Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic

被引:6
|
作者
Moiz, Bushra [1 ]
Rashid, Anila [1 ]
Hasan, Muhammad [1 ]
Jafri, Lena [1 ]
Raheem, Ahmed [1 ]
机构
[1] Aga Khan Univ Hosp, Dept Pathol & Lab Med, Karachi, Pakistan
关键词
international normalized ratio; point-of-care testing device; anticoagulation; warfarin; SELF-MANAGEMENT; COAGUCHEK XS; ACCURACY; RELIABILITY; EMERGENCY;
D O I
10.1177/1076029617752247
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients' self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements between CoaguChek XS Pro and Sysmex CS2000i was excellent (correlation coefficient = 0.973). The overall mean difference was 0.21 INR +/- 0.32 (range: 1.7-0.44). The mean difference was found to get increased as INR results increased and was 0.09 in the subtherapeutic range (1.9 INR), 0.29 INR in the therapeutic range (2.0-3.0 INR), while 0.4 INR in the supratherapeutic range (>3.0 INR). The overall agreement was excellent ( = 0.916) and overall 11 (5.5%) of 200 INR measurements showed a difference in dosing decision between the 2 instruments. The positive bias of POC-INR is evident in the supratherapeutic range which could affect the dosing decision requiring confirmation with the laboratory INR measurement.
引用
收藏
页码:1153 / 1158
页数:6
相关论文
共 50 条
  • [31] Measuring international normalized ratios in long-term care: A comparison of commercial laboratory and point-of-care device results
    Bluestein, Daniel
    Brantley, Cynthia
    Barnes-Eley, Myra
    Gravenstein, Stefan
    Basta, Sameh
    JOURNAL OF THE AMERICAN MEDICAL DIRECTORS ASSOCIATION, 2007, 8 (06) : 404 - 408
  • [32] Monitoring oral anticoagulant therapy with point-of-care devices: Correlations and caveats.
    Gosselin, RC
    Owings, JT
    Larkin, F
    White, RH
    Hutchinson, R
    Branch, J
    CLINICAL CHEMISTRY, 1997, 43 (09) : 1785 - 1786
  • [33] International normalized ratio testing with a point-of-care coagulometer in 1-month-old infants: A comparison with Normotest
    Iijima, Shigeo
    Baba, Toru
    Ueno, Daizo
    Ohishi, Akira
    THROMBOSIS RESEARCH, 2016, 145 : 72 - 77
  • [34] Prospective comparison of point-of-care international normalised ratio measurement versus plasma international normalised ratio for acute traumatic coagulopathy
    Mitra, Biswadev
    O'Reilly, Gerard
    Collecutt, Margaret
    Cameron, Peter A.
    Phillips, Louise
    Davis, Amanda
    EMERGENCY MEDICINE AUSTRALASIA, 2012, 24 (04) : 363 - 368
  • [35] ADOPTION AND VALIDATION OF THE INTERNATIONAL NORMALIZED RATIO FOR MONITORING ORAL ANTICOAGULANT-THERAPY - THE SITUATION IN 1989
    VANDENBESSELAAR, AMHP
    RESEARCH IN CLINIC AND LABORATORY, 1990, 20 (01): : 75 - 81
  • [36] Point-of-Care International Normalized Ratio Testing Accelerates Thrombolysis in Patients With Acute Ischemic Stroke Using Oral Anticoagulants
    Rizos, Timolaos
    Herweh, Christian
    Jenetzky, Ekkehart
    Lichy, Christoph
    Ringleb, Peter Arthur
    Hacke, Werner
    Veltkamp, Roland
    STROKE, 2009, 40 (11) : 3547 - 3551
  • [37] Effect of dabigatran treatment at therapeutic levels on point-of-care international normalized ratio (INR)
    Lokkebo, Elisabeth Sellevoll
    Kristoffersen, Ann Helen
    Stavelin, Anne
    Solvik, Una Orvim
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2017, 55 (07) : E143 - E145
  • [38] External quality assessment of point-of-care International Normalized Ratio (INR) testing in Europe
    Stavelin, Anne
    Meijer, Piet
    Kitchen, Dianne
    Sandberg, Sverre
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2012, 50 (01) : 81 - 88
  • [39] Establishing an external quality assessment scheme for point-of-care international normalized ratio in Thailand
    Tientadakul, P.
    Chuntarut, A.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2015, 37 (04) : 509 - 514
  • [40] Quality assessment of CoaguChek point-of-care international normalized ratio monitors: A note of caution
    Kitchen, Steve
    Kitchen, Dianne
    Jennings, Ian
    Woods, Tim
    Walker, Isobel
    CLINICAL CHEMISTRY, 2007, 53 (08) : 1555 - 1556