Prospective Comparison of Point-of-Care Device and Standard Analyzer for Monitoring of International Normalized Ratio in Outpatient Oral Anticoagulant Clinic

被引:6
|
作者
Moiz, Bushra [1 ]
Rashid, Anila [1 ]
Hasan, Muhammad [1 ]
Jafri, Lena [1 ]
Raheem, Ahmed [1 ]
机构
[1] Aga Khan Univ Hosp, Dept Pathol & Lab Med, Karachi, Pakistan
关键词
international normalized ratio; point-of-care testing device; anticoagulation; warfarin; SELF-MANAGEMENT; COAGUCHEK XS; ACCURACY; RELIABILITY; EMERGENCY;
D O I
10.1177/1076029617752247
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Point-of-care testing (POCT) coagulometers are increasingly being used in the hospital setting and patients' self-testing. We determined the agreement of prothrombin time international normalized ratio (INR) results by POCT coagulometer and laboratory instrument through a comparative analysis and investigated whether the results of POCT coagulometer can reliably be used without being confirmed by standard laboratory analyzer. A total of 200 INR measurements by POCT coagulometer (CoaguChek XS Pro) and laboratory analyzer (Sysmex CS2000i) were compared using Passing-Bablok regression analysis and Bland-Altman plot. Agreement of the INR measurement was further analyzed in relation to dosing decision. The correlation of INR measurements between CoaguChek XS Pro and Sysmex CS2000i was excellent (correlation coefficient = 0.973). The overall mean difference was 0.21 INR +/- 0.32 (range: 1.7-0.44). The mean difference was found to get increased as INR results increased and was 0.09 in the subtherapeutic range (1.9 INR), 0.29 INR in the therapeutic range (2.0-3.0 INR), while 0.4 INR in the supratherapeutic range (>3.0 INR). The overall agreement was excellent ( = 0.916) and overall 11 (5.5%) of 200 INR measurements showed a difference in dosing decision between the 2 instruments. The positive bias of POC-INR is evident in the supratherapeutic range which could affect the dosing decision requiring confirmation with the laboratory INR measurement.
引用
收藏
页码:1153 / 1158
页数:6
相关论文
共 50 条
  • [1] Verification of the qLabs international normalized ratio point-of-care device for monitoring of patients attending an anticoagulation clinic
    Mbokota, N.
    Schapkaitz, E.
    Louw, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 (05) : 508 - 514
  • [2] Clinically significant differences between point-of-care analysers and a standard analyser for monitoring the International Normalized Ratio in oral anticoagulant therapy: a multi-instrument evaluation in a hospital outpatient setting
    Moore, Gary W.
    Henley, Alexis
    Cotton, Shawn S.
    Tugnait, Sanjiv
    Rangarajan, Savita
    BLOOD COAGULATION & FIBRINOLYSIS, 2007, 18 (03) : 287 - 292
  • [3] Reliability of Point-of-Care International Normalized Ratio Testing in an Academic Family Medicine Clinic
    Christenberry, Emily J.
    Padilla, Margie E.
    Aguirre, Michelle
    Loya, Amanda M.
    Aragon, Lorenzo
    POINT OF CARE, 2018, 17 (02): : 55 - 58
  • [4] Bias and uncertainty of the International Normalized Ratio determined with a whole blood point-of-care prothrombin time test device by comparison to a new International Standard for thromboplastin
    van den Besselaar, Antonius M. H. P.
    van Rijn, Claudia J. J.
    Abdoel, Charmane F.
    Cobbaert, Christa M.
    THROMBOSIS RESEARCH, 2021, 202 : 1 - 7
  • [5] Comparison of two point-of-care international normalized ratio devices and laboratory method
    Vazquez, Sara R.
    Fleming, Ryan P.
    Johnson, Stacy A.
    BLOOD COAGULATION & FIBRINOLYSIS, 2017, 28 (07) : 534 - 539
  • [6] A Comparison of International Normalized Ratio Results by Point-of-Care Device and Clinical Laboratory Analyzers in a Vascular Surgery Department
    Alserr, Ayman H. K.
    Menshawey, Rahma
    Kotb, Ahmed
    Hussein, Nourhan
    Kotp, Noha
    Ashraf-Taha, Mahmoud
    Anwar, Nouran
    Abdalla, Asmaa
    Abdullah, Menna
    Ela, Shaimaa Abol
    Khairy, Hussein
    POINT OF CARE, 2020, 19 (04): : 106 - 111
  • [7] Point-of-care versus laboratory measurement of the International Normalized Ratio
    Pierce, MT
    Crain, L
    Smith, J
    Mehta, V
    AMERICAN JOURNAL OF HEALTH-SYSTEM PHARMACY, 2000, 57 (24) : 2271 - 2274
  • [8] Validation of the international normalized ratio (INR) in a new point-of-care system designed for home monitoring of oral anticoagulation therapy
    Plesch, W.
    van den Besselaar, A. M. H. P.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2009, 31 (01) : 20 - 25
  • [9] Importance of device evaluation for point-of-care prothrombin time international normalized ratio testing programs
    McBane, RD
    Felty, CL
    Hartgers, ML
    Chaudhry, R
    Beyer, LK
    Santrach, PJ
    MAYO CLINIC PROCEEDINGS, 2005, 80 (02) : 181 - 186
  • [10] A comparison study of point of care testing versus conventional international normalised ratio monitoring at a medical outpatient clinic
    Ng, Seng Loong
    Mohamed, Mohd Sapawi
    Ibrahim, Suhairi
    Mustaffa, Nazri
    Draman, Nani
    Yusof, Zurkurnai
    CIRCULATION, 2012, 125 (19) : E898 - E899