Microtube-based electrode arrays for low invasive extracellular recording with a high signal-to-noise ratio

被引:12
|
作者
Takei, Kuniharu [1 ]
Kawano, Takeshi [1 ]
Kawashima, Takahiro [2 ]
Sawada, Kazuaki [1 ]
Kaneko, Hidekazu [3 ]
Ishida, Makoto [1 ]
机构
[1] Toyohashi Univ Technol, Dept Elect & Elect Engn, Aichi 4418580, Japan
[2] Toyohashi Univ Technol, Dept Prod Syst Engn, Aichi 4418580, Japan
[3] Natl Inst Adv Ind Sci & Technol, Tsukuba, Ibaraki 3058566, Japan
基金
日本科学技术振兴机构;
关键词
Microtube; Microneedle; Array; Extracellular recording; Low invasiveness; High signal-to-noise ratio; METAL-OXIDE-SEMICONDUCTOR; MICROELECTRODE ARRAYS; BRAIN-TISSUE; SI(111); GROWTH; IMPROVEMENT; INTERFACES; DEVICES;
D O I
10.1007/s10544-009-9356-y
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We report on the development of a microtube electrode array as a neural interface device. To combine the desired properties for the neural interface device, such as low invasiveness with a small needle and a good signal-to-noise ratio in neural recordings, we applied the structure of a glass pipette electrode to each microtube electrode. The device was fabricated as sub-5-mu m-diameter out-of-plane silicon dioxide microtube arrays using silicon microneedle templates, which are grown by the selective vapor-liquid-solid method. The microtubes had inner diameters of 1.9-6.4 mu m and a length of 25 mu m. Impedances ranged from 220 k Omega to 1.55 M Omega, which are less than those for conventional microneedles. In addition, the microtube electrodes had less signal attenuation than conventional microneedle electrodes. We confirmed that the effects of parasitic capacitances between neighboring microtubes and channels were sufficiently small using a test signal. Finally, neural responses evoked from a rat peripheral nerve were recorded in vivo using a microtube electrode to confirm that this type of electrode can be used for both electrophysiological measurements and as a neural interface device.
引用
收藏
页码:41 / 48
页数:8
相关论文
共 50 条
  • [21] ENHANCEMENT OF THE SIGNAL-TO-NOISE RATIO DURING HOLOGRAPHIC RECORDING IN SILLENITE
    TONTCHEV, D
    JIVKOVA, ST
    OPTICS LETTERS, 1992, 17 (23) : 1715 - 1717
  • [22] MEASUREMENT OF THE INTRINSIC SIGNAL-TO-NOISE RATIO FOR HIGH-PERFORMANCE RIGID RECORDING MEDIA
    BELK, NR
    GEORGE, PK
    MOWRY, GS
    JOURNAL OF APPLIED PHYSICS, 1986, 59 (02) : 557 - 563
  • [23] MUTUAL COUPLING IN SIGNAL-TO-NOISE RATIO OPTIMIZATION OF ANTENNA ARRAYS
    DELISLE, GY
    CUMMINS, JA
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 1973, EM15 (02) : 38 - 44
  • [24] OPTIMIZATION OF SIGNAL-TO-NOISE RATIO AND GAIN OF ARBITRARY ANTENNA ARRAYS
    LO, YT
    LEE, SW
    PROCEEDINGS OF THE INSTITUTE OF ELECTRICAL AND ELECTRONICS ENGINEERS, 1965, 53 (06): : 655 - &
  • [25] OPTIMAL SIGNAL-TO-NOISE RATIO FOR LINEAR ARRAYS BY SCHWARTZ INEQUALITY
    KRITIKOS, HN
    JOURNAL OF THE FRANKLIN INSTITUTE-ENGINEERING AND APPLIED MATHEMATICS, 1963, 276 (04): : 295 - &
  • [26] An innovative detection method of high frequency BPSK signal with low signal-to-noise ratio
    Shi, Shuo
    Xu, Li-Zhen
    Gu, Xue-Mai
    Zhang, Hong-Li
    Journal of Harbin Institute of Technology (New Series), 2012, 19 (06) : 93 - 99
  • [27] Optimal Signal Discrimination in a Low Signal-to-Noise Ratio Environment
    Ciodaro, Thiago
    2011 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS), 2011, : 1085 - 1088
  • [28] An innovative detection method of high frequency BPSK signal with low signal-to-noise ratio
    石硕
    徐立振
    顾学迈
    张宏莉
    Journal of Harbin Institute of Technology(New series), 2012, (06) : 93 - 99
  • [29] Improving the signal-to-noise ratio for heat-assisted magnetic recording by optimizing a high/low Tc bilayer structure
    Muthsam, O.
    Slanovc, F.
    Vogler, C.
    Suess, D.
    JOURNAL OF APPLIED PHYSICS, 2019, 126 (12)
  • [30] Nano Stripe Electrode of Pyrolyzed Polymer for High Signal-to-Noise Ratio Electrochemical Detector
    Naka, Keisuke
    Hayashi, Hiroki
    Senda, Mitsugi
    Shiraishi, Haruki
    Konishi, Satoshi
    JAPANESE JOURNAL OF APPLIED PHYSICS, 2009, 48 (03) : 037001