A diamond target for megavoltage cone-beam CT

被引:18
|
作者
Sawkey, D. [1 ]
Lu, M. [1 ]
Morin, O. [1 ]
Aubin, M. [1 ]
Yom, S. S. [1 ]
Gottschalk, A. R. [1 ]
Bani-Hashemi, A. [2 ]
Faddegon, B. A. [1 ]
机构
[1] Univ Calif San Francisco, Dept Radiat Oncol, Ctr Comprehens Canc, San Francisco, CA 94143 USA
[2] Siemens Healthcare Sector, Concord, CA 94520 USA
关键词
biomedical equipment; computerised tomography; diamond; dosimetry; linear accelerators; Monte Carlo methods; X-ray imaging; X-RAY-BEAMS; MONTE-CARLO; COMPUTED-TOMOGRAPHY; SIMULATION; SYSTEM; IMPLEMENTATION; LINE;
D O I
10.1118/1.3302831
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Methods: A sintered diamond target was used in place of a graphite target as part of an imaging beam line (an unflattened beam from a graphite target) installed on a linear accelerator. The diamond target, with a greater density than the graphite target, permitted imaging at the lower beam energy (4 MV) required with the graphite target and the higher beam energy (6 MV) conventionally used with the tungsten/stainless steel target and stainless steel flattening filter. Images of phantoms and patients were acquired using the different beam lines and compared. The beam spectra and dose distributions were determined using Monte Carlo simulation. Results: The diamond target allowed use of the same beam energy as for treatment, simplifying commissioning and quality assurance. Images acquired with the diamond target at 4 MV were similar to those obtained with the graphite target at 4 MV. The slight reduction in low energy photons due to the higher-Z sintering material in the diamond target had minimal effect on image quality. Images acquired at 6 MV with the diamond target showed a small decrease in contrast-to-noise ratio, resulting from a decrease in the fraction of photons in the beam in the energy range to which the detector is most sensitive. Conclusions: The diamond target provides images of a similar quality to the graphite target. Diamond allows use of the higher beam energy conventionally used for treatment, provides a higher dose rate for the same beam current, and potentially simplifies installation and maintenance of the beam line.
引用
收藏
页码:1246 / 1253
页数:8
相关论文
共 50 条
  • [21] Breath-hold Target Localization with simultaneous Kilovoltage/Megavoltage Cone-Beam CT and fast Reconstruction
    Blessing, M.
    Stsepankou, D.
    Wertz, H.
    Arns, A.
    Lohr, F.
    Hesser, J.
    Wenz, F.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 201 - 204
  • [22] VOLUME OF INTEREST MEGAVOLTAGE CONE-BEAM CT WITH A LOW ATOMIC NUMBER X-RAY TARGET
    Robar, J.
    Berman, A.
    MacDonald, A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S48 - S48
  • [23] Diaphragm motion quantification in megavoltage cone-beam CT projection images
    Chen, Mingqing
    Siochi, R. Alfredo
    MEDICAL PHYSICS, 2010, 37 (05) : 2312 - 2320
  • [24] Various procedures for the reduction of cupping artifacts in MegaVoltage Cone-Beam CT
    Petit, S.
    van Elmpt, W.
    Nijsten, S.
    Dekker, A.
    Lambin, P.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S232 - S233
  • [25] Patient dose considerations for routine megavoltage cone-beam CT imaging
    Morin, Olivier
    Gillis, Amy
    Descovich, Martina
    Chen, Josephine
    Aubin, Michele
    Aubry, Jean-Francois
    Chen, Hong
    Gottschalk, Alexander R.
    Xia, Ping
    Pouliot, Jean
    MEDICAL PHYSICS, 2007, 34 (05) : 1819 - 1827
  • [26] Low-dose megavoltage cone-beam CT for radiation therapy
    Pouliot, J
    Bani-Hashemi, A
    Chen, J
    Svatos, M
    Ghelmansarai, F
    Mitschke, M
    Aubin, M
    Xia, P
    Morin, O
    Bucci, K
    Roach, M
    Hernandez, P
    Zheng, ZR
    Hristov, D
    Verhey, L
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 61 (02): : 552 - 560
  • [27] Integrating respiratory gating into a megavoltage cone-beam CT (CBCT) system
    Sillanpaa, J
    Chang, J
    Ling, C
    Seppi, E
    Mageras, G
    Amols, H
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2004, 60 (01): : S197 - S198
  • [28] Characteristics of megavoltage cone-beam digital tomosynthesis
    Descovich, M.
    Morin, O.
    Aubry, J. F.
    Aubin, M.
    Chen, J.
    Bani-Hashemi, A.
    Pouliot, J.
    MEDICAL PHYSICS, 2008, 35 (04) : 1310 - 1316
  • [29] Evaluation of imaging performance of megavoltage cone-beam CT over an extended period
    Sumida, Iori
    Yamaguchi, Hajime
    Kizaki, Hisao
    Yamada, Yuji
    Koizumi, Masahiko
    Yoshioka, Yasuo
    Ogawa, Kazuhiko
    Kakimoto, Naoya
    Murakami, Shumei
    Furukawa, Souhei
    JOURNAL OF RADIATION RESEARCH, 2014, 55 (01) : 191 - 199
  • [30] Dose calculation in presence of a metallic object using megavoltage cone-beam CT
    Aubin, M.
    Morin, O.
    Aubry, J.
    Chen, J.
    Descovich, M.
    Pouliot, J.
    MEDICAL PHYSICS, 2006, 33 (06) : 2222 - 2222