FILTERED BACKPROJECTION RECONSTRUCTION OF COMBINED PARALLEL BEAM AND CONE-BEAM SPECT DATA

被引:22
|
作者
DEFRISE, M [1 ]
CLACK, R [1 ]
机构
[1] UNIV UTAH, DEPT RADIOL, SALT LAKE CITY, UT 84132 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 1995年 / 40卷 / 09期
关键词
D O I
10.1088/0031-9155/40/9/010
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The reconstruction problem for a combined parallel beam (PB) and cone beam (CB) imaging geometry has been addressed. The general algorithm (CB-FBP) of Defrise and Clack has been applied to this geometry and shown to provide accurate images as expected. A second algorithm specifically tailored to the PB-CB geometry was developed. It uses the general principles of the CB-FBP method to combine a shift-variant filtering of the PB data with a standard reconstruction of the CB data using the algorithm of Feldkamp, Davis and Kress (FDK). This 'mixed' algorithm has the advantage of fewer interpolation steps, thereby reducing reconstruction time and providing more accurate reconstructions. The algorithms were applied to noiseless data from a computer-generated 3D Shepp phantom. Both the CB-FBP algorithm and the mixed algorithm successfully combine the CB and the PB data to correct the well known artefacts observed when reconstructing CB data acquired with a circular orbit. The mixed algorithm is about twice as fast as CB-FBP and results in better image quality, due to decreased discretization errors. However, the two algorithms yield comparable image quality when applied to a disc phantom measured with a two-headed SPECT system.
引用
收藏
页码:1517 / 1537
页数:21
相关论文
共 50 条
  • [1] Cone-beam reconstruction using the backprojection of locally filtered projections
    Pack, JD
    Noo, F
    Clackdoyle, R
    [J]. IEEE TRANSACTIONS ON MEDICAL IMAGING, 2005, 24 (01) : 70 - 85
  • [2] SHIFT-VARIANT FILTERED-BACKPROJECTION FOR A DUAL CONE-BEAM AND PARALLEL-BEAM COLLIMATOR SPECT CAMERA
    DEFRISE, M
    CLACK, R
    CHRISTIAN, PE
    GULLBERG, GT
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1994, 35 (05) : P5 - P5
  • [3] SPECT RECONSTRUCTION OF COMBINED CONE-BEAM AND PARALLEL HOLE COLLIMATION WITH EXPERIMENTAL-DATA
    LI, JY
    JASZCZAK, RJ
    TURKINGTON, TG
    GREER, KL
    COLEMAN, RE
    [J]. IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1993, 40 (03) : 300 - 306
  • [4] Combined cone-beam and parallel-beam approach to SPECT
    Krol, A
    Salgado, RB
    Feiglin, DH
    Gangal, KR
    Hardikar, AS
    Thomas, FD
    Lipson, ED
    Bordikar, A
    Coman, IL
    [J]. 2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1032 - 1034
  • [5] Cone-beam reconstruction by backprojection and filtering
    Bronnikov, AV
    [J]. JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 2000, 17 (11) : 1993 - 2000
  • [6] Cone-beam filtered-backprojection algorithm for truncated helical data
    Kudo, H
    Noo, F
    Defrise, M
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1998, 43 (10): : 2885 - 2909
  • [7] Filtered backprojection algorithms for attenuated parallel and cone-beam projections sampled on a sphere
    Weng, Y
    Zeng, GL
    Gullberg, GT
    [J]. THREE-DIMENSIONAL IMAGE RECONSTRUCTION IN RADIOLOGY AND NUCLEAR MEDICINE, 1996, 4 : 19 - 34
  • [8] Cone-beam filtered backprojection image reconstruction using a factorized weighting function
    Chen, Guang-Hong
    Zhuang, Ting-Liang
    Leng, Shuai
    Nett, Brian E.
    [J]. OPTICAL ENGINEERING, 2007, 46 (08)
  • [9] A parallel backprojection algorithm for spiral cone-beam ct exact reconstruction
    Xue, Guowei
    Zhang, Yan
    [J]. International Journal of Digital Content Technology and its Applications, 2010, 4 (06) : 134 - 143
  • [10] 3-DIMENSIONAL SPECT RECONSTRUCTION OF COMBINED CONE BEAM AND PARALLEL BEAM DATA
    JASZCZAK, RJ
    LI, JY
    WANG, H
    COLEMAN, RE
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1992, 37 (03): : 535 - 548