Thermal conductivity measurement under hydrostatic pressure using the 3ω method

被引:49
|
作者
Chen, F [1 ]
Shulman, J
Xue, Y
Chu, CW
Nolas, GS
机构
[1] Univ Houston, Texas Ctr Superconduct & Adv Mat, Houston, TX 77204 USA
[2] Univ S Florida, Dept Phys, Tampa, FL 33620 USA
[3] Hong Kong Univ Sci & Technol, Lawrence Berkeley Natl Lab, Hong Kong, Hong Kong, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2004年 / 75卷 / 11期
关键词
D O I
10.1063/1.1805771
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
We have designed and modeled new techniques, based on the 3omega method, to measure thermal conductivity of liquids (kappa(l)) and solids (kappa(s)) under hydrostatic pressure (P). The system involves a solid sample immersed in a liquid pressure medium, both of which have unknown thermal properties. The temperature (T) and P dependance of kappa(l) are first determined through the use of a modified 3omega technique. This method uses a conducting wire (Pt, in this work), which is immersed in the pressure medium, as the heater/sensor. In addition to kappa(l), this allows for the accurate determination of the specific heat per volume of the liquid and Pt, (rhoC)(l) and (rhoC)(Pt), respectively. The information of kappa(l) and (rhoC)(l) can then be used to make corrections to measurements of kappa(s), in which the sample is immersed in the pressure medium, and a metal strip acts as the heater/sensor. We present the T and P dependence of kappa(l) and (rhoC)(l) for the widely used pressure medium 3M Fluorinert FC77 up to 0.8 GPa. The measurement of kappa(s) for a thermoelectric clathrate material, Sr8Ga16Ge30, in FC77 is analyzed in detail, and the refined data achieves an accuracy of 1%. The setup can be modified to measure kappa and rhoC up to 3.5 GPa. (C) 2004 American Institute of Physics.
引用
收藏
页码:4578 / 4584
页数:7
相关论文
共 50 条
  • [31] Error factors in precise thermal conductivity measurement using 3ω method for wire samples
    Yuki Sekimoto
    Ryo Abe
    Hirotaka Kojima
    Hiroaki Benten
    Masakazu Nakamura
    Journal of Thermal Analysis and Calorimetry, 2023, 148 : 2285 - 2296
  • [32] Thermal conductivity measurement and sedimentation detection of aluminum oxide nanofluids by using the 3ω method
    Oh, Dong-Wook
    Jain, Ankur
    Eaton, John K.
    Goodson, Kenneth E.
    Lee, Joon Sik
    INTERNATIONAL JOURNAL OF HEAT AND FLUID FLOW, 2008, 29 (05) : 1456 - 1461
  • [33] Measurement of thermal conductivity of fluids using 3-ω method in a suspended micro wire
    R. Karthik
    N. Harish Nagarajan
    B. Raja
    P. Damodharan
    Journal of Engineering Thermophysics, 2012, 21 : 60 - 68
  • [34] INFLUENCE OF HYDROSTATIC PRESSURE ON THERMAL-CONDUCTIVITY OF SEMICONDUCTING MATERIALS
    AVERKIN, AA
    ZHAPAROV, ZZ
    STILBANS, LS
    SOVIET PHYSICS SEMICONDUCTORS-USSR, 1972, 5 (11): : 1954 - &
  • [35] Hydrostatic Pressure Effect on Lattice Thermal Conductivity in Si Nanofilms
    Hamarashid, M. M.
    Omar, M. S.
    Qader, Ibrahim Nazem
    SILICON, 2022, 14 (18) : 12789 - 12798
  • [36] Hydrostatic Pressure Effect on Lattice Thermal Conductivity in Si Nanofilms
    M. M. Hamarashid
    M. S. Omar
    Ibrahim Nazem Qader
    Silicon, 2022, 14 : 12789 - 12798
  • [37] Thermal Conductivity Measurement of Fuzzy W Using a Thermoreflectance Method
    Kajita, Shin
    Yagi, Takashi
    Ohno, Noriyasu
    IEEE TRANSACTIONS ON PLASMA SCIENCE, 2024, 52 (09) : 4365 - 4370
  • [38] Thermoelectric power measurement under hydrostatic pressure using a self-clamped pressure cell
    Choi, ES
    Kang, H
    Jo, YJ
    Kang, W
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2002, 73 (08): : 2999 - 3002
  • [39] Analysis of the 3-omega method for thermal conductivity measurement
    Wang, Hainan
    Sen, Mihir
    INTERNATIONAL JOURNAL OF HEAT AND MASS TRANSFER, 2009, 52 (7-8) : 2102 - 2109
  • [40] MEASUREMENT OF THERMAL-CONDUCTIVITY OF NONWOVENS USING A DYNAMIC METHOD
    MARTIN, JR
    LAMB, GER
    TEXTILE RESEARCH JOURNAL, 1987, 57 (12) : 721 - 727