Estimation of mean glandular dose for mammography of augmented breasts

被引:8
|
作者
Beckett, JR [1 ]
Kotre, CJ [1 ]
机构
[1] Newcastle Gen Hosp, Dept Reg Med Phys, Newcastle Upon Tyne NE4 6BE, Tyne & Wear, England
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2000年 / 45卷 / 11期
关键词
D O I
10.1088/0031-9155/45/11/309
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The standard quantity used to relate breast surface exposure to radiation risk is the mean dose received by the radiation sensitive tissue contained within the female breast, the mean glandular dose (MGD). At present, little is known about the MGD received by women with breast implants as there is no technique available to facilitate its calculation. The present work has involved modification of the conventional method for MGD estimation to make it applicable to women with augmented breasts. The technique was used to calculate MGDs for a cohort of 80 women with breast implants, which were compared with similar data calculated for a total of 1258 non-augmented women. Little difference was found in median MGD at low compressed breast thickness. At high breast thickness, however, the MGDs received by women with augmented breasts were found to be considerably lower than those relating to their non-augmented counterparts.
引用
收藏
页码:3241 / 3252
页数:12
相关论文
共 50 条
  • [21] Estimation of glandular dose in mammography based on artificial neural networks
    Trevisan Massera, Rodrigo
    Tomal, Alessandra
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2020, 65 (09):
  • [22] MODEL FOR COMPUTING TUNGSTEN TUBE SPECTRA AND MEAN GLANDULAR DOSE FOR DIGITAL MAMMOGRAPHY
    SPIVEY, BA
    ANDRE, MP
    STOLLER, M
    TRAN, J
    [J]. RADIOLOGY, 1995, 197 : 291 - 291
  • [23] A Monte Carlo model for mean glandular dose evaluation in spot compression mammography
    Sarno, Antonio
    Dance, David R.
    van Engen, Ruben E.
    Young, Kenneth C.
    Russo, Paolo
    Di Lillo, Francesca
    Mettivier, Giovanni
    Bliznakova, Kristina
    Fei, Baowei
    Sechopoulos, Ioannis
    [J]. MEDICAL PHYSICS, 2017, 44 (07) : 3848 - 3860
  • [24] Dosimetric evaluation of the mean glandular dose for mammography in Korean women: A preliminary report
    Oh, KK
    Hur, J
    Kim, EK
    Choo, SS
    [J]. YONSEI MEDICAL JOURNAL, 2003, 44 (05) : 863 - 868
  • [25] Investigations on a Computer Application for Tracking the Mean Glandular Breast Dose Profile in Mammography
    Schiabel, Homero
    Barufaldi, Bruno
    Ruberti Filha, Eny M.
    [J]. XV MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING - MEDICON 2019, 2020, 76 : 869 - 873
  • [26] The Impact of Mammography Physics Instrumentation On Entrance Air Kerma and Mean Glandular Dose
    Brateman, L.
    Heintz, P.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [27] A Study of Mean Glandular Dose during Diagnostic Mammography in Hospitals in Hanoi, Vietnam
    Nguyen Thai Ha
    Nguyen Duc Thuan
    Nguyen Thu Van
    [J]. THIRD INTERNATIONAL CONFERENCE ON THE DEVELOPMENT OF BIOMEDICAL ENGINEERING IN VIETNAM, 2010, 27 : 263 - 265
  • [28] Investigation of mean glandular dose versus compressed breast thickness relationship for mammography
    Bor, Dogan
    Tukel, Selma
    Olgar, Turan
    Toklu, Tuerkay
    Aydin, Elif
    Akyol, Oya
    [J]. RADIATION PROTECTION DOSIMETRY, 2008, 129 (1-3) : 160 - 164
  • [29] Digital full field mammography for imaging augmented breasts
    Diekmann, S
    Diekmann, F
    Hauschild, M
    Hamm, B
    [J]. RADIOLOGE, 2002, 42 (04): : 275 - 279
  • [30] Evaluation of mean glandular dose from mammography screening: A single-center study
    Alahmad, Haitham
    AlEnazi, Khaled
    Alshahrani, Abdulrhman
    Albariqi, Saleh
    Alnafea, Mohammed
    [J]. JOURNAL OF RADIATION RESEARCH AND APPLIED SCIENCES, 2023, 16 (04)