Bend relief fenestration might prevent outflow graft obstruction in patients with left ventricular assist device

被引:4
|
作者
Faerber, Gloria [1 ]
Kirov, Hristo [1 ]
Schwan, Imke [1 ]
Graeger, Stephanie [2 ]
Diab, Mahmoud [1 ]
Tkebuchava, Sophie [1 ]
Doenst, Torsten [1 ]
机构
[1] Friedrich Schiller Univ, Jena Univ Hosp, Dept Cardiothorac Surg, Klinikum 1, D-07747 Jena, Germany
[2] Friedrich Schiller Univ, Jena Univ Hosp, Dept Radiol, Jena, Germany
关键词
Left ventricular assist device; External outflow graft obstruction; SOCIETY;
D O I
10.1093/icvts/ivac149
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES: Patients with left ventricular assist devices may experience external obstruction of the outflow graft through a gelatinous substance within the bend relief (BR; a stiff tube graft guiding the outflow graft). Preventative strategies have been missing. Having faced this problem, we decided to fenestrate the BR to avoid outflow graft obstruction (OGO). METHODS: Since December 2010, 167 patients underwent left ventricular assist device implantation using HeartMate II or 3. BR fenestration was introduced on July 2018 (108 patients before, 59 after the introduction of BR fenestration). Follow-up computed tomography scans were obtained from all patients and were screened for OGO by 3 independent investigators. Results were correlated with log file history, echocardiographic and clinical outcomes. RESULTS: Demographic data were comparable between groups, with mostly male patients. Patients with BR fenestration were older [63 (standard deviation (SD):10.6) vs 58 (SD: 10.7) years] and had shorter support duration [494 (SD: 383) vs 951 (SD: 875) days]. OGO was observed in 5 patients and occurred only in patients without fenestration. Importantly, it occurred late on postoperative Days 412, 462, 1043, 1184 and 1506. Three patients are still asymptomatic. Surgical revision was required in the other 2 patients for pump thrombosis or continuous low flow. One of them died 36 days after revision due to right heart failure. CONCLUSIONS: Our results suggest that fenestration of the BR may be a preventative strategy to avoid external OGO. OGO occurred late, which suggests a careful long-term follow-up.
引用
收藏
页数:6
相关论文
共 50 条
  • [31] Outflow Graft Occlusion of the HeartMate 3 Left Ventricular Assist Device
    Posada, Juan G. Duero
    Moayedi, Yasbanoo
    Alhussein, Mosaad
    Rodger, Marnie
    Alvarez, Juglans
    Wintersperger, Bernd J.
    Ross, Heather J.
    Butany, Jagdish
    Billia, Filio
    Rao, Vivek
    CIRCULATION-HEART FAILURE, 2017, 10 (09)
  • [32] Outflow Graft Narrowing of the HeartMate 3 Left Ventricular Assist Device
    Jain, Sneha S.
    Clerkin, Kevin J.
    Anstey, Edmund
    Liu, Qi
    Fried, Justin A.
    Raikhelkar, Jayant
    Griffin, Jan M.
    Marshall, Dylan
    Colombo, Paolo
    Yuzefpolskaya, Melana
    Topkara, Veli
    Naka, Yoshifumi
    Takeda, Koji
    Sayer, Gabriel
    Uriel, Nir
    Leb, Jay
    ANNALS OF THORACIC SURGERY, 2023, 115 (05): : 1282 - 1288
  • [33] Endovascular management of left ventricular assist device outflow graft stenosis
    Robinson, Emilie C.
    Mudy, Karol
    Mirza, Aleem K.
    JOURNAL OF CARDIAC SURGERY, 2022, 37 (09) : 2894 - 2896
  • [34] Retained outflow graft following the explantation of left ventricular assist device
    Ornek, Ender
    Kanal, Yucel
    Dundar, Cihan
    Kundi, Harun
    Kervan, Umit
    ANATOLIAN JOURNAL OF CARDIOLOGY, 2020, 24 (05): : E16 - E17
  • [35] A Clinicopathological Case of Left Ventricular Assist Device Outflow Graft Stenosis
    Sener, Yusuf Ziya
    van der Heiden, Cornelis W.
    Sjatskig, Jelena
    von der Thusen, Jan
    Caliskan, Kadir
    ASAIO JOURNAL, 2024, 70 (11) : e162 - e164
  • [36] Endovascular repair of left ventricular assist device outflow graft defect
    Teman, Nicholas R.
    Shah, Shawn M.
    Downs, Emily
    Ailawadi, Gorav
    Mazimba, Sula
    Fritz Angle, John
    Khaja, Minhaj S.
    Yarboro, Leora T.
    JOURNAL OF CARDIAC SURGERY, 2020, 35 (11) : 3235 - 3238
  • [37] Left ventricular outflow tract obstruction associated with chronic ventricular assist device support
    Baradarian, S
    Dembitsky, WP
    Jaski, B
    Abolhoda, A
    Adamson, R
    Chillcot, S
    Daily, PO
    ASAIO JOURNAL, 2002, 48 (06) : 665 - 667
  • [38] A Multi-Center Evaluation of Outflow Graft Obstruction with a Fully Magnetically Levitated Left Ventricular Assist Device
    Wert, L.
    Stewart, G. C.
    Mehra, M. R.
    Milwidsky, A.
    Jorde, U. P.
    Goldstein, D. J.
    Selzman, C. H.
    Stehlik, J.
    Alshamdin, F. D.
    Khaliel, F. H.
    Gustafsson, F.
    Boschi, S.
    Loforte, A.
    Ajello, S.
    Scandroglio, A. M.
    Tucanova, Z.
    Netuka, I.
    Schloglhofer, T.
    Zimpfer, D.
    Dogan, G.
    Schmitto, J. D.
    Maier, S.
    Schibilsky, D.
    Jawad, K.
    Saeed, D.
    Faerber, G.
    Morshuis, M.
    Hanuna, M.
    Muller, C. S.
    Mulzer, J.
    Kempfert, J.
    Falk, V.
    Potapov, E. V.
    JOURNAL OF HEART AND LUNG TRANSPLANTATION, 2022, 41 (04): : S24 - S25
  • [39] Left Ventricular Assist Device Outflow Graft Obstruction: A Complication Specific to Polytetrafluoroethylene Covering. A Word of Caution!
    Alnabelsi, Talal
    Shafii, Alexis E.
    Gurley, John C.
    Dulnuan, Kenneth
    Harris, Dwight D.
    Guglin, Maya
    ASAIO Journal, 2019, 65 (06):
  • [40] Left ventricular assist device outflow graft obstruction: A complication specific to polytetrafluoroethylene covering. A word of caution!
    Alnabelsi, Talal
    Shafii, Alexis E.
    Gurley, John C.
    Dulnuan, Kenneth
    Harris, Dwight D.
    Guglin, Maya
    ASAIO Journal, 2019, 65 (05)