Estimation of HVL in computed tomography using CTDI measurements

被引:0
|
作者
Yang, Z. [1 ]
Chu, R. [1 ]
Ruan, C. [1 ]
机构
[1] Univ Oklahoma, Hlth Sci Ctr, Oklahoma City, OK USA
关键词
D O I
10.1118/1.2760387
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To determine the HVL of CT systems from CTDI dosimetry measurements without additional test, and to overcome the technical difficulties in CT HVL measurements. Method and Materials: The HVL of one multi‐detector CT scanner was measured when the scanner is static, over the available energy range (80∼140 kVp) and with different bow‐tie filter configurations. Commercial CT dose phantoms in compliance with FDA specifications were used to measure the exposures. Nine doses at the central, peripheral, and intermediate positions were measured for both head and body phantoms. Averages were taken for doses measured at axially symmetric positions of each phantom. These averages were compared to the central position dose measurements. The ratios were analyzed along with the HVL values measured under exactly the same conditions and fitted to second power polynomials. Combining the fitting parameters and dosimetry measurements from different machines, the corresponding HVL were predicted and compared to the measured value of each machine. Results: The HVL of the CT scanner were determined to be between 5 to 10 mm Al over the energy range assessed. When the values were plotted against the dose at the peripheral or intermediate position, normalized by center dose, the curves all fit to second power polynomials with R2>0.99. When applying the correlation obtained from the fittings to dosimetry measurements from different machines, the calculated HVLs are within 5% or 0.5 mm Al error of the measured values. Conclusion: This study showed that the data collected for the CTDI measurements can be directly used to estimate the HVL of CT scanners to certain accuracy. This will reduce the time and effort needed for medical physicists to determine the HVL of CT systems during acceptance testing and annual quality assurance surveys. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2340 / 2340
页数:1
相关论文
共 50 条
  • [31] Proposal of a Reference Object in Dimensional Measurements using Computed Tomography
    Novak, Amalija Horvatic
    Runje, Biserka
    Samardzic, Ivan
    Maglic, Leon
    TEHNICKI VJESNIK-TECHNICAL GAZETTE, 2021, 28 (05): : 1735 - 1741
  • [32] Measurements of the coronary blood flow using Computed Tomography (CT)
    Grimme, S
    Krug, B
    Just, T
    Bovenschulte, H
    Paperno, S
    Lackner, K
    Cardio-Visions 2004: JUNGE EXZELLENZE IN DER KARDIOVASKULAREN FORSCHUNG, 2005, : 197 - 199
  • [33] Reproducibility of Lung and Lobar Volume Measurements Using Computed Tomography
    Brown, Matthew S.
    Kim, Hyun J.
    Abtin, Fereidoun
    Da Costa, Irene
    Pais, Richard
    Ahmad, Shama
    Angel, Erin
    Ni, Chiayi
    Kleerup, Eric C.
    Gjertson, David W.
    McNitt-Gray, Michael F.
    Goldin, Jonathan G.
    ACADEMIC RADIOLOGY, 2010, 17 (03) : 316 - 322
  • [34] Stature estimation based on femoral measurements in the modern Japanese population: a cadaveric study using multidetector computed tomography
    Chiba, Fumiko
    Makino, Yohsuke
    Torimitsu, Suguru
    Motomura, Ayumi
    Inokuchi, Go
    Ishii, Namiko
    Hoshioka, Yumi
    Abe, Hiroko
    Yamaguchi, Rutsuko
    Sakuma, Ayaka
    Nagasawa, Sayaka
    Saito, Hisako
    Yajima, Daisuke
    Fukui, Yuta
    Iwase, Hirotaro
    INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2018, 132 (05) : 1485 - 1491
  • [35] Stature estimation based on measurements of the sternal medullary cavity using multidetector computed tomography images of Japanese cadavers
    Torimitsu, Suguru
    Makino, Yohsuke
    Saitoh, Hisako
    Sakuma, Ayaka
    Ishii, Namiko
    Hayakawa, Mutsumi
    Yajima, Daisuke
    Inokuchi, Go
    Motomura, Ayumi
    Chiba, Fumiko
    Iwase, Hirotaro
    FORENSIC SCIENCE INTERNATIONAL, 2014, 242 : 299.e1 - 299.e5
  • [36] Estimation of population affinity using cranial measurements acquired in multidetector computed tomography images of Japanese and Malay individuals
    Torimitsu, Suguru
    Nakazawa, Akari
    Flavel, Ambika
    Iwase, Hirotaro
    Makino, Yohsuke
    Hisham, Salina
    Franklin, Daniel
    INTERNATIONAL JOURNAL OF LEGAL MEDICINE, 2025, 139 (02) : 863 - 873
  • [37] Stature estimation based on femoral measurements in the modern Japanese population: a cadaveric study using multidetector computed tomography
    Fumiko Chiba
    Yohsuke Makino
    Suguru Torimitsu
    Ayumi Motomura
    Go Inokuchi
    Namiko Ishii
    Yumi Hoshioka
    Hiroko Abe
    Rutsuko Yamaguchi
    Ayaka Sakuma
    Sayaka Nagasawa
    Hisako Saito
    Daisuke Yajima
    Yuta Fukui
    Hirotaro Iwase
    International Journal of Legal Medicine, 2018, 132 : 1485 - 1491
  • [38] ESTIMATION OF FELINE RENAL VOLUME USING COMPUTED TOMOGRAPHY AND ULTRASOUND
    Tyson, Reid
    Logsdon, Stacy A.
    Werre, Stephen R.
    Daniel, Gregory B.
    VETERINARY RADIOLOGY & ULTRASOUND, 2013, 54 (02) : 127 - 132
  • [39] Ion range estimation by using dual energy computed tomography
    Huenemohr, Nora
    Krauss, Bernhard
    Dinkel, Julien
    Gillmann, Clarissa
    Ackermann, Benjamin
    Jaekel, Oliver
    Greilich, Steffen
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2013, 23 (04): : 300 - 313
  • [40] REGIONAL MYOCARDIAL FLOW ESTIMATION USING COMPUTED-TOMOGRAPHY
    NASSI, M
    BRODY, WR
    MEDICAL PHYSICS, 1981, 8 (03) : 302 - 307