Dual-energy contrast enhanced digital mammography using a new approach for breast tissue canceling

被引:5
|
作者
Puong, Sylvie [1 ,2 ]
Bouchevreau, Xavier [3 ]
Patoureaux, Fanny [2 ]
Iordache, Razvan [2 ]
Muller, Serge [2 ]
机构
[1] Univ Paris 11, Bat 425, Orsay, France
[2] GE Hlthcare, Buc, France
[3] Altran, Paris, France
关键词
digital mammography; contrast medium; contrast enhanced digital mammography; dual-energy; algorithm description;
D O I
10.1117/12.710133
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
In this study, we propose a novel approach to dual-energy contrast-enhanced digital mammography, with the development of a dual-energy recombination algorithm based on an image chain model and the determination of the associated optimal low and high-energy techniques. Our method produces clutter-free iodine-equivalent images and includes thickness correction near the breast border. After the algorithm description, the optimal low and high-energy acquisition techniques are determined to obtain a compromise between image quality and glandular dose. The low and high-energy techniques were chosen to minimize the glandular dose for a target Signal Difference to Noise Ratio (SDNR) in the dual-energy recombined image. The theoretical derivation of the iodine SDNR in the recombined image allowed the prediction of the optimal low and high-energy techniques. Depending on the breast thickness and glandular percentage, the optimal low-energy kVp and mAs ranged from 24kVp (Mo/Mo or Mo/Rh) to 35kVp (Rh/Rh), and from 60 to 90mAs respectively, and the high-energy kVp and mAs ranged from 40kVp to 47kVp (Mo/Cu), and from 80mAs to 290mAs. We proved the better :performance of our algorithm compared to the classic weighted logarithmic subtraction method in terms of patient dose and also in terms of texture cancelation, through the use of artificial textured images. Values of iodine contrast measured on phantom were close to the expected iodine thickness. Good correlation was found between the measured and theoretical iodine SDNR in the dual-energy images, which validates our theoretical optimization of the acquisition techniques.
引用
收藏
页数:12
相关论文
共 50 条
  • [21] Motion artifacts in dual-energy contrast-enhanced mammography
    Allec, Nicholas
    Abbaszadeh, Shiva
    Lewin, John M.
    Karim, Karim S.
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [22] Dual-energy contrast-enhanced spectral mammography (CESM)
    Martin Daniaux
    Tobias De Zordo
    Wolfram Santner
    Birgit Amort
    Florian Koppelstätter
    Werner Jaschke
    Clarisse Dromain
    Willi Oberaigner
    Michael Hubalek
    Christian Marth
    Archives of Gynecology and Obstetrics, 2015, 292 : 739 - 747
  • [23] Glandular dose in contrast-enhanced dual-energy mammography
    Cossu, E.
    Castellani, F.
    Fiaschetti, V.
    Mariateresa, M.
    Tiziana, R.
    Angelo, T.
    CANCER RESEARCH, 2019, 79 (04)
  • [24] A novel approach to background subtraction in contrast-enhanced dual-energy digital mammography with commercially available mammography devices: Noise minimization
    Contillo, Adriano
    Di Domenico, Giovanni
    Cardarelli, Paolo
    Gambaccini, Mauro
    Taibi, Angelo
    MEDICAL PHYSICS, 2016, 43 (06) : 3080 - 3089
  • [25] A novel approach to background subtraction in contrast-enhanced dual-energy digital mammography with commercially available mammography devices: Polychromaticity correction
    Contillo, Adriano
    Di Domenico, Giovanni
    Cardarelli, Paolo
    Gambaccini, Mauro
    Taibi, Angelo
    MEDICAL PHYSICS, 2015, 42 (11) : 6641 - 6650
  • [26] Feasibility Study for the Improvement of Microcalcification Visualization in Different Breast Thicknesses and Tissue Components Using a Dual-Energy Approach in Digital Mammography
    Tsai, Chia-Jung
    Chen, Ran-Chou
    Hung, Shuo-Hui
    Wu, Jay
    Peng, Hui-Ling
    Lee, Jason J. S.
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2012, 36 (04) : 488 - 494
  • [27] Establishing a Quality Control protocol for dual-energy based contrast-enhanced digital mammography
    Cockmartin, L.
    Bosmans, H.
    Marshall, N. W.
    MEDICAL IMAGING 2021: PHYSICS OF MEDICAL IMAGING, 2021, 11595
  • [28] Dual-energy contrast-enhanced digital breast tomosynthesis - a feasibility study
    Carton, A-K
    Gavenonis, S. C.
    Currivan, J. A.
    Conant, E. F.
    Schnall, M. D.
    Maidment, A. D. A.
    BRITISH JOURNAL OF RADIOLOGY, 2010, 83 (988): : 344 - 350
  • [29] A method for material decomposition in dual-energy contrast enhancement digital mammography
    Kahani, M.
    Kamali-asl, A.
    Ghadiri, H.
    Hashemi, S.
    MEASUREMENT, 2016, 88 : 87 - 95
  • [30] Technique and clinical applications of dual-energy contrast-enhanced digital mammography (CEDM) in breast cancer evaluation: a pictorial essay
    Sudhir, Rashmi
    Koppula, Veeraiah
    Mandava, Anitha
    Kamala, Sannapareddy
    Potlapalli, Alekya
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2021, 27 (01) : 28 - 36