Free-breathing high-pitch 80 kVp dual-source computed tomography of the pediatric chest: Image quality, presence of motion artifacts and radiation dose

被引:22
|
作者
Bodelle, Boris [1 ]
Fischbach, Constanze [1 ]
Booz, Christian [1 ]
Yel, Ibrahim [1 ]
Frellesen, Claudia [1 ]
Beeres, Martin [1 ]
Vogl, Thomas J. [1 ]
Scholtz, Jan-Erik [1 ]
机构
[1] Goethe Univ Frankfurt, Dept Diagnost & Intervent Radiol, Theodor Stern Kai 7, D-60590 Frankfurt, Germany
关键词
High pitch CT; Third generation dual-source CT; Free breathing; Pediatric CT; Chest CT; FILTERED BACK-PROJECTION; AFFIRMED ITERATIVE RECONSTRUCTION; ADVERSE EVENTS; CT; SEDATION; CHILDREN; CANCER; RISKS;
D O I
10.1016/j.ejrad.2017.01.027
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objectives: To investigate image quality, presence of motion artifacts and effects on radiation dose of 80 kVp high-pitch dual-source CT (DSCT) in combination with an advanced modeled iterative reconstruction algorithm (ADMIRE) of the pediatric chest compared to single-source CT (SSCT). Materials and methods: The study was approved by the institutional review board. Eighty-seven consecutive pediatric patients (mean age 9.1 +/- 4.9 years) received either free-breathing high-pitch (pitch 3.2) chest 192-slice DSCT (group 1, n = 31) or standard-pitch (pitch 1.2) 128-slice SSCT (group 2, n = 56) with breathing-instructions by random assignment. Tube settings were similar in both groups with 80 kVp and 74 ref. mAs. Images were reconstructed using FBP for both groups. Additionally, ADMIRE was used in group 1. Effective thorax diameter, image noise, and signal-to-noise ratio (SNR) of the pectoralis major muscle and the thoracic aorta were calculated. Motion artifacts were measured as doubling boarders of the diaphragm and the heart. Images were rated by two blinded readers for overall image quality and presence of motion artifacts on 5-point-scales. Size specific dose estimates (SSDE, mGy) and effective dose (ED, mSv) were calculated. Results: Age and effective thorax diameter showed no statistically significant differences in both groups. Image noise and SNR were comparable (p > 0.64) for SSCT and DSCT with ADMIRE, while DSCT with FBP showed inferior results (p < 0.01). Motion artifacts were reduced significantly (p = 0.001) with DSCT. DSCT with ADMIRE showed the highest overall IQ (p < 0.0001). Radiation dose was lower for DSCT compared to SSCT (median SSDE: 0.82 mGy vs. 0.92 mGy, p < 0.02; median ED: 0.4 mSv vs. 0.48 mSv, p = 0.02). Conclusions: High-pitch 80 kVp chest DSCT in combination with ADMIRE reduces motion artifacts and increases image quality while lowering radiation exposure in free-breathing pediatric patients without sedation. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:208 / 214
页数:7
相关论文
共 50 条
  • [21] The ascending aortic image quality and the whole aortic radiation dose of high-pitch dual-source CT angiography
    Liu, Ying
    Xu, Jian
    Li, Jian
    Ren, Jing
    Liu, Hongtao
    Xu, Junqing
    Wei, Mengqi
    Hao, Yuewen
    Zheng, Minwen
    JOURNAL OF CARDIOTHORACIC SURGERY, 2013, 8
  • [22] Effect of High-Pitch Dual-Source CT to Compensate Motion Artifacts: A Phantom Study
    Farshad-Amacker, Nadia A.
    Alkadhi, Hatem
    Leschka, Sebastian
    Frauenfelder, Thomas
    ACADEMIC RADIOLOGY, 2013, 20 (10) : 1234 - 1239
  • [23] Image Quality of ECG-Triggered High-Pitch, Dual-Source Computed Tomography Angiography for Cardiovascular Assessment in Children
    Barrera, Christian A.
    Otero, Hansel J.
    White, Ammie M.
    Saul, David
    Biko, David M.
    CURRENT PROBLEMS IN DIAGNOSTIC RADIOLOGY, 2020, 49 (01) : 23 - 28
  • [24] Radiation doses from phantom measurements at high-pitch dual-source computed tomography coronary angiography
    Goetti, Robert
    Leschka, Sebastian
    Boschung, Markus
    Mayer, Sabine
    Wyss, Christophe
    Stolzmann, Paul
    Frauenfelder, Thomas
    EUROPEAN JOURNAL OF RADIOLOGY, 2012, 81 (04) : 773 - 779
  • [25] High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
    Korn, A.
    Fenchel, M.
    Bender, B.
    Danz, S.
    Thomas, C.
    Ketelsen, D.
    Claussen, C. D.
    Moonis, G.
    Krauss, B.
    Heuschmid, M.
    Ernemann, U.
    Brodoefel, H.
    NEURORADIOLOGY, 2013, 55 (04) : 423 - 430
  • [26] High-pitch dual-source CT angiography of supra-aortic arteries: assessment of image quality and radiation dose
    A. Korn
    M. Fenchel
    B. Bender
    S. Danz
    C. Thomas
    D. Ketelsen
    C. D. Claussen
    G. Moonis
    B. Krauss
    M. Heuschmid
    U. Ernemann
    H. Brodoefel
    Neuroradiology, 2013, 55 : 423 - 430
  • [27] High-pitch Helical Dual-source Computed Tomographic Pulmonary Angiography Comparing Image Quality in Inspiratory Breath-hold and During Free Breathing
    Ajlan, Amr M.
    Binzaqr, Salma
    Jadkarim, Dalia A.
    Jamjoom, Lamia G.
    Leipsic, Jonathon
    JOURNAL OF THORACIC IMAGING, 2016, 31 (01) : 56 - 62
  • [28] Computed tomography pulmonary angiography using high-pitch dual-source scanner technology
    Al Hassan, Donya A.
    Waheed, Khawaja B.
    El Sirafy, Mohamed N.
    Khattab, Mohamad A.
    Al-Hammadi, Hassan I.
    Ibrahim, Motassem F.
    Ibrahim, Motassem F.
    Arulanantham, Zechariah J.
    SAUDI MEDICAL JOURNAL, 2019, 40 (03) : 230 - 237
  • [29] Radiation Dose and Image Quality of Prospective Triggering With Dual-Source Cardiac Computed Tomography
    Blankstein, Ron
    Shah, Amar
    Pale, Rodrigo
    Abbara, Suhny
    Bezerra, Hiram
    Bolen, Michael
    Mamuya, Wilfred S.
    Hoffmann, Udo
    Brady, Thomas J.
    Cury, Ricardo C.
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 103 (08): : 1168 - 1173
  • [30] Image quality and radiation dose of lower extremity CT angiography at 70 kVp on an integrated circuit detector dual-source computed tomography
    Qi, Li
    Zhao, Yan'E
    Zhou, Chang Sheng
    Spearman, James V.
    Renker, Matthias
    Schoepf, U. Joseph
    Zhang, Long Jiang
    Lu, Guang Ming
    ACTA RADIOLOGICA, 2015, 56 (06) : 659 - 665