A Method to Improve the Accuracy of the Bioelectrical Impedance Phase-Angle Measuring System

被引:0
|
作者
Nguyen Ngoc Tram [1 ]
Trinh Duc Nam [2 ]
Truong Xuan Thanh [3 ]
Tran Quang Do [4 ]
Dinh Hong Thai [5 ]
Vu Duy Hai [1 ]
机构
[1] Hanoi Univ Sci & Technol, Sch Elect Telecommun, Hanoi, Vietnam
[2] BME HUST, Minist Hlth, Dept Med Equipment & Contruct, Hanoi, Vietnam
[3] BME HUST, Friendship Hosp, Informat Technol Dept, Hanoi, Vietnam
[4] BME HUST, Bach Mai Hosp, Dept Med Equipment & Consumables, Hanoi, Vietnam
[5] BME HUST, Viet Duc Univ Hosp, Dept Med Equipment & Consumables, Hanoi, Vietnam
关键词
Bioelectrical impedance; phase-angle; nutritional assessment; voltage to current converter; comparator; phase lock loop; phase-difference detector; BODY-COMPOSITION; NUTRITIONAL-STATUS; CANCER; MALNUTRITION; MASS;
D O I
10.1109/ACOMP53746.2021.00030
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
The bioelectric impedance phase difference (BIPA) is a characteristic quantity for the changes in the electrical properties of the tissue, obtained from the Bioelectrical Impedance Analysis (BIA). Nowadays, the need for an automatic, specialized and safe device to measure BIPA in medical facilities is very pressing. One method that has been proposed to measure this parameter is to use a multiplier and calculate the value of the phase difference between the two signals. The bias (mean) (+/- SD) of the measurement of the phase difference between two signals with programmed phases using this method was 0.5 +/- 0.2. Consequently, to improve the accuracy of the measurement results, this study was conducted to propose another signal processing method which is based on the comparator and the operating principle of the phase lock loop for the BIPA measurement system. This system still consists of an analog signal generator at 100kHz, a current source circuit. However, the signal processors which process signal received from the body consist of an analog-to-pulse converter and a phase-difference detector. The phase angle between the two signals is then calculated from the DC voltage obtained at the output of the phase-difference detector and compared with the programmed one. The phase difference between the two programmed signals ranges from 5 degrees to 8 degrees with an increment of 0.5 degrees. The obtained results show that the bias of the measurement performed by the system using the proposed method is -0.01, the precision was 0.07 and percentage error was 4.31%.
引用
收藏
页码:169 / 175
页数:7
相关论文
共 50 条
  • [31] Bioelectrical impedance phase angle as a prognostic marker in chronic heart failure
    Colin-Ramirez, Eloisa
    Castillo-Martinez, Lilia
    Orea-Tejeda, Arturo
    Vazquez-Duran, Marisela
    Rodriguez, Ana E.
    Keirns-Davis, Candace
    NUTRITION, 2012, 28 (09) : 901 - 905
  • [32] Bioelectrical impedance analysis (BIA) and phase angle: how relevant in cancer?
    Almeida, Ana Isabel
    Ferreira, Catarina
    Grillo, Isabel Monteiro
    Camilo, Maria
    Ravasco, Paula
    FASEB JOURNAL, 2013, 27
  • [33] Phase Angle and Bioelectrical Impedance Vectors in Adolescent and Adult Male Athletes
    Koury, Josely C.
    Trugo, Nadia M. F.
    Torres, Alexandre G.
    INTERNATIONAL JOURNAL OF SPORTS PHYSIOLOGY AND PERFORMANCE, 2014, 9 (05) : 798 - 804
  • [34] BIOELECTRICAL IMPEDANCE DERIVED PHASE ANGLE AS MORTALITY PREDICTOR IN HEMODIALYSIS PATIENTS
    Gincaite, Alvita
    Vicka, Vaidas
    Sukackiene, Diana
    Pavinic, Jelena
    Laucyte-Cibulskiene, Agne
    Rimsevicius, Laurynas
    Miglinas, Marius
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2017, 32 : 704 - 704
  • [35] Re. "Bioelectrical impedance vector analysis, phase-angle assessment and relationship with malnutrition risk in a cohort of frail older hospital patients in the United Kingdom"
    Alicandro, Gianfranco
    NUTRITION, 2015, 31 (02) : 416 - 417
  • [36] PHASE-ANGLE MEASURING DEVICES ERRORS DUE TO REPEATED SEQUENTIAL DISPLACEMENT OF FREQUENCY
    AKHMADOV, AA
    PRONENKO, VI
    MEASUREMENT TECHNIQUES, 1973, 16 (04) : 542 - 544
  • [37] HYPERACUITY FOR LUMINANCE PHASE-ANGLE IN THE HUMAN VISUAL-SYSTEM
    MORGAN, MJ
    WATT, RJ
    VISION RESEARCH, 1982, 22 (07) : 863 - 866
  • [38] Technique for the standing phase-angle reduction in power system restoration
    Shahidehpour, SM
    Yamin, HY
    ELECTRIC POWER COMPONENTS AND SYSTEMS, 2005, 33 (03) : 277 - 286
  • [39] GPS SYNCHRONIZED POWER-SYSTEM PHASE-ANGLE MEASUREMENTS
    WILSON, RE
    STERLINA, PS
    INTERNATIONAL JOURNAL OF SATELLITE COMMUNICATIONS, 1994, 12 (05): : 499 - 505
  • [40] A Simple Rapid Method for Measuring Liver Steatosis Using Bioelectrical Impedance
    Yoshimoto-Haramura, Tomoko
    Hara, Takanobu
    Soyama, Akihiko
    Kugiyama, Tota
    Matsushima, Hajime
    Matsuguma, Kunihito
    Imamura, Hajime
    Tanaka, Takayuki
    Adachi, Tomohiko
    Hidaka, Masaaki
    Okabe, Shuichi
    Murata, Masakazu
    Eguchi, Susumu
    IN VIVO, 2022, 36 (02): : 570 - 575