A diamond target for megavoltage cone-beam CT

被引:18
|
作者
Sawkey, D. [1 ]
Lu, M. [1 ]
Morin, O. [1 ]
Aubin, M. [1 ]
Yom, S. S. [1 ]
Gottschalk, A. R. [1 ]
Bani-Hashemi, A. [2 ]
Faddegon, B. A. [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, Ctr Comprehens Canc, San Francisco, CA 94143 USA
[2] Siemens Healthcare Sector, Concord, CA 94520 USA
关键词
biomedical equipment; computerised tomography; diamond; dosimetry; linear accelerators; Monte Carlo methods; X-ray imaging; X-RAY-BEAMS; MONTE-CARLO; COMPUTED-TOMOGRAPHY; SIMULATION; SYSTEM; IMPLEMENTATION; LINE;
D O I
10.1118/1.3302831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A sintered diamond target was used in place of a graphite target as part of an imaging beam line (an unflattened beam from a graphite target) installed on a linear accelerator. The diamond target, with a greater density than the graphite target, permitted imaging at the lower beam energy (4 MV) required with the graphite target and the higher beam energy (6 MV) conventionally used with the tungsten/stainless steel target and stainless steel flattening filter. Images of phantoms and patients were acquired using the different beam lines and compared. The beam spectra and dose distributions were determined using Monte Carlo simulation. Results: The diamond target allowed use of the same beam energy as for treatment, simplifying commissioning and quality assurance. Images acquired with the diamond target at 4 MV were similar to those obtained with the graphite target at 4 MV. The slight reduction in low energy photons due to the higher-Z sintering material in the diamond target had minimal effect on image quality. Images acquired at 6 MV with the diamond target showed a small decrease in contrast-to-noise ratio, resulting from a decrease in the fraction of photons in the beam in the energy range to which the detector is most sensitive. Conclusions: The diamond target provides images of a similar quality to the graphite target. Diamond allows use of the higher beam energy conventionally used for treatment, provides a higher dose rate for the same beam current, and potentially simplifies installation and maintenance of the beam line.
引用
收藏
页码:1246 / 1253
页数:8
相关论文
共 50 条
  • [31] Towards optimization of megavoltage cone-beam CT for image-guided radiotherapy
    Pang, G
    O'Brien, P
    Mei, X
    Tambasco, M
    Yeboah, C
    Beachey, D
    Rowlands, J
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S623 - S623
  • [32] Dose-guided radiation therapy strategies with megavoltage cone-beam CT
    Pouliot, J.
    Aubry, J.
    Aubin, M.
    Morin, O.
    Chen, J.
    Nunna, C.
    Celi, J.
    MEDICAL PHYSICS, 2007, 34 (06) : 2379 - 2379
  • [33] The effect of megavoltage field size on intrafraction cone-beam CT image quality
    Friend, Georgia
    O'Connor, Patrick
    Charles, Paul
    PHYSICAL AND ENGINEERING SCIENCES IN MEDICINE, 2020, 43 (02) : 711 - 717
  • [34] Investigation of the optimal number of projections for head and neck megavoltage cone-beam CT
    Hristov, D
    Khamene, A
    Gangadharan, B
    Zheng, Z
    Svatos, M
    Banihashemi, A
    Mitschke, M
    Sauer, F
    MEDICAL PHYSICS, 2004, 31 (06) : 1722 - 1722
  • [35] The effect of megavoltage field size on intrafraction cone-beam CT image quality
    Georgia Friend
    Patrick O’Connor
    Paul Charles
    Physical and Engineering Sciences in Medicine, 2020, 43 : 711 - 717
  • [36] Quantitative cone beam megavoltage CT
    Spies, L
    Ebert, M
    Groh, BA
    Hesse, BM
    Bortfeld, T
    USE OF COMPUTERS IN RADIATION THERAPY, 2000, : 564 - 566
  • [37] Comparison of verification accuracy and radiation dose between megavoltage CT and kilovoltage cone-beam CT
    Wu, Vincent W. C.
    Ho, M. L.
    Yuen, L. Y.
    Li, K. L.
    Lai, N. F.
    Tsang, Y. C.
    Lee, W. Y.
    Chiu, G.
    JOURNAL OF RADIOTHERAPY IN PRACTICE, 2011, 10 (01) : 35 - 44
  • [38] Correction of megavoltage cone-beam CT images for dose calculation in the head and neck region
    Aubrya, Jean-Francois
    Pouliot, Jean
    Beaulieu, Luc
    MEDICAL PHYSICS, 2008, 35 (03) : 900 - 907
  • [39] A Prototype Fiber-Optic Scintillation Glass Array for Megavoltage Cone-Beam CT
    Gopal, A.
    Samant, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [40] Observation of interfractional variations in lung tumor position using megavoltage cone-beam CT
    Amols, HI
    Chang, J
    De Arruda, F
    Sillanpaa, J
    Yorke, E
    Hertanto, A
    Pham, H
    Rosenzweig, K
    Ling, C
    Seppi, E
    Mageras, G
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S88 - S88