Newborn screening for hepatorenal tyrosinemia by tandem mass spectrometry: analysis of succinylacetone extracted from dried blood spots

被引:66
|
作者
Allard, P
Grenier, A
Korson, MS
Zytkovicz, TH
机构
[1] Tufts New England Med Ctr, Div Genet, Boston, MA 02111 USA
[2] Tufts Univ, Sch Med, Boston, MA 02111 USA
[3] Univ Laval, Ctr Med, Dept Med Genet, St Foy, PQ G1V 4G2, Canada
[4] Univ Massachusetts, Sch Med, New England Newborn Screening Program, Jamaica Plain, MA 02130 USA
关键词
tyrosinemia; succinylacetone; newborn screening; dried blood spots; filter paper; tandem mass spectrometry;
D O I
10.1016/j.clinbiochem.2004.07.006
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Objectives: To develop a method for the determination of succinylacetone (SA) in dried blood spots (DBS) using tandem mass spectrometry (MS/MS). Methods: SA was extracted from DBS with an acetonitrile and water solution (80:20 by volume) containing formic acid and hydrazine hydrate (both at 0.1% by Volume), and analyzed by MS/MS with a total run time per sample under 2 min. The reference range for SA in newborns was determined by analyzing a control group of 3199 DBS. SA was also measured in stored newborn specimens from three patients diagnosed clinically with hepatorenal tyrosinernia (FIT). Results: The within-run precision was less than or equal to7.4% and total precision was less than or equal to 12.2% on blood spots fortified with SA at 2, 10, and 50 [mumol/L. The limit of quantitation was 1 mumol/L and the calibration was linear from 1 to 50 mumol/L. A comparison of SA analysis of this MS/MS method with an established enzyme assay indirectly quantitating SA by inhibition of delta-aminolevulinic acid clehydratase demonstrated a strong con-elation. The reference range in the control group of unaffected newborn was determined to be <2 mumol/L, while SA in retrieved DBS Frorri FIT patients was significantly elevated, measuring 46.7, 36.5. and 23.2 mumol/L. Conclusions: We report a simple method that can be Used for screening HT in newborns. Based on our limited experience, a 2 mumol/L Cutoff could result in up to 100% sensitivity and specificity. (C) 2004 The Canadian Society of Clinical Chemists. All rights reserved.
引用
收藏
页码:1010 / 1015
页数:6
相关论文
共 50 条
  • [41] Quantification of derivatized phenylalanine and tyrosine in dried blood spots using liquid chromatography with tandem spectrometry for newborn screening of phenylketonuria
    Duh, Tsai-Hui
    Liang, Yu-Ching
    Shen, Po Tsun
    Ke, Yi-Wen
    Nian, Yan-Tian
    Liang, Shih-Shin
    EUROPEAN JOURNAL OF MASS SPECTROMETRY, 2024, 30 (02) : 133 - 140
  • [42] Determination of EDTA in dried blood samples by tandem mass spectrometry avoids serious errors in newborn screening
    Fingerhut, Ralph
    Dame, Torsten
    Olgemoeller, Bernhard
    EUROPEAN JOURNAL OF PEDIATRICS, 2009, 168 (05) : 553 - 558
  • [43] Determination of EDTA in dried blood samples by tandem mass spectrometry avoids serious errors in newborn screening
    Ralph Fingerhut
    Torsten Dame
    Bernhard Olgemöller
    European Journal of Pediatrics, 2009, 168 : 553 - 558
  • [44] Screening for mucopolysaccharidoses' subtypes by tandem mass spectrometry determination of disease-specific oligosaccharides in dried blood spots
    Lemos da Silva, Layzon Antonio
    Faqueti, Larissa
    Iop, Gabrielle
    Soares, Juliano
    Borges, Henrique B. L.
    Kubaski, Francyne
    Trapp, Franciele B.
    Giugliani, Roberto
    MOLECULAR GENETICS AND METABOLISM, 2023, 138 (02) : 79 - 80
  • [45] Multiplex enzyme assay screening of dried blood spots for lysosomal storage disorders by using tandem mass spectrometry
    Zhang, X. Kate
    Elbin, Carole S.
    Chuang, Wei-Lien
    Cooper, Samantha K.
    Marashio, Carla A.
    Beauregard, Christa
    Keutzer, Joan M.
    CLINICAL CHEMISTRY, 2008, 54 (10) : 1725 - 1728
  • [46] Tandem mass spectrometry in newborn screening
    不详
    JOURNAL OF MEDICAL SCREENING, 2000, 7 (03) : 164 - 164
  • [47] Development of tandem mass spectrometry-based creatinine measurement using dried blood spot for newborn mass screening
    Masaru Nakano
    Osamu Uemura
    Masataka Honda
    Tetsuya Ito
    Yoko Nakajima
    Shinji Saitoh
    Pediatric Research, 2017, 82 : 237 - 243
  • [48] Development of tandem mass spectrometry-based creatinine measurement using dried blood spot for newborn mass screening
    Nakano, Masaru
    Uemura, Osamu
    Honda, Masataka
    Ito, Tetsuya
    Nakajima, Yoko
    Saitoh, Shinji
    PEDIATRIC RESEARCH, 2017, 82 (02) : 237 - 243
  • [49] Tandem mass spectrometry in newborn screening
    Matern, D
    ENDOCRINOLOGIST, 2002, 12 (01): : 50 - 57
  • [50] Use of tandem mass spectrometry for multianalyte screening of dried blood specimens from newborns
    Chace, DH
    Kalas, TA
    Naylor, EW
    CLINICAL CHEMISTRY, 2003, 49 (11) : 1797 - 1817