Topography and longitudinal chromatic aberration characterizations of refractive-diffractive multifocal intraocular lenses

被引:41
|
作者
Loicq, Jerome [1 ]
Willet, Nicolas [2 ]
Gatinel, Damien [3 ]
机构
[1] Univ Liege, STAR Inst, Ctr Spatial Liege, Av Pre Aily, B-4031 Angleur, Belgium
[2] Univ Liege, Dept Chem, Liege, Belgium
[3] Fdn Rothschild Hosp, Paris, France
来源
关键词
EXTENDED DEPTH; PERFORMANCE;
D O I
10.1016/j.jcrs.2019.06.002
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: Most optical systems present chromatic aberration quantified along the optical axis by the longitudinal chromatic aberration (LCA). LCA is controlled by the biomaterial Abbe number combined with diffractive effects, driven by the intraocular lens (IOL) topography. This study experimentally aimed at describing the effect in vitro of LCA in diffractive multifocal IOLs, with the help of dedicated optical benches and topographic characterization. Setting: Centre Spatial de Liege, Belgium. Design: Optical and topology analysis of various multifocal diffractive IOLs. Methods: Seven diffractive multifocal IOLs, available on the market and exhibiting different diffractive profiles, made from various biomaterials, were characterized under different wavelengths. Results: Through-focus modulation transfer function (MTF) curves and IOL diffraction efficiency depends on the incident light wavelength. In this study, the topology properties of various multi-focal IOLs were investigated and their characteristics were correlated to their optical behavior for various wavelengths. Chromatic properties and their origins were then compared. As expected, diffractive and refractive effects were found to act in opposite ways, and could be partially or completely compensated. Conclusions: The LCA of each of the IOLs was evaluated in vitro. In most of the multifocal IOLs studied, some of the foci were found to be refractive, whereas others were diffractive. Although the results were not extrapolated to clinical relevance, it was shown, in some of the cases, that LCA could be fully compensated. (C) 2019 ASCRS and ESCRS
引用
收藏
页码:1650 / 1659
页数:10
相关论文
共 50 条
  • [21] Influence of ocular longitudinal chromatic aberration on the selection of aspheric intraocular lenses
    Hong, Xin
    Choi, Myoung
    OPTICS EXPRESS, 2010, 18 (25): : 26175 - 26183
  • [22] Apodized diffractive versus refractive multifocal intraocular lenses: Optical and visual evaluation
    Zelichowska, Beata
    Rekas, Marek
    Stankiewicz, Andrzej
    Cervino, Alejandro
    Montes-Mico, Robert
    JOURNAL OF CATARACT AND REFRACTIVE SURGERY, 2008, 34 (12): : 2036 - 2042
  • [23] Reading ability and stereoacuity with combined implantation of refractive and diffractive multifocal intraocular lenses
    Chen, Weirong
    Meng, Qianli
    Ye, Haiyun
    Liu, Yizhi
    ACTA OPHTHALMOLOGICA, 2011, 89 (04) : 376 - 381
  • [24] Diffractive-Refractive Intraocular Lenses
    Lenkova, G. A.
    Korolkov, V. P.
    Koronkevich, V. P.
    Nasyrov, R. K.
    Gutman, A. S.
    Iskakov, I. A.
    Treushnikov, V. M.
    OPTOELECTRONICS INSTRUMENTATION AND DATA PROCESSING, 2008, 44 (04)
  • [25] Diffractive-refractive intraocular lenses
    G. A. Lenkova
    V. P. Korolkov
    V. P. Koronkevich
    R. K. Nasyrov
    A. S. Gutman
    I. A. Iskakov
    V. M. Treushnikov
    Optoelectronics, Instrumentation and Data Processing, 2008, 44 (4) : 342 - 352
  • [26] In vitro assessment of longitudinal chromatic aberration in different optical designs of intraocular lenses
    Labuz, Grzegorz
    Hyeck-Soo, Son
    Yildirim, Timur Mert
    Auffarth, Gerd
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2018, 59 (09)
  • [27] Halos in refractive multifocal intraocular lenses
    Pieh, S
    Zöhrer, R
    Backner, B
    Hanselmayer, G
    Ruhswurm, I
    Scholz, U
    Weghaupt, H
    Skorpik, C
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S298 - S298
  • [28] Design and Optimization of Diffractive Multifocal Intraocular Lenses
    Dong Bo
    Yang Ying
    Xue Changxi
    ACTA OPTICA SINICA, 2023, 43 (19)
  • [29] Simulation and measurement of chromatic aberration reduction with diffractive-refractive hybrid lenses for medical applications
    Brezna, W.
    Lux, K.
    Dragostinoff, N.
    Jung, G.
    BIOMEDICAL ENGINEERING-BIOMEDIZINISCHE TECHNIK, 2014, 59 : S954 - S954
  • [30] BIOMETRIC ASPECTS OF DIFFRACTIVE MULTIFOCAL INTRAOCULAR LENSES
    BELLUCCI, R
    ANNALS OF OPHTHALMOLOGY, 1992, 24 (10): : 374 - 377