Stability of volar locking plate systems for AO type C3 fractures of the distal radius: biomechanical study in a cadaveric model

被引:25
|
作者
Kamei, Shuzo [1 ]
Osada, Denju [1 ]
Tamai, Kazuya [1 ]
Kato, Nakayuki [1 ]
Takai, Morimitsu [1 ]
Kameda, Masahiro [1 ]
Nohara, Yutaka [1 ]
机构
[1] Dokkyo Med Univ, Sch Med, Dept Orthopaed Surg, Mibu, Tochigi 3210293, Japan
关键词
FRAGMENT-SPECIFIC FIXATION; COMPRESSION PLATE; EXTERNAL FIXATION; DORSAL; SCREW;
D O I
10.1007/s00776-010-1466-0
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of the present study was to compare the relative stability of five volar locking plates (all of which are available for the treatment of intraarticular fractures of the distal radius) under loading conditions simulating the physiological forces that occur during early active mobilization. Five plating techniques were applied to surgically simulated AO type C3.2 distal radius fractures in formalin-fixed cadavers. The specimens were tested with a servohydraulic materials testing machine with 250 N of axial compression load for 3000 cycles. After cyclic loading, the specimens were loaded until they demonstrated failure in axial compression. The five fixation systems studied included a DRV locking plate (group 1), a Stellar plate (group 2), an Acu-Loc plate (group 3), AO Locking Distal Radius System 2.4 (group 4); and a Matrix SmartLock plate (group 5). None of the plate fixations tested failed during the cyclic loading. Group 2 had a higher elastic limit than groups 4 and 5. There were no significant differences among the five groups for the failure load. Failure occurred at the distal portion of the fixation system, at the ulnar side locking pin, or the locking screw was bent (groups 1, 2, 3); the ulnar side locking screw was broken (groups 3, 4); the locking screw became loose (group 4); and the ulnar side locking screw was uncoupled from the screw hole (group 5). All of the five volar plate fixation systems provided sufficient stability to permit 3000 repeated motions of the digits after surgery for AO type C3 distal radius fractures.
引用
收藏
页码:357 / 364
页数:8
相关论文
共 50 条
  • [31] Outcomes of Primary Volar Locking Plate Fixation of Open Distal Radius Fractures
    Joo, Min Su
    Kang, Hong Je
    Yu, Hyun Kyu
    Lee, Jae Sung
    JOURNAL OF HAND SURGERY-ASIAN-PACIFIC VOLUME, 2022, 27 (03): : 517 - 523
  • [32] Volar plate fixation of AQ type C2 and C3 distal radius fractures, a single-center study of 55 patients
    Gerald, Gruber
    Karl, Gruber
    Christian, Giessauf
    Heimo, Clar
    Max, Zacherl
    Florentine, Fuerst
    Alexander, Bernhardt G.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2008, 22 (07) : 467 - 472
  • [33] Biomechanical Study on the Cyclic Stability of Distal Radius C2 Type Osteoporotic Fractures Using Locking Plates in a Cadaver Model
    Chih-Kun Hsiao
    Yuan-Kun Tu
    Cheng-Hung Lee
    Cheng-Yo Yen
    Feng-Chen Kao
    Yi-Jung Tsai
    Journal of Medical and Biological Engineering, 2018, 38 : 707 - 714
  • [34] Biomechanical Study on the Cyclic Stability of Distal Radius C2 Type Osteoporotic Fractures Using Locking Plates in a Cadaver Model
    Hsiao, Chih-Kun
    Tu, Yuan-Kun
    Lee, Cheng-Hung
    Yen, Cheng-Yo
    Kao, Feng-Chen
    Tsai, Yi-Jung
    JOURNAL OF MEDICAL AND BIOLOGICAL ENGINEERING, 2018, 38 (05) : 707 - 714
  • [35] Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model
    Hoegel, F.
    Mair, S.
    Eberle, S.
    Weninger, P.
    von Oldenburg, G.
    Augat, P.
    ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2013, 133 (01) : 51 - 57
  • [36] Distal radius fracture fixation with volar locking plates and additional bone augmentation in osteoporotic bone: a biomechanical study in a cadaveric model
    F. Högel
    S. Mair
    S. Eberle
    P. Weninger
    G. von Oldenburg
    P. Augat
    Archives of Orthopaedic and Trauma Surgery, 2013, 133 : 51 - 57
  • [37] Biomechanical evaluation of the stability of extra-articular distal radius fractures fixed with volar locking plates according to the length of the distal locking screw
    Oh, Gyung-Hwan
    Kim, Hak-Sung
    Lee, Jung Il
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2021, 24 (08) : 922 - 932
  • [38] Biomechanical Properties of Volar Hybrid and Locked Plate Fixation in Distal Radius Fractures
    Sokol, Shima C.
    Amanatullah, Derek F.
    Curtiss, Shane
    Szabo, Robert M.
    JOURNAL OF HAND SURGERY-AMERICAN VOLUME, 2011, 36A (04): : 591 - 597
  • [39] CT-Based Micromotion Analysis After Locking Plate Fixation of AO Type C Distal Radius Fractures
    Eva Lundqvist
    Henrik Olivecrona
    Per Wretenberg
    Marcus Sagerfors
    Indian Journal of Orthopaedics, 2023, 57 : 2031 - 2039
  • [40] CT-Based Micromotion Analysis After Locking Plate Fixation of AO Type C Distal Radius Fractures
    Lundqvist, Eva
    Olivecrona, Henrik
    Wretenberg, Per
    Sagerfors, Marcus
    INDIAN JOURNAL OF ORTHOPAEDICS, 2023, 57 (12) : 2031 - 2039