Safety and Efficacy of Implantation of the Bonebridge Active Transcutaneous Bone-Conduction Device Using Implant Lifts

被引:6
|
作者
Brkic, Faris F. [1 ]
Riss, Dominik [1 ]
Arnoldner, Christoph [1 ]
Liepins, Rudolfs [1 ]
Gstoettner, Wolfgang [1 ]
Baumgartner, Wolf-Dieter [1 ]
Vyskocil, Erich [1 ]
机构
[1] Med Univ Vienna, Dept Otorhinolaryngol Head & Neck Surg, Vienna, Austria
关键词
implant lifts; Bonebridge; complications; audiologic outcomes; HEARING IMPLANT; INDICATION CRITERIA; COMPLICATIONS; OUTCOMES; CHILDREN;
D O I
10.1055/s-0041-1723038
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Background Implant lifts were recently introduced to facilitate implantation of the Bonebridge and to reduce the risk of uncovering the sigmoid sinus and/or dura. Purpose The current study analyzed medical, technical, and audiological outcomes of implantation with the Bonebridge implant using lifts. Research Design This was a retrospective study on all consecutive patients implanted with a bone-conduction hearing implant at a tertiary medical referral center between March 2012 and October 2018. Outcome measures were complications, explantations, and revisions and the mean time of implant use. Audiological results were assessed as well. Outcomes were evaluated for devices implanted with BCI Lifts and compared with those implanted without lifts. Results In the study period, 13 out of a total of 54 implantations were conducted using one or two 1- to 4-mm BCI Lifts. During the follow-up period, two complications occurred and both in patients implanted without lifts (2/41; 4.9%). All patients in the lifts group were using the implant at the end of observation period. No statistically significant difference was observed in functional hearing gain or word-recognition improvement at 65dB between two groups. Conclusions The use of BCI Lifts in Bonebridge implantations was not associated with adverse events during the observation period. The clinical follow-up revealed no complications in implantations requiring lifts. Furthermore, the functional hearing gain and the word-recognition improvement did not differ from those of devices implanted without lifts. Data indicate safety and efficacy for Bonebridge implantations using lifts.
引用
收藏
页码:290 / 294
页数:5
相关论文
共 50 条
  • [41] Simultaneous auricular reconstruction and transcutaneous bone conduction device implantation in patients with microtia
    Chan, Kai-Chieh
    Wallace, Christopher Glenn
    Ho, Valerie Wai-Yee
    Wu, Che-Ming
    Chen, Hsin-Yu
    Chen, Zung-Chung
    JOURNAL OF THE FORMOSAN MEDICAL ASSOCIATION, 2019, 118 (08) : 1202 - 1210
  • [42] Safety and efficacy of transcutaneous bone conduction implant surgery for hearing improvement in microtia patients with bilateral hearing impairment
    Cheon, Jeong Hyun
    Lee, Hyung Chul
    Im, Gi Jung
    Park, Jung Youl
    Park, Chul
    ARCHIVES OF PLASTIC SURGERY-APS, 2019, 46 (06): : 525 - 534
  • [43] Transportation Noise Reduction Through Analysis and using Bone-Conduction Device
    Park, Hyung-Woo
    Bae, Myung-Jin
    2015 SAI INTELLIGENT SYSTEMS CONFERENCE (INTELLISYS), 2015, : 1049 - 1050
  • [44] Audiological Results in SSD With an Active Transcutaneous Bone Conduction Implant at a Retrosigmoidal Position
    Salcher, Rolf
    Zimmermann, Denise
    Giere, Thomas
    Lenarz, Thomas
    Maier, Hannes
    OTOLOGY & NEUROTOLOGY, 2017, 38 (05) : 642 - 647
  • [45] Early Outcomes of a New Active Transcutaneous Bone Conduction Implant in Pediatric Patients
    You, Peng
    Choi, Alexander
    Drob, Jennifer
    Hunsaker, Sabrina Marciante
    Liu, Yi-Chun Carol
    Silva, Rodrigo
    OTOLOGY & NEUROTOLOGY, 2022, 43 (02) : 212 - 218
  • [46] An Experimental Objective Method to Determine Maximum Output and Dynamic Range of an Active Bone Conduction Implant: The Bonebridge
    Mertens, Griet
    Desmet, Jolien
    Snik, Ad F. M.
    Van de Heyning, Paul
    OTOLOGY & NEUROTOLOGY, 2014, 35 (07) : 1126 - 1130
  • [47] y Image-guided surgical navigation for bone-conduction hearing device implant placement
    Shih, Michael
    Gitomer, Sarah A.
    Barton, Geran
    Liu, Yi-Chun Carol
    INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2020, 139
  • [48] Outcomes After Transcutaneous Bone-Conduction Implantation in Adults and Children: Erratum (vol 44, pg 317, 2023)
    Harris, M. K.
    Kaul, V. F.
    Bergman, M.
    OTOLOGY & NEUROTOLOGY, 2024, 45 (09) : 1093 - 1093
  • [49] A bone conduction implant using self-drilling screws Self-drilling screws as a new fixation method of an active transcutaneous bone conduction hearing implant
    Koitschev, Assen
    Neudert, Marcus
    Lenarz, Thomas
    HNO, 2023, 71 (SUPPL 1) : 61 - 66
  • [50] Transcutaneous bone-conduction hearing device: audiological and surgical aspects in a first series of patients with mixed hearing loss
    Barbara, Maurizio
    Perotti, Marco
    Gioia, Barbara
    Volpini, Luigi
    Monini, Simonetta
    ACTA OTO-LARYNGOLOGICA, 2013, 133 (10) : 1058 - 1064