3-d view weighted cone-beam filtered backprojection reconstruction for digital tomosynthesis

被引:6
|
作者
Li, Baojun [1 ]
Avinash, Gopal [1 ]
Claus, Bernhard [2 ]
Metz, Stephen [3 ]
机构
[1] Gen Elect Healthcare, Appl Sci Lab, Waukesha, WI 53188 USA
[2] Gen Elect, Global Res Ctr, Niskayuna, NY 12309 USA
[3] Gen Elect Healthcare, Diagnost Xray Engn, Waukesha, WI 53188 USA
关键词
tomosynthesis; filtered backprojection; view weighting; cone beam; artifact;
D O I
10.1117/12.708084
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Cone-beam filtered backprojection (CB-FBP) is one of the major reconstruction algorithms for digital tomosynthesis. In conventional FBP, the photon fluxes in projections are evenly distributed along the X-ray beam. Due to the limited view angles and finite detector dimensions, this uniform weighting causes non-uniformity in the recon images and leads to cone-beam artifact. In this paper, we propose a 3-D view weighting technique in combination with FBP to combat this artifact. An anthropomorphic chest phantom was placed at supine position to enable the imaging of chest PA view. During a linear sweep of X-ray source, 41 X-ray images at different projection angles were acquired with the following protocol: 120kVp, 160mA, and 0.64mAs/exposure. To create the worst scenario for testing, we chose 60 degrees as the sweep angle in this exam. The data set was reconstructed with conventional CB-FBP and proposed algorithm under the same parameters: FOV = 40x40 cm boolean AND 2, and slice thickness = 4mm. 3 recon slices were randomly selected for review with slice height = 10.5/14.5/17.5cm. Results were assessed qualitatively by human observers and quantitatively through ROI measurement. In each slice, three pre-defined ROIs (50x50 pixels)--ROI A and B are in artifact more pronounced area, and ROI C is in relatively artifact-free area-are extracted and measured. The non-uniformity error was defined as the ratio of MEAN(AVG(C-A), AVG(C-B)) / AVG(C). The average non-uniformity error over the three test images was 0.428 for without view weighting and only 0.041 for with view weighting.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] Cone-beam reconstruction using the backprojection of locally filtered projections
    Pack, JD
    Noo, F
    Clackdoyle, R
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (01) : 70 - 85
  • [2] Regularized iterative weighted filtered backprojection for helical cone-beam CT
    Sunnegardh, Johan
    Danielsson, Per-Erik
    [J]. MEDICAL PHYSICS, 2008, 35 (09) : 4173 - 4185
  • [3] FILTERED BACKPROJECTION RECONSTRUCTION OF COMBINED PARALLEL BEAM AND CONE-BEAM SPECT DATA
    DEFRISE, M
    CLACK, R
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (09): : 1517 - 1537
  • [4] Nonlinear Regularization of Iterative Weighted Filtered Backprojection for Helical Cone-Beam CT
    Sunnegardh, Johan
    Grasruck, Michael
    [J]. 2008 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (2008 NSS/MIC), VOLS 1-9, 2009, : 4356 - +
  • [5] Generalized Filtered Backprojection for Digital Breast Tomosynthesis Reconstruction
    Erhard, Klaus
    Grass, Michael
    Hitziger, Sebastian
    Iske, Armin
    Nielsen, Tim
    [J]. MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [6] Cone-beam reconstruction by backprojection and filtering
    Bronnikov, AV
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2000, 17 (11) : 1993 - 2000
  • [7] Cone-beam filtered backprojection image reconstruction using a factorized weighting function
    Chen, Guang-Hong
    Zhuang, Ting-Liang
    Leng, Shuai
    Nett, Brian E.
    [J]. OPTICAL ENGINEERING, 2007, 46 (08)
  • [8] WEIGHTED BACKPROJECTION APPROACH TO CONE-BEAM 3D PROJECTION RECONSTRUCTION FOR TRUNCATED SPHERICAL DETECTION GEOMETRY
    CHO, ZH
    WU, EX
    HILAL, SK
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 1994, 13 (01) : 110 - 122
  • [9] 3D filtered backprojection for curved detector axial cone-beam data on a PlayStation® 3
    Benson, Thomas M.
    Dixon, Walter
    Basu, Samit
    [J]. 2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 3522 - 3525
  • [10] Image reconstruction on PI-lines by use of filtered backprojection in helical cone-beam CT
    Zou, Y
    Pan, XC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (12): : 2717 - 2731