Influence of developer exhaustion on accuracy of quantitative digital subtraction radiography - An in vitro study

被引:6
|
作者
Fidler, A
Likar, B
Pernus, F
Skaleric, U
机构
[1] Univ Ljubljana, Fac Med, Dept Oral Med & Periodontol, SI-1000 Ljubljana, Slovenia
[2] Univ Ljubljana, Fac Med, Dept Stomatol, SI-1000 Ljubljana, Slovenia
[3] Univ Ljubljana, Fac Med, Dept Elect Engn, SI-1000 Ljubljana, Slovenia
关键词
D O I
10.1067/moe.2000.107362
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objective. The aim of the study was to evaluate the influence of developer exhaustion on accuracy of quantitative digital subtraction radiography. Study design. Six objects, each incorporating a section of dry human mandible, were radiographed with 4 exposure times. Baseline films were processed in fresh solutions, whereas follow-up films were processed in fresh and in increasingly exhausted solutions (ie, 1, 2, and 3 weeks old). Bone loss and bone gain were computer simulated in 17 regions of interest on baseline radiographs. Area and volume of changes in mineralization were measured in subtracted images, obtained by subtraction of baseline from their corresponding follow-up radiographs. Friedman's 2-way analysis of variance by ranks and Wilcoxon signed-rank test were used for statistical analysis. Results. Because of exhausted developer, bone loss was relatively underestimated from 6.6% to 16.5% (P < .05), whereas bone gain was relatively overestimated from 9.7% to 16.7% (P < .05). Conclusions. This in vitro study demonstrates that films for quantitative digital subtraction radiography should be processed in fresh developer or error might be introduced.