Personnel exposure to scattered radiation during radiography of the distal interphalangeal joint in the horse using a portable X-ray machine

被引:1
|
作者
Sterc, J. [1 ]
Lepkova, R. [1 ]
机构
[1] Univ Vet & Pharmaceut Sci, Brno, Czech Republic
关键词
radiation exposure; radiation dosage; pre-purchase examination; radiological examination; scatter radiation;
D O I
10.2754/avb200776010105
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Radiography of the distal interphalangeal joints of two limbs of a cadaver of a horse weighing 550 kg was conducted in the present study. The examination was performed on lateromedial, palmaroproximal-palmarodistal views and on dorsoproximal-palmarodistal, dorsolateral-palmaromedial, dorsomedial-palmarolateral views of a raised limb placed on a navicular block, and of weight-bearing limbs. During the examination, doses of scattered radiation were measured at the sites of radiographer, assistant holding a film cassette and assistant positioning the examined limb or the opposite one. The lowest radiation dose was received by the assistant holding the cassette behind the examined region; a total equivalent dose of 183.6 - 201.7 nSv was received by his hands and genitals, and 110.8 - 113.3 nSv by his eye tens and neck during the examination of the distal interphalangeal joint. The radiographer was exposed to higher radiation; an equivalent dose of 846.6 - 854.1 nSv was received by his hands and genitals, and 271.8 - 328.2 nSv by the eye lens and the neck. The highest scattered radiation dose was received by the assistant positioning the examined limbs; equivalent doses were 7751 - 9354 nSv (hands), 1117.3 - 1119.5 nSv (genitals), and 880.6 - 1096.2 nSv (eye lens and neck). The equivalent dose values measured, received by the radiographer and the assistants during the examination of the distal interphalangeal joint seem to be very low relative to radiation exposure limits. However, it must be taken into account that the personnel involved in radiography is also exposed to scattered radiation during other examinations where radiation doses are often much higher. These results indicate the necessity to use protective lead aprons, gloves and collars during radiography of the distal interphalangeal joint.
引用
收藏
页码:105 / +
页数:11
相关论文
共 50 条
  • [41] Material analysis with a multiple X-ray tomography scanner using transmitted and scattered radiation
    Cesareo, R
    Brunetti, A
    Golosio, B
    Lopes, RT
    Barroso, RC
    Castellano, S
    Quarta, S
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2004, 525 (1-2): : 336 - 341
  • [42] Neutron radiography characteristics of transfer exposure method using X-ray imaging plate
    Uno H.
    Uemoto R.
    Nittoh K.
    Sogabe S.
    Sonoda Y.
    Kawashima M.
    Transactions of the Atomic Energy Society of Japan, 2018, 17 (02) : 33 - 41
  • [43] SCATTERED RADIATION EXPOSURE RATES FROM VARIOUS PHANTOM MATERIALS IRRADIATED AT DIAGNOSTIC X-RAY ENERGIES
    CHAMBERLAIN, CC
    BASSANO, DA
    MOYER, RF
    HEALTH PHYSICS, 1976, 31 (06): : 557 - 557
  • [44] Scattered radiation doses absorbed by technicians at different distances from X-ray exposure: Experiments on prosthesis
    Chiang, Hsien-Wen
    Liu, Ya-Ling
    Chen, Tou-Rong
    Chen, Chun-Lon
    Chiang, Hsien-Jen
    Chao, Shin-Yu
    BIO-MEDICAL MATERIALS AND ENGINEERING, 2015, 26 : S1641 - S1650
  • [45] Correction of the Contribution of Scattered Photon Radiation to the Ionization Chamber Readings During X-Ray Radiation Quality Assessment
    Zaharadniuk, A. A.
    Senkovsky, K. G.
    Lukashevich, R. V.
    DEVICES AND METHODS OF MEASUREMENTS, 2022, 13 (03): : 180 - 188
  • [46] Assessment of scattered and leakage radiation from ultra-portable X-ray systems in chest imaging: An independent study
    Paulis, Leonie E.
    Schnerr, Roald S.
    Halton, Jarred
    Qin, Zhi Zhen
    Chua, Arlene
    PLOS GLOBAL PUBLIC HEALTH, 2025, 5 (01):
  • [47] Precision and accuracy of stress measurement with a portable X-ray machine using an area detector
    Lee, Seung-Yub
    Ling, Jinjing
    Wang, Shenghe
    Ramirez-Rico, Joaquin
    JOURNAL OF APPLIED CRYSTALLOGRAPHY, 2017, 50 : 131 - 144
  • [48] Evaluation of the radiation dose to a phantom for various X-ray exposure factors performed using the dose area product in digital radiography
    Kweon, D. C.
    Chung, W. K.
    Dong, K. R.
    Lee, J. W.
    Choi, J. W.
    Goo, E. H.
    Lee, J. S.
    Kim, S. G.
    Cho, J. H.
    Chung, J. E.
    RADIATION EFFECTS AND DEFECTS IN SOLIDS, 2012, 167 (12): : 954 - 970
  • [49] RADIATION EXPOSURE DURING X-RAY EXAMINATIONS IN A LARGE PAEDIATRIC HOSPITAL IN SERBIA
    Ciraj-Bjelac, Olivera
    Gavrilovic, Marijana
    Arandjic, Danijela
    Vujovic, Milan
    Bozovic, Predrag
    RADIATION PROTECTION DOSIMETRY, 2015, 165 (1-4) : 220 - 225
  • [50] INVITED GUEST EDITORIAL IN RESPONSE TO: ESTIMATED OPERATOR EXPOSURE FOR HAND HOLDING PORTABLE X-RAY UNITS DURING IMAGING OF THE EQUINE DISTAL EXTREMITY
    Barbee, Dave
    VETERINARY RADIOLOGY & ULTRASOUND, 2011, 52 (02) : 213 - 214