Non-invasive characterization and quality assurance of silicon micro-strip detectors using pulsed infrared laser

被引:0
|
作者
Ghosh, P. [1 ,2 ,3 ]
机构
[1] Goethe Univ Frankfurt, Inst Kernphys IKP, Max von Laue Str 1, D-60438 Frankfurt, Germany
[2] GSI Helmholtzzentrum Schwerionenforsch GmbH, Planckstr 1, D-64291 Darmstadt, Germany
[3] Facil Antiproton & Ion Res Europe GmbH, Planckstr 1, D-64291 Darmstadt, Germany
来源
关键词
Interaction of radiation with matter; Radiation-hard detectors; Detection of defects; Charge induction;
D O I
10.1088/1748-0221/11/01/C01019
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The Compressed Baryonic Matter (CBM) experiment at FAIR is composed of 8 tracking stations consisting of roughly 1300 double sided silicon micro-strip detectors of 3 different dimensions. For the quality assurance of prototype micro-strip detectors a non-invasive detector charaterization is developed. The test system is using a pulsed infrared laser for charge injection and characterization, called Laser Test System (LTS). The system is aimed to develop a set of characterization procedures which are non-invasive (non-destructive) in nature and could be used for quality assurances of several silicon micro-strip detectors in an efficient, reliable and reproducible way. The procedures developed (as reported here) uses the LTS to scan sensors with a pulsed infra-red laser driven by step motor to determine the charge sharing in-between strips and to measure qualitative uniformity of the sensor response over the whole active area. The prototype detector modules which are tested with the LTS so far have 1024 strips with a pitch of 58 mu m on each side. They are read-out using a self-triggering prototype read-out electronic ASIC called n-XYTER. The LTS is designed to measure sensor response in an automatized procedure at several thousand positions across the sensor with focused infra-red laser light (spot size approximate to 12 mu m, wavelength = 1060 nm). The pulse with a duration of approximate to 10 ns and power approximate to 5mW of the laser pulse is selected such, that the absorption of the laser light in the 300 mu m thick silicon sensor produces approximate to 24000 electrons, which is similar to the charge created by minimum ionizing particles (MIP) in these sensors. The laser scans different prototype sensors and various non-invasive techniques to determine characteristics of the detector modules for the quality assurance is reported.
引用
收藏
页数:9
相关论文
共 34 条
  • [21] Non-invasive histologic characterization of skin diseases using in vivo confocal scanning laser microscopy.
    González, S
    Rajadhyaksha, M
    Rubinstein, G
    White, WM
    González, E
    Anderson, RR
    JOURNAL OF INVESTIGATIVE DERMATOLOGY, 1998, 110 (04) : 580 - 580
  • [22] Non-invasive prediction of mango quality using near-infrared spectroscopy: Assessment on spectral interferences of different packaging materials
    Ding, Fangchen
    Zuo, Changzhou
    Garcia-Martin, Juan Francisco
    Ge, Yan
    Tu, Kang
    Peng, Jing
    Xiao, Hongmei
    Lan, Weijie
    Pan, Leiqing
    JOURNAL OF FOOD ENGINEERING, 2023, 357
  • [23] Principal component analysis-assisted energy dispersive X-ray fluorescence spectroscopy for non-invasive quality assurance characterization of complex matrix materials
    Angeyo, K. H.
    Gari, S.
    Mangala, J. M.
    Mustapha, A. O.
    X-RAY SPECTROMETRY, 2012, 41 (05) : 321 - 327
  • [24] Measurement of blood flow velocity in a model of stenosis in vitro and in mesenteric vessels in vivo using non-invasive micro multipoint laser Doppler velocimetry
    Akiguchi, Shunsuke
    Ishida, Hiroki
    Andoh, Tsugunobu
    Hachiga, Tadashi
    Shimizu, Tadamichi
    Kuraishi, Yasushi
    Shirakawa, Hiroki
    Ueyama, Koji
    MEASUREMENT SCIENCE AND TECHNOLOGY, 2012, 23 (04)
  • [25] Validation of non-invasive fetal sex determination using cell free fetal DNA analysis and development of a fetal RASSF1A assay for quality assurance
    Evans, Elaine
    Warner, J.
    Porteous, M.
    Maher, E.
    Delaney, C. A.
    JOURNAL OF MEDICAL GENETICS, 2012, 49 : S98 - S98
  • [26] Characterization and Diagnostics of Active Phased Array Modules Using Non-Invasive Electro-Optic Field Probes with a CW Laser Source
    Sabet, Kazem
    Darragh, Richard
    Sabet, Ali
    Sarabandi, Kamal
    Jamil, Khalid
    Alhumaidi, Sami
    2017 IEEE MTT-S INTERNATIONAL MICROWAVE SYMPOSIUM (IMS), 2017, : 1617 - 1619
  • [27] Rapid and non-invasive detection of high-thickness glucose solution concentrations using quantum cascade laser-based transmission infrared spectroscopy
    Chen, Junyi
    Furukawa, Hiromitsu
    INFRARED PHYSICS & TECHNOLOGY, 2023, 131
  • [28] Microstructural characterization of deep-fat fried formulated products using confocal scanning laser microscopy and a non-invasive double staining procedure
    Carolina Moreno, Maria
    Bouchon, Pedro
    JOURNAL OF FOOD ENGINEERING, 2013, 118 (02) : 238 - 246
  • [29] Non-invasive treatment of the rabbit carotid artery atherothrombotic occlusion using combined Nd:YAG laser and pulsed- focused ultrasound accompanied by thrombolytic agent administration
    Mehrad, H.
    Akrami, A.
    ATHEROSCLEROSIS, 2016, 252 : E26 - E26
  • [30] Development of non-invasive Breath Analyzer : A proposed system for detection of Nitric Oxide (NO) in exhaled human breath, using InfraRed Quantum Cascade Laser (QCL) absorption technique
    Arekar, Sonali
    Daruwala, R. D.
    2015 INTERNATIONAL CONFERENCE ON NASCENT TECHNOLOGIES IN THE ENGINEERING FIELD (ICNTE), 2015,