Hemodynamic study of overlapping bare-metal stents intervention to aortic aneurysm

被引:27
|
作者
Zhang, Peng [1 ]
Sun, Anqiang [1 ]
Zhan, Fan [1 ]
Luan, Jingyuan [2 ]
Deng, Xiaoyan [1 ]
机构
[1] Beihang Univ, Sch Biol Sci & Med Engn, Minist Educ, Key Lab Biomech & Mechanobiol, Beijing 100191, Peoples R China
[2] Peking Univ, Dept Intervent Radiol & Vasc Surg, Hosp 3, Beijing 100191, Peoples R China
基金
中国国家自然科学基金;
关键词
Thoracic aortic aneurysm; Overlapping bare-metal stents; Hemodynamics; Numerical simulation; ENDOVASCULAR REPAIR; MULTILAYER STENT; THORACOABDOMINAL ANEURYSMS; INTERACTION SIMULATION; ARTERY ANEURYSM; FLOW PATTERNS; ENDOLEAKS; PERFORMANCE; OUTCOMES; STRESS;
D O I
10.1016/j.jbiomech.2014.08.025
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
To investigate the hemodynamic performance of overlapping bare-metal stents intervention treatment to thoracic aortic aneurysms (TAA), three simplified TAA models, representing, no stent, with a single stent and 2 overlapped stents deployed in the aneurismal sac, were studied and compared in terms of flow velocity, wall shear stress (WSS) and pressure distributions by means of computational fluid dynamics. The results showed that overlapping stents intervention induced a flow field of slow velocity near the aneurismal wall. Single stent deployment in the sac reduced the jet-like flow formed prior to the proximal neck of the aneurysm, which impinged on the internal wall of the aneurysm. This jet-like flow vanished completely in the overlapping double stents case. Overlapping stents intervention led to an evident decrease in WSS; meanwhile, the pressure acting on the wall of the aneurysm was reduced slightly and presented more uniform distribution. The results therefore indicated that overlapping stents intervention may effectively isolate the thoracic aortic aneurysm, protecting it from rupture. In conclusion, overlapping bare-metal stents may serve a purpose similar to that of the multilayer aneurysm repair system (MARS) manufactured by Cardiatis SA (Isnes, Belgium). (C) 2014 Elsevier Ltd. All rights reserved.
引用
收藏
页码:3524 / 3530
页数:7
相关论文
共 50 条
  • [1] Hemodynamic insight into overlapping bare-metal stents strategy in the treatment of aortic aneurysm
    Zhang, Peng
    Liu, Xiao
    Sun, Anqiang
    Fan, Yubo
    Deng, Xiaoyan
    JOURNAL OF BIOMECHANICS, 2015, 48 (10) : 2041 - 2046
  • [2] Overlapping Bare-Metal Stents Strategy in Treatment of Thoracic Aortic Aneurysm: A Numerical Study
    Zhang, P.
    Sun, A. Q.
    Zhan, F.
    Deng, X. Y.
    1ST GLOBAL CONFERENCE ON BIOMEDICAL ENGINEERING & 9TH ASIAN-PACIFIC CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING, 2015, 47 : 53 - 56
  • [3] Covered stents versus bare-metal stents in kissing reconstructions of the aortic bifurcation
    D'Ospina, Rita M.
    Vento, Vincenzo
    Ghariani, Mohamed Z.
    Kuntz, Salome
    Schwein, Adeline
    Sonetto, Alessia
    Lejay, Anne
    Bernardini, Giulia
    Freyrie, Antonio
    Gargiulo, Mauro
    Chakfe, Nabil
    ITALIAN JOURNAL OF VASCULAR AND ENDOVASCULAR SURGERY, 2024, 31 (01): : 1 - 9
  • [4] In Vitro Modelling of Aortic Dissection for Bare-Metal Stent Intervention
    Kobayashi, Mayo
    Shiraishi, Yasuyuki
    Chikweto, Francis
    Fukai, Miharu
    Hou, Xiaoxi
    Fukaya, Aoi
    Ikura, Maiko
    Hanzawa, Kazuhiko
    Yambe, Tomoyuki
    6TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, ICOBE 2023, 2025, 115 : 10 - 18
  • [5] Drug-Eluting Stents Versus Bare-Metal Stents for Treatment of Bare-Metal In-Stent Restenosis
    Singh, Inder M.
    Filby, Steven J.
    El Sakr, Fredy
    Gorodeski, Eiran Z.
    Lincoff, A. Michael
    Ellis, Stephen G.
    Shishehbor, Mehdi H.
    CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2010, 76 (02) : 257 - 262
  • [6] Drug-Eluting vs Bare-Metal Stents for Percutaneous Coronary Intervention
    Rodriguez-Granillo, Alfredo M.
    Fernandez-Pereira, Carlos
    Rodriguez, Alfredo E.
    JAMA INTERNAL MEDICINE, 2021, 181 (07) : 1012 - 1013
  • [7] Drug-Eluting Versus Bare-Metal Stents for Bare-Metal In-Stent Restenosis
    Singh, Inder M.
    Filbly, Steven J.
    El Sakr, Fredy
    Gorodeski, Eiran Z.
    Lincoff, A. Michael
    Ellis, Stephen G.
    Shishehbor, Mehdi H.
    AMERICAN JOURNAL OF CARDIOLOGY, 2009, 104 (6A): : 145D - 145D
  • [8] Outcomes of Overlapping Bare-Metal Stents with Sirolimus-Eluting Stents for Long Lesions in Small Coronary Vessels
    Kronsteiner, Emily
    Lee, Richard W.
    Sweeney, John P.
    Metallo, Jennifer R.
    Stoehr, James D.
    Fortuin, F. David
    JOURNAL OF INVASIVE CARDIOLOGY, 2010, 22 (02): : 76 - 79
  • [9] Drug-eluting or bare-metal stents?
    Irene Fernández-Ruiz
    Nature Reviews Cardiology, 2016, 13 (11) : 631 - 631
  • [10] Clinical outcome after coronary angioplasty with overlapping drug-eluting and bare-metal stents
    Kubler, P.
    Reczuch, K.
    Jankowska, E. A.
    Luciuk, D.
    Banasiak, W.
    Ponikowski, P.
    EUROPEAN HEART JOURNAL, 2010, 31 : 998 - 998