment
    Cheruvu, M. S.
    Dass, D.
    Ford, D. J.
    Roushdi, I.
    BRITISH JOURNAL OF SURGERY, 2021, 108
  • [47] Dorsal Screw Penetration With the Use of Volar Plating of Distal Radius Fractures: How Can You Best Detect?
    Hill, Brian W.
    Shakir, Irshad
    Cannada, Lisa K.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2015, 29 (10) : e408 - e413
  • [48] Locking Plate for AO Type C Intra-articular Distal Radius Fracture: Volar or Dorsal Approach?
    Ka-Ho, Tsang
    Kai-Yiu, Choi
    Yat-Fai, Chan
    Shun-Chung, Lau
    Yuk-Yin, Chow
    JOURNAL OF ORTHOPAEDICS TRAUMA AND REHABILITATION, 2014, 18 (02) : 83 - 88
  • [49] Corrective Osteotomy with Volar and Dorsal Fixation for Malunion of Intra-Articular Fracture of the Distal Radius: A Retrospective Study
    Zhang, Huihao
    Zhu, Yong
    Fu, Fangda
    Gou, Lingyun
    Zhu, Yonglin
    Zhang, Zhiguo
    Zhou, Chengcong
    Yao, Sai
    Yue, Ming
    Li, Xiaofeng
    Tong, Peijian
    Ruan, Hongfeng
    Wu, Chengliang
    ORTHOPAEDIC SURGERY, 2022, 14 (08) : 1751 - 1758
  • [50] Functional and Radiological Outcomes in Intra-Articular Fractures of Distal Radius with Volar Variable Angle Locking Plates
    Arora, Sahil
    Govil, Vasudha
    Neogi, Avik Kr
    Bishnoi, Sanju
    Paul, Shagnik
    Gupta, R. K.