共 50 条
Selective and Sensitive Chemiluminescence Determination of MCPB: Flow Injection and Liquid Chromatography
被引:5
|作者:
Meseguer-Lloret, Susana
[1
]
Torres-Cartas, Sagrario
[1
]
Catala-Icardo, Monica
[1
]
Gomez-Benito, Carmen
[1
]
机构:
[1] Univ Politecn Valencia, Inst Invest Gest Integrada Zonas Costeras, E-46022 Valencia, Spain
关键词:
MCPB;
Flow injection;
Liquid chromatography;
High-performance liquid chromatography;
HPLC;
Chemiluminescence;
Photoreactor;
Solid phase extraction;
SPE;
SOLID-PHASE EXTRACTION;
CHLOROPHENOXY ACID HERBICIDES;
DIODE-ARRAY DETECTION;
GC-MS ANALYSIS;
PHOTOINDUCED CHEMILUMINESCENCE;
WATER SAMPLES;
POSTCOLUMN CHEMILUMINESCENCE;
PHENOXYACID HERBICIDES;
ENVIRONMENTAL WATERS;
AQUEOUS SAMPLES;
D O I:
10.1177/0003702815620133
中图分类号:
TH7 [仪器、仪表];
学科分类号:
0804 ;
080401 ;
081102 ;
摘要:
Two new chemiluminescence (CL) methods are described for the determination of the herbicide 4-(4-chloro-o-tolyloxy) butyric acid (MCPB). First, a flow injection chemiluminescence (FI-CL) method is proposed. In this method, MCPB is photodegraded with an ultraviolet (UV) lamp and the photoproducts formed provide a great CL signal when they react with ferricyanide in basic medium. Second, a high-performance liquid chromatography chemiluminescence (HPLC-CL) method is proposed. In this method, before the photodegradation and CL reaction, the MCPB and other phenoxyacid herbicides are separated in a C18 column. The experimental conditions for the FI-CL and HPLC-CL methods are optimized. Both methods present good sensitivity, the detection limits being 0.12 mu g L-1 and 0.1 mu g L-1 (for FI-CL and HPLC-CL, respectively) when solid phase extraction (SPE) is applied. Intra-and interday relative standard deviations are below 9.9%. The methods have been satisfactorily applied to the analysis of natural water samples. FI-CL method can be employed for the determination of MCPB in simple water samples and for the screening of complex water samples in a fast, economic, and simple way. The HPLC-CL method is more selective, and allows samples that have not been resolved with the FI-CL method to be solved.
引用
收藏
页码:312 / 321
页数:10
相关论文