Monitoring of heparin therapy beyond the anti-Xa activity assay: Evaluation of a thrombin generation assay

被引:2
|
作者
Vermeiren, Pauline [1 ]
Vandevelde, Arne [1 ,2 ]
Peperstraete, Harlinde [3 ]
Devreese, Katrien M. J. [1 ,2 ]
机构
[1] Ghent Univ Hosp, Coagulat Lab, Dept Lab Med, Ghent, Belgium
[2] Univ Ghent, Dept Diagnost Sci, Ghent, Belgium
[3] Ghent Univ Hosp, Dept Intens Care Med, Ghent, Belgium
关键词
blood coagulation tests; heparin; reference values; thrombin; MOLECULAR-WEIGHT HEPARIN; UNFRACTIONATED HEPARIN; ANTICOAGULANT-THERAPY; VARIABILITY; PROTEINS; BINDING;
D O I
10.1111/ijlh.13836
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Global coagulation assays may be of added value to the anti-Xa assay for monitoring heparin therapy. Unlike most testing methods, the thrombin generation assay ( TGA) has the ability to assess the overall function of the hemostatic system, which provides information on the anticoagulation status of patients. We compared the TGA, measured with ST Genesia (R) STG-DrugScreen (R) reagent, with the anti-Xa assay for monitoring heparin therapy in inflammatory and non-inflammatory patients. We also determined reference values for STG-DrugScreen (R) thrombin generation (TG) parameters. Methods: Reference values were determined on 120 healthy donors. Furthermore, a spiking experiment with unfractionated heparin (UFH) and low molecular weight heparin (LMWH) was performed, and samples of patients receiving UFH or LMWH were analyzed with ST Genesia (R) and the anti-Xa assay. Results: High discrepancy between TG parameters and anti-Xa activity was observed for low LMWH anti-Xa levels. TG parameters were affected in 36/46 (time to peak) to 42/46 (peak height) patients during UFH therapy with sub-target anti-Xa activity levels. Conclusion: TGA seems insufficiently sensitive for low concentrations of LMWH. There may be an added value of the TGA for monitoring UFH in so-called heparin-resistant patients. Therefore, the TGA has the potential to be introduced as an additional tool for monitoring heparin therapy.
引用
收藏
页码:785 / 795
页数:11
相关论文
共 50 条
  • [41] Anti-Xa Assay Monitoring Improves the Precision of Anticoagulation in Venovenous Extracorporeal Membrane Oxygenation
    Hla, Teddy Tun Win
    Christou, Silvana
    Sanderson, Barnaby
    Hanks, Fraser
    Cameron, Lynda
    Camporota, Luigi
    Doyle, Andrew J.
    Retter, Andrew
    ASAIO JOURNAL, 2024, 70 (04) : 313 - 320
  • [42] Establishing the heparin therapeutic range using aPTT and anti-Xa measurements for monitoring unfractionated heparin therapy
    Byun, Jung-Hyun
    Jang, In-Seok
    Kim, Jong Woo
    Koh, Eun-Ha
    BLOOD RESEARCH, 2016, 51 (03) : 171 - 174
  • [43] Comparison of Commercial Low Molecular Weight Heparin and Homemade Anti-Xa Calibrators to a Commercial Specific Anti-Xa Calibrator for Plasma Rivaroxaban Quantification by Anti-Xa Oral Anticoagulant Plasma Concentration Chromogenic Assay
    Divsalar, Bita
    Kalantari, Tahereh
    Mohebbi, Soheila
    Bahmanimehr, Ardeshir
    Shahsavani, Amin
    Borhani-Haghighi, Afshin
    LABORATORY MEDICINE, 2023, 54 (04) : 400 - 405
  • [44] ANTI-XA MONITORING FOR HEPARIN DOSING IN PATIENTS WITH LIVER CIRRHOSIS
    Fuentes, Amaris
    Hall, Jeffrey
    Strapp, Ellen
    Putney, David
    Gordon-Burroughs, Sherilyn
    Monsour, Howard
    CRITICAL CARE MEDICINE, 2012, 40 (12) : U247 - U248
  • [45] Clinical laboratory measurement of direct factor Xa inhibitors: Anti-Xa assay is preferable to prothrombin time assay
    Barrett, Yu Chen
    Wang, Zhaoqing
    Frost, Charles
    Shenker, Andrew
    THROMBOSIS AND HAEMOSTASIS, 2010, 104 (06) : 1263 - 1271
  • [46] LABORATORY VALIDATION OF EDOXABAN CALIBRATED CHROMOGENIC ANTI-XA ASSAY
    Leger, Rachel
    Tange, Julie
    Chen, Dong
    Pruthi, Rajiv
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2019, 41 : 99 - 99
  • [47] Inter-assay and instrument variability of anti-Xa - Results
    Kovacs, MJ
    Keeney, M
    THROMBOSIS AND HAEMOSTASIS, 2000, 84 (01) : 138 - 138
  • [48] A MECHANIZED AMIDOLYTIC ASSAY FOR THE DETERMINATION OF ANTI-XA LEVELS IN PLASMA
    TENCATE, H
    LAMPING, RJ
    HENNY, CP
    PRINS, A
    TENCATE, JW
    HAEMOSTASIS, 1982, 12 (1-2) : 51 - 51
  • [49] Universal anti-Xa assay for rivaroxaban and apixaban plasma concentration
    Studt, Jan-Dirk
    Stein, Philipp
    Hegemann, Inga
    Koch, Karin
    Sangalli-Baruffaldi, Jessica
    Nagler, Michael
    SWISS MEDICAL WEEKLY, 2018, 148 : 37S - 37S
  • [50] EVALUATION OF ANTI-XA AND APTT MONITORING OF HEPARIN IN ADULT PATIENTS RECEIVING ECMO SUPPORT
    Arnouk, Serena
    Altshuler, Diana
    Merchan, Cristian
    Zakhary, Bishoy
    Papadopoulos, John
    CRITICAL CARE MEDICINE, 2018, 46 (01) : 88 - 88