Dual-energy contrast enhanced digital mammography using a new approach for breast tissue canceling

被引:5
|
作者
Puong, Sylvie [1 ,2 ]
Bouchevreau, Xavier [3 ]
Patoureaux, Fanny [2 ]
Iordache, Razvan [2 ]
Muller, Serge [2 ]
机构
[1] Univ Paris 11, Bat 425, Orsay, France
[2] GE Hlthcare, Buc, France
[3] Altran, Paris, France
关键词
digital mammography; contrast medium; contrast enhanced digital mammography; dual-energy; algorithm description;
D O I
10.1117/12.710133
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, we propose a novel approach to dual-energy contrast-enhanced digital mammography, with the development of a dual-energy recombination algorithm based on an image chain model and the determination of the associated optimal low and high-energy techniques. Our method produces clutter-free iodine-equivalent images and includes thickness correction near the breast border. After the algorithm description, the optimal low and high-energy acquisition techniques are determined to obtain a compromise between image quality and glandular dose. The low and high-energy techniques were chosen to minimize the glandular dose for a target Signal Difference to Noise Ratio (SDNR) in the dual-energy recombined image. The theoretical derivation of the iodine SDNR in the recombined image allowed the prediction of the optimal low and high-energy techniques. Depending on the breast thickness and glandular percentage, the optimal low-energy kVp and mAs ranged from 24kVp (Mo/Mo or Mo/Rh) to 35kVp (Rh/Rh), and from 60 to 90mAs respectively, and the high-energy kVp and mAs ranged from 40kVp to 47kVp (Mo/Cu), and from 80mAs to 290mAs. We proved the better :performance of our algorithm compared to the classic weighted logarithmic subtraction method in terms of patient dose and also in terms of texture cancelation, through the use of artificial textured images. Values of iodine contrast measured on phantom were close to the expected iodine thickness. Good correlation was found between the measured and theoretical iodine SDNR in the dual-energy images, which validates our theoretical optimization of the acquisition techniques.
引用
收藏
页数:12
相关论文
共 50 条
  • [31] Segmentation methods for breast vasculature in dual-energy contrast-enhanced digital breast tomosynthesis
    Lau, Kristen C.
    Lee, Hyo Min
    Singh, Tanushriya
    Maidment, Andrew D. A.
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [32] DUAL-ENERGY CONTRAST-ENHANCED DIGITAL MAMMOGRAPHY: PATIENT RADIATION DOSE ESTIMATION USING A MONTE CARLO CODE
    Yakoumakis, E.
    Tzamicha, E.
    Dimitriadis, A.
    Georgiou, E.
    Tsapaki, V.
    Chalazonitis, A.
    RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 369 - 372
  • [33] Classification of breast microcalcifications using dual-energy mammography
    Ghammraoui, Bahaa
    Makeev, Andrey
    Zidan, Ahmed
    Alayoubi, Alaadin
    Glick, Stephen J.
    JOURNAL OF MEDICAL IMAGING, 2019, 6 (01)
  • [34] Classification of breast microcalcifications using dual-energy mammography
    Ghammraoui, B.
    Makeev, A.
    Glick, S. J.
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [35] Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study
    Clarisse Dromain
    Fabienne Thibault
    Felix Diekmann
    Eva M Fallenberg
    Roberta A Jong
    Marcia Koomen
    R Edward Hendrick
    Anne Tardivon
    Alicia Toledano
    Breast Cancer Research, 14
  • [36] Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study
    Dromain, Clarisse
    Thibault, Fabienne
    Diekmann, Felix
    Fallenberg, Eva M.
    Jong, Roberta A.
    Koomen, Marcia
    Hendrick, R. Edward
    Tardivon, Anne
    Toledano, Alicia
    BREAST CANCER RESEARCH, 2012, 14 (03):
  • [37] A scatter correction method for contrast-enhanced dual-energy digital breast tomosynthesis
    Lu, Yihuan
    Peng, Boyu
    Lau, Beverly A.
    Hu, Yue-Houng
    Scaduto, David A.
    Zhao, Wei
    Gindi, Gene
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (16): : 6323 - 6354
  • [38] Dual-energy contrast enhanced digital breast tomosynthesis: concept, method and evaluation on phantoms
    Puong, Sylvie
    Patoureaux, Fanny
    Iordache, Razvan
    Bouchevreau, Xavier
    Muller, Serge
    MEDICAL IMAGING 2007: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2007, 6510
  • [39] Digital breast tomosynthesis and contrast-enhanced dual-energy digital mammography alone and in combination compared to 2D digital synthetized mammography and MR imaging in breast cancer detection and classification
    Petrillo, Antonella
    Fusco, Roberta
    Vallone, Paolo
    Filice, Salvatore
    Granata, Vincenza
    Petrosino, Teresa
    Rubulotta, Maria Rosaria
    Setola, Sergio Venanzio
    Raso, Mauro Mattace
    Maio, Francesca
    Raiano, Concetta
    Siani, Claudio
    Di Bonito, Maurizio
    Botti, Gerardo
    BREAST JOURNAL, 2020, 26 (05): : 860 - 872
  • [40] Breast Glandular Tissue Volume and Volumetric Glandular Ratio Using Dual-Energy Mammography
    Pacheco-Guevara, G.
    Castillo-Lopez, J.
    Villasenor-Navarro, Y.
    Brandan, M. E.
    MEDICAL PHYSICS, 2020, 47 (06) : E308 - E308