Assessment of Device-Related Thrombus and Associated Clinical Outcomes With the WATCHMAN Left Atrial Appendage Closure Device for Embolic Protection in Patients With Atrial Fibrillation (from the PROTECT-AF Trial)

被引:96
|
作者
Main, Michael L. [1 ]
Fan, Dali [2 ]
Reddy, Vivek Y. [3 ]
Holmes, David R. [4 ]
Gordon, Nicole T. [5 ]
Coggins, Tina R. [1 ]
House, John A. [1 ]
Liao, Lawrence [6 ]
Rabineau, Dawn [6 ]
Latus, George G. [5 ]
Huber, Kenneth C. [1 ]
Sievert, Horst [7 ]
Wright, Richard F. [8 ]
Doshi, Shephal K. [8 ]
Douglas, Pamela S. [6 ]
机构
[1] St Lukes Mid Amer Heart Inst, Kansas City, MO USA
[2] Univ Calif Davis, Med Ctr, Sacramento, CA 95817 USA
[3] Mt Sinai Sch Med, New York, NY USA
[4] Mayo Clin, Rochester, MN USA
[5] Boston Sci Corp, St Paul, MN USA
[6] Duke Univ, Med Ctr, Durham, NC USA
[7] Cardiovasc Ctr Frankfurt, Frankfurt, Germany
[8] St Johns Hlth Ctr, Pacific Heart Inst & Providence, Santa Monica, CA USA
来源
AMERICAN JOURNAL OF CARDIOLOGY | 2016年 / 117卷 / 07期
关键词
SPONTANEOUS ECHO CONTRAST; AMPLATZER CARDIAC PLUG; THROMBOEMBOLIC RISK; AMERICAN-SOCIETY; ECHOCARDIOGRAPHY; STROKE; STANDARDS; VELOCITY; WARFARIN; SYSTEM;
D O I
10.1016/j.amjcard.2016.01.039
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Left atrial appendage closure with the WATCHMAN device is an alternative to anticoagulation for stroke prevention in selected patients with atrial fibrillation (AF). LA device-related thrombus (DRT) is poorly defined and understood. We aimed to (1) develop consensus echocardiographic diagnostic criteria for DRT; (2) estimate the incidence of DRT; and (3) determine clinical event rates in patients with DRT. In phase 1 (training), a training manual was developed and reviewed by 3 echocardiographers with left atrial appendage closure device experience. All available transesophageal (TEE) studies in the WATCHMAN left atrial appendage system for embolic protection in patients with atrial fibrillation (PROTECT-AF) trial patients with suspected DRT were reviewed in 2 subsequent phases. In phase 2 (primary blind read), each reviewer independently scored each study for DRT, and final echo criteria were developed. Unanimously scored studies were considered adjudicated, whereas all others were reevaluated by all reviewers in phase 3 (group adjudication read). DRT was suspected in 35 of 485 patients by the site investigator, the echocardiography core laboratory, or both; 93 of the individual TEE studies were available for review. In phase 2, 3 readers agreed on 67 (72%) of time points. Based on phases 1 and 2, 5 DRT criteria were developed. In phase 3, studies without agreement in phase 2 were adjudicated using these Criteria. Overall, at least 1 TEE was DRT positive in 27 (5.7%) PROTECT-AF patients. Stroke, peripheral embolism, or cardiac/unexplained death occurred in subjects with DRT at a rate of 3.4 per 100 patient-years follow-up. In conclusion, DRT were identified on at least 1 TEE in 27 PROTECT-AF patients, indicating a DRT incidence of 5.7%. Primary efficacy events in patients with DRT occurred at a rate of 3.4 per 100 patient-years follow-up, intermediate in frequency between event rates previously reported for the overall device and warfarin arms in PROTECT-AF. (C) 2016 Elsevier Inc. All rights reserved.
引用
收藏
页码:1127 / 1134
页数:8
相关论文
共 50 条
  • [11] Device-Related Thrombus After Left Atrial Appendage Closure: Incidence, Predictors, and Outcomes
    Dukkipati, Srinivas R.
    Kar, Saibal
    Holmes, David R.
    Doshi, Shephal K.
    Swarup, Vijendra
    Gibson, Douglas N.
    Maini, Brijeshwar
    Gordon, Nicole T.
    Main, Michael L.
    Reddy, Vivek Y.
    CIRCULATION, 2018, 138 (09) : 874 - 885
  • [12] Percutaneous Left Atrial Appendage Closure for Stroke Prophylaxis in Patients With Atrial Fibrillation 2.3-Year Follow-up of the PROTECT AF (Watchman Left Atrial Appendage System for Embolic Protection in Patients With Atrial Fibrillation) Trial
    Reddy, Vivek Y.
    Doshi, Shephal K.
    Sievert, Horst
    Buchbinder, Maurice
    Neuzil, Petr
    Huber, Kenneth
    Halperin, Jonathan L.
    Holmes, David
    CIRCULATION, 2013, 127 (06) : 720 - 729
  • [13] Safety of Amulet Left Atrial Appendage Occluder and Watchman Device for Left Atrial Appendage Closure in Patients With Atrial Fibrillation
    Sawaya, Fadi
    Abi-Saleh, Bernard
    Hoteit, Abbas
    Jdaidany, Jennifer
    Moumneh, Mohamad B.
    Harbieh, Bernard
    Khoury, Maurice
    Aramouni, Salim
    Abdulhai, Farah
    Refaat, Marwan
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2024, 16 (03)
  • [14] REPLY: Device-Related Thrombus After Percutaneous Left Atrial Appendage Occlusion in Patients With Atrial Fibrillation
    Fauchier, Laurent
    Cinaud, Alexandre
    Brigadeau, Francois
    Lepillier, Antoine
    Defaye, Pascal
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2018, 72 (04) : 474 - 475
  • [15] Recurrence of Device-Related Thrombus After Percutaneous Left Atrial Appendage Closure
    Asmarats, Lluis
    Cruz-Gonzalez, Ignacio
    Nombela-Franco, Luis
    Arzamendi, Dabit
    Peral, Vicente
    Nietlispach, Fabian
    Latib, Azeem
    Maffeo, Diego
    Gonzalez-Ferreiro, Rocio
    Rodriguez-Gabella, Tania
    Agudelo, Victor
    Alamar, Marta
    Ghenzi, Raffael A.
    Mangieri, Antonio
    Bernier, Mathieu
    Rodes-Cabau, Josep
    CIRCULATION, 2019, 140 (17) : 1441 - 1443
  • [16] Percutaneous left atrial appendage closure using the LAmbre device in patients with atrial fibrillation and left atrial appendage thrombus
    Jin, Lu-Shen
    Ke, Jin-Yan
    Lin, Yuan-Nan
    Li, Ling
    Fu, Jia-Yang
    Chen, Yi-Lian
    Qiu, Yi-Xuan
    Li, Xiao-Wei
    Pan, Yang-Qi
    Li, Yue-Chun
    FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
  • [17] Incidence, predictors and outcomes of device-related thrombus after left atrial appendage closure with the WATCHMAN device-Insights from the EWOLUTION real world registry
    Sedaghat, Alexander
    Nickenig, Georg
    Schrickel, Jan Wilko
    Ince, Huseyin
    Schmidt, Boris
    Protopopov, Alexey V.
    Betts, Timothy R.
    Gori, Tommaso
    Sievert, Horst
    Mazzone, Patrizio
    Grygier, Marek
    Wald, Christof
    Vireca, Elisa
    Allocco, Dominic
    Boersma, Lucas V. A.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2021, 97 (07) : E1019 - E1024
  • [18] Left Atrial Appendage Closure With the Use of a WATCHMAN FLX Device in a Patient with Persistent Left Atrial Appendage Thrombus
    Torres-Saura, Francisco
    Cruz-Gonzalez, Ignacio
    Perez-Berbel, Patricio
    Raul Centurion-Inda, Ernesto
    Farid Dau-Villarreal, Derek
    Romero-Vazquianez, Manuela
    Arroyo-Ucar, Eduardo
    CANADIAN JOURNAL OF CARDIOLOGY, 2020, 36 (06) : 968.e5 - 968.e7
  • [19] Predictors of Left Atrial Appendage Occlusion Device-Related Thrombus
    Morgan, Ethan
    Baghdadi, Sarah
    Smith, Ross
    Tabak, Carine
    Parikh, Riya
    Bajaj, Matthew
    Enders, Robert
    Uhlich, Cody
    Baer, Jacob
    Jiwani, Sania
    Harvey, Christopher
    Gunasekaran, Prasad
    CIRCULATION, 2023, 148
  • [20] Left atrial appendage closure with the watchman device reduces atrial fibrillation management costs
    D'Ancona, Giuseppe
    Safak, Erdal
    Weber, Denise
    Arslan, Fatih
    Kische, Stephan
    Darius, Harald
    Behrens, Steffen
    Zohlenhoefer-Momm, Dietlind
    Ortak, Jasmin
    Kugler, Joachim
    Ince, Huseyin
    CLINICAL RESEARCH IN CARDIOLOGY, 2022, 111 (01) : 105 - 113