Optimization of conformal thoracic radiotherapy using cone-beam CT imaging for treatment verification

被引:35
|
作者
Sidhu, K
Ford, EC
Spirou, S
Yorke, E
Chang, JW
Mueller, K
Todor, D
Rosenzweig, K
Mageras, G
Chui, C
Ling, CC
Amols, H
机构
[1] Mem Sloan Kettering Canc Ctr, Dept Med Phys, New York, NY 10021 USA
[2] Mem Sloan Kettering Canc Ctr, Dept Radiat Oncol, New York, NY 10021 USA
[3] SUNY Stony Brook, Dept Comp Sci, Stony Brook, NY 11794 USA
来源
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS | 2003年 / 55卷 / 03期
关键词
megavoltage CT; cone-beam CT; treatment verification; radiotherapy planning; electronic portal imaging device;
D O I
10.1016/S0360-3016(02)04152-4
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: Megavoltage cone-beam computed tomography (MVCBCT) has been proposed for treatment verification in conformal radiotherapy. However, the doses required for such imaging may compromise the quality of the delivered dose distribution. The present paper explores the effect of cone-beam imaging on dose homogeneity and critical organ dose and the use of our new tool, adapted intensity-modulated radiation therapy (AIMRT). Methods and Materials: Three types of treatment plans were devised (3D-CRT [three-dimensional conformal radiotherapy], IMRT [intensity-modulated radiotherapy], and AIMRT) based on 4 patients with thoracic malignancies. MVCBCT fields were then integrated into the plans. The MVCBCT technique used 21 imaging portals at 10degrees intervals. The MVCBCT apertures were shaped to conform to the planning target volume with a 6-mm margin. In a second set of plans, the field size was expanded by a further 2 cm. The unoptimized MVCBCT dose distribution was incorporated into the IMRT plan using AIMRT. Results: Normal-tissue complication probability with MVCBCT is acceptable for all plans at the 66.6 Gy level, but exceeds tolerance for both 3D-CRT alone and 3D-CRT with MVCBCT at higher doses. In contrast, the use of AIMRT planning with MVCBCT allowed safe dose escalation to 85 Gy. Expanding the MVCBCT aperture provided better anatomic visibility with an acceptable lung dose. The results using IMRT with MVCBCT fell between the values measured for 3D-CRT and AIMRT with MVCBCT. Conclusion: The present study is the first to demonstrate that MVCBCT can be incorporated into 3D-CRT and IMRT planning with minimal effect on planning target volume homogeneity and dose to critical structures. This paves the way for highly conformal radiotherapy at greater doses delivered with increased confidence and safety. (C) 2003 Elsevier Science Inc.
引用
收藏
页码:757 / 767
页数:11
相关论文
共 50 条
  • [1] A cone-beam megavoltage CT scanner for treatment verification in conformal radiotherapy
    Mosleh-Shirazi, MA
    Evans, PM
    Swindell, W
    Webb, S
    Partridge, M
    RADIOTHERAPY AND ONCOLOGY, 1998, 48 (03) : 319 - 328
  • [2] ESTIMATING THE RISKS ASSOCIATED WITH BREAST RADIOTHERAPY AND MEGAVOLTAGE CONE-BEAM CT IMAGING VERIFICATION
    Quinn, A.
    Holloway, L. C.
    Koh, E. S.
    Arumugam, S.
    Batumalai, V.
    Nelson, V.
    Owen, K.
    Goozee, G.
    Vial, P.
    Metcalfe, P.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S236 - S236
  • [3] Cone-beam CT for radiotherapy applications
    Cho, P.S.
    Johnson, R.H.
    Griffin, T.W.
    1863, Institute of Physics Publishing Ltd, Bristol, Engl (40):
  • [4] CONE-BEAM CT FOR RADIOTHERAPY APPLICATIONS
    CHO, PS
    JOHNSON, RH
    GRIFFIN, TW
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (11): : 1863 - 1883
  • [5] Personalized organ dose assessment of cone-beam CT imaging for image-guided thoracic radiotherapy
    Zhang, Y.
    Wu, H.
    Chen, Z.
    Knisely, J.
    Nath, R.
    Feng, Z.
    Bao, S.
    Deng, J.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S266 - S266
  • [6] Respiratory gated cone-beam CT volumetric imaging for external beam radiotherapy
    Moseley, DJ
    Hawkins, M
    Eccles, C
    Euler, C
    White, EA
    Bissonnette, J
    Dawson, LA
    Jaffray, DA
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S27 - S28
  • [7] Aspects of simulator cone-beam CT for radiotherapy treatment planning
    Irvine, Denise
    McJury, Mark
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2010, 9 (03) : 165 - 174
  • [8] Selection of lung cancer patients for adaptive radiotherapy using cone-beam CT imaging
    Hattu, D.
    RADIOTHERAPY AND ONCOLOGY, 2019, 133 : S177 - S177
  • [9] CONE-BEAM CT VERIFICATION FOR PROSTATE RADIOTHERAPY: MANUAL VERSUS AUTOMATIC REGISTRATION
    Rosenfelder, N.
    Aitken, A.
    van As, N.
    Khoo, V.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S398 - S399
  • [10] The Use of Cone Beam CT for Treatment Verification in Prostate Radiotherapy
    Khaira, M. K.
    Huang, V. W.
    Liu, M.
    Cheung, A. K.
    Laberge, T.
    Kwan, W.
    CLINICAL ONCOLOGY, 2011, 23 (03) : S42 - S42