Measured and Monte Carlo simulated electron backscatter to the monitor chamber for the Varian TrueBeam Linac

被引:2
|
作者
Lloyd, Samantha A. M. [1 ]
Gagne, Isabelle M. [1 ,2 ]
Bazalova-Carter, Magdalena [1 ]
Zavgorodni, Sergei [1 ,2 ]
机构
[1] Univ Victoria, Dept Phys & Astron, Victoria, BC, Canada
[2] BC Canc Agcy, Vancouver Isl Ctr, Dept Med Phys, Victoria, BC, Canada
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2016年 / 61卷 / 24期
关键词
electron radiation therapy; Monte Carlo; TrueBeam; backscatter; PHOTON BEAMS; MULTILEAF COLLIMATOR; DOSE CALCULATIONS; BREAST-CANCER; THERAPY; ACCELERATORS;
D O I
10.1088/1361-6560/61/24/8779
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
To accurately simulate therapeutic electron beams using Monte Carlo methods, backscatter from jaws into the monitor chamber must be accounted for via the backscatter factor, S-b. Measured and simulated values of Sb for the TrueBeam are investigated. Two approaches for measuring Sb are presented. Both require service mode operation with the dose and pulse forming networking servos turned off in order to assess changes in dose rate with field size. The first approach samples an instantaneous dose rate, while the second approach times the delivery of a fixed number of monitor units to assess dose rate. Dose rates were measured for 6, 12 and 20 MeV electrons for jaw- or MLC-shaped apertures between 1x1 and 40x40 cm(2). The measurement techniques resulted in values of Sb that agreed within 0.21% for square and asymmetric fields collimated by the jaws. Measured values of Sb were used to calculate the forward dose component in a virtual monitor chamber using BEAMnrc. Based on this forward component, simulated values of Sb were calculated and compared to measurement and Varian's VirtuaLinac simulations. BEAMnrc results for jaw- shaped fields agreed with measurements and with VirtuaLinac simulations within 0.2%. For MLC-shaped fields, the respective measurement techniques differed by as much as 0.41% and BEAMnrc results differed with measurement by as much as 0.4%, however, all measured and simulated values agreed within experimental uncertainty. Measurement sensitivity was not sufficient to capture the small backscatter effect due to the MLC, and Monte Carlo predicted backscatter from the MLC to be no more than 0.3%. Backscatter from the jaws changed the electron dose rate by up to 2.6%. This reinforces the importance of including a backscatter factor in simulations of electron fields shaped with secondary collimating jaws, but presents the option of ignoring it when jaws are retracted and collimation is done with the MLC.
引用
收藏
页码:8779 / 8793
页数:15
相关论文
共 50 条
  • [1] Measured and Monte Carlo simulated electron backscatter to the monitor chamber for the varian TrueBeam linac (vol 61, pg 8779, 2016)
    Lloyd, Samantha A. M.
    Gagne, Isabelle M.
    Bazalova-Carter, Magdalena
    Zavgorodni, Sergei
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (12): : 5198 - 5199
  • [2] TrueBeam Monte Carlo Absolute Dose Calculations Using Monitor Chamber Backscatter Simulations and Linac-Logged Target Current
    Popescu, I. A.
    Lobo, J.
    Sawkey, D.
    Svatos, M.
    MEDICAL PHYSICS, 2014, 41 (06) : 567 - 567
  • [3] Monte Carlo Modeling and Simulation of the Varian TrueBeam LINAC Using Heterogeneous Computing
    Lin, H.
    Liu, T.
    Su, L.
    Shi, C.
    Tang, X.
    Adam, D.
    Bednarz, B.
    Xu, X.
    MEDICAL PHYSICS, 2017, 44 (06) : 3003 - 3003
  • [4] Monitor backscatter factors for the Varian 21EX and TrueBeam linear accelerators: measurements and Monte Carlo modelling
    Zavgorodni, Sergei
    Alhakeem, Eyad
    Townson, Reid
    PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (04): : 911 - 924
  • [5] Simple and Efficient Modeling of the Varian TrueBeam Linac Using the GATE Monte Carlo Platform
    Roth, B.
    Ploquin, N.
    Wiebe, J.
    MEDICAL PHYSICS, 2015, 42 (06) : 3497 - 3497
  • [6] A Monte Carlo Dose Calculation Framework for Electron Beams On Varian TrueBeam
    Rodrigues, A.
    Sawkey, D.
    Yin, F.
    Wu, Q.
    MEDICAL PHYSICS, 2014, 41 (06) : 314 - +
  • [7] A Comprehensive Geant4 Monte Carlo Model of Multileaf Collimator for Varian TrueBeam Linac
    Kandlakunta, P.
    Zhang, T.
    Mutic, S.
    Khan, R.
    MEDICAL PHYSICS, 2017, 44 (06) : 2909 - 2909
  • [8] Monte Carlo Simulation of a 6MV Varian Truebeam Without Flattening Filter Linac
    Song, T.
    Zhou, L.
    Jiang, S.
    MEDICAL PHYSICS, 2012, 39 (06) : 3819 - 3819
  • [9] Monte Carlo simulation of a Varian's TrueBeam linac as a Clinac 2100: A feasibility study.
    Castillo, M. L. Rodriguez
    Brualla, L.
    Sempau, J.
    RADIOTHERAPY AND ONCOLOGY, 2020, 152 : S740 - S740
  • [10] A Monte Carlo study of backscatter in the monitor ion chamber for clinical photon and electron beams
    Verhaegen, F
    Symonds-Tayler, R
    Mubata, C
    Bidmead, M
    Nahum, A
    Liu, HH
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 466 - 468