An Experimental Objective Method to Determine Maximum Output and Dynamic Range of an Active Bone Conduction Implant: The Bonebridge

被引:0
|
作者
Mertens, Griet [1 ,2 ]
Desmet, Jolien [1 ,2 ]
Snik, Ad F. M. [3 ,4 ]
Van de Heyning, Paul [1 ,2 ]
机构
[1] Univ Antwerp Hosp, Univ Dept Otorhinolaryngol Head & Neck Surg, B-2650 Edegem, Belgium
[2] Univ Antwerp, Fac Med, Antwerp, Belgium
[3] Radboud Univ Nijmegen, Med Ctr, Dept Otorhinolaryngol, NL-6525 ED Nijmegen, Netherlands
[4] Donders Inst Brain Cognit & Behav, Nijmegen, Netherlands
关键词
Bone conduction device; Bone conduction implant; Maximum output; Mixed hearing loss; ANCHORED HEARING-AID; SOUND PRESSURE; BAHA SURGERY; LEVEL;
D O I
暂无
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Introduction: Recently, a new active bone conduction implant, the Bonebridge, was introduced. This transcutaneous device is proposed as an alternative to previous percutaneous systems. The current study aims to determine the maximum output (MO) of the Bonebridge by making use of Bonebridge-generated sound pressure levels in the occluded ear canal of the unaided ear. Methodology: The test setup consisted of audiometry and input-output measurements. These tests were performed on 3 Bonebridge users with conductive or mixed hearing loss (bone-conduction thresholds, <= 45 dB HL) at least 3 months after implantation surgery. All the patients were implanted and were evaluated in the Antwerp University Hospital. The MO of the device was determined by measuring input-output functions with a microphone placed in the occluded contralateral ear canal using the Aurical REM system. During testing, the sound processor was fitted in linear amplification mode and with unlimited output to determine the MO and the input dynamic range of the Bonebridge. This experimental setup intends to evaluate the device in a fitting program without compression. Results: The mean MO of the device was 55 dB HL (SD, 6 dB HL) at 0.5 kHz, 61 dB HL (SD, 18 dB HL), 71 dB HL (SD, 10 dB HL) at 2 kHz, and 60 dB HL (SD, 10 dB HL) at 4 kHz. The mean dynamic range of the Bonebridge was 41 (SD, 5) dB HL, 46 (SD, 10) dB HL, 46 (SD, 5) dB HL, and 37 (SD, 16) dB HL for 0.5, 1, 2, and 4 kHz, respectively. Conclusion: In summary, ear canal measures can effectively be used to assess input-output behavior of the Bonebridge. The present study indicates that the MO of the Bonebridge ranges from 55 to 71 dB HL, depending on frequency. Accepting a minimum dynamic range of 35 dB with the Bonebridge, fitting of the Bonebridge in a linear program is advocated in patients with a sensorineural hearing loss component of up to 30 dB HL.
引用
收藏
页码:1126 / 1130
页数:5
相关论文
共 19 条
  • [1] The Bonebridge BCI 602 Active Transcutaneous Bone Conduction Implant in Children: Objective and Subjective Benefits
    Cywka, Katarzyna B.
    Skarzynski, Henryk
    Krol, Bartlomiej
    Skarzynski, Piotr H.
    JOURNAL OF CLINICAL MEDICINE, 2021, 10 (24)
  • [2] A Novel Method to Determine the Maximum Output of Individual Patients for an Active Transcutaneous Bone Conduction Implant Using Clinical Routine Data
    Ghoncheh, Mohammad
    Busch, Susan
    Lenarz, Thomas
    Maier, Hannes
    EAR AND HEARING, 2024, 45 (01): : 219 - 226
  • [3] Objective and subjective results of the Bonebridge transcutaneous active direct-drive bone conduction hearing implant
    Alzhrani, Farid
    SAUDI MEDICAL JOURNAL, 2019, 40 (08) : 797 - 801
  • [4] Postoperative pain in patients undergoing a transcutaneous active bone conduction implant (Bonebridge)
    Lassaletta, Luis
    Calvino, Miryam
    Zernotti, Mario
    Gavilan, Javier
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2016, 273 (12) : 4103 - 4110
  • [5] Postoperative pain in patients undergoing a transcutaneous active bone conduction implant (Bonebridge)
    Luis Lassaletta
    Miryam Calvino
    Mario Zernotti
    Javier Gavilán
    European Archives of Oto-Rhino-Laryngology, 2016, 273 : 4103 - 4110
  • [6] Audiologic, surgical and subjective outcomes of active transcutaneous bone conduction implant system (Bonebridge)
    Oh, Se-Joon
    Goh, Eui-Kyung
    Choi, Sung-Won
    Lee, Seokhwan
    Lee, Hyun-Min
    Lee, Il-Woo
    Kong, Soo-Keun
    INTERNATIONAL JOURNAL OF AUDIOLOGY, 2019, 58 (12) : 956 - 963
  • [7] Comparison of artefact reduction possibilities with the new active transcutaneous bone conduction implant (Bonebridge)
    Edlinger, S.
    Tenner, E.
    Fruehwald, J.
    Sprinzl, G.
    JOURNAL OF LARYNGOLOGY AND OTOLOGY, 2023, 137 (03): : 263 - 269
  • [8] The BONEBRIDGE active transcutaneous bone conduction implant: effects of location, lifts and screws on sound transmission
    Seyed Alireza Rohani
    Mandolin Li Bartling
    Hanif M. Ladak
    Sumit K. Agrawal
    Journal of Otolaryngology - Head & Neck Surgery, 49
  • [9] The BONEBRIDGE active transcutaneous bone conduction implant: effects of location, lifts and screws on sound transmission
    Rohani, Seyed Alireza
    Bartling, Mandolin Li
    Ladak, Hanif M.
    Agrawal, Sumit K.
    JOURNAL OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 2020, 49 (01)
  • [10] Safety and Efficacy of Implantation of the Bonebridge Active Transcutaneous Bone-Conduction Device Using Implant Lifts
    Brkic, Faris F.
    Riss, Dominik
    Arnoldner, Christoph
    Liepins, Rudolfs
    Gstoettner, Wolfgang
    Baumgartner, Wolf-Dieter
    Vyskocil, Erich
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2021, 32 (05) : 290 - 294