Effects of abutment screw preload in two implant connection systems: A 3D finite element study

被引:7
|
作者
Jung, Won-kyung [1 ,2 ]
Lee, Won-sup [3 ,4 ]
Kwon, Ho-Beom [1 ,2 ]
机构
[1] Seoul Natl Univ, Sch Dent, Dent Res Inst, 101 Daehak Ro, Seoul 110744, South Korea
[2] Seoul Natl Univ, Sch Dent, Dept Prosthodont, 101 Daehak Ro, Seoul 110744, South Korea
[3] Seoul Natl Univ, Grad Sch Dent, Dept Prosthodont, Seoul, South Korea
[4] Catholic Univ Korea, Dent Hosp, Dept Prosthodont, St Marys Hosp, Seoul, South Korea
来源
JOURNAL OF PROSTHETIC DENTISTRY | 2019年 / 122卷 / 05期
基金
新加坡国家研究基金会;
关键词
STRESS-DISTRIBUTION; LOADING CONDITIONS; BONE;
D O I
10.1016/j.prosdent.2019.04.025
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Statement of problem. Finite element analysis (FEA) has been used to evaluate the biomechanical behaviors of dental implants. However, in some FEA studies, the influence of the preload condition has been omitted to simplify the analysis. This might affect the results of biomechanical analysis significantly. The preload condition requires analysis. Purpose. The purpose of this FEA study was to evaluate and verify the effects of the presence of the preload condition on abutment screws under the occlusal load for external and internal hexagonal connection systems. Material and method. The finite element models consisting of bone blocks, 2 different implant systems (Osstem US and GS system; Osstem Implant Co), and crowns were created. With these components, a total of 6 models with different conditions were constructed for FEA: external hexagonal connection system only with preload (E0), external hexagonal connection system with no preload but occlusal load (EN), external hexagonal system with both preload and occlusal load (EP), internal hexagonal system only with preload (10), internal hexagonal system with no preload but occlusal load (IN), and internal hexagonal system with both preload and occlusal load (IP). An 113-degree oblique load (100 N) to the axis of the implant was applied on the occlusal surface of the crown for the models with occlusal load. A preload of 825 N was applied in the abutment screw of the models EO, EP, IO, and IP. The maximum von Mises stress, maximum principal stress, and maximum displacement of the components of the models were evaluated. Results. Both external and internal connection systems resulted in higher maximum von Mises stress and maximum principal stress values in the presence of preload in the abutment screw. The internal connection system showed higher displacement values than the external system with or without occlusal loading, and values tended to increase with the preload condition. Conclusion. The presence of a preload condition significantly affected the biomechanical behaviors of the components of 2 different connection systems. The preload condition should be included in FEA to achieve more realistic results.
引用
收藏
页码:474.e1 / 474.e8
页数:8
相关论文
共 50 条
  • [21] The effect of thread design on stress distribution in a solid screw implant: a 3D finite element analysis
    Oğuz Eraslan
    Özgür İnan
    Clinical Oral Investigations, 2010, 14 : 411 - 416
  • [22] Biomechanical analysis and comparison of 12 dental implant systems using 3D finite element study
    Liang, Rui
    Guo, Weihua
    Qiao, Xiangchen
    Wen, Hailin
    Yu, Mei
    Tang, Wei
    Liu, Lei
    Wei, Yongtao
    Tian, Weidong
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2015, 18 (12) : 1340 - 1348
  • [23] Influence of Implant/Abutment Connection on Stress Distribution to Implant-Surrounding Bone: A Finite Element Analysis
    Hanaoka, Marcia
    Alexandre Gehrke, Sergio
    Mardegan, Fabio
    Gennari, Cesar Roberto
    Taschieri, Silvio
    Del Fabbro, Massimo
    Corbella, Stefano
    JOURNAL OF PROSTHODONTICS-IMPLANT ESTHETIC AND RECONSTRUCTIVE DENTISTRY, 2014, 23 (07): : 565 - 571
  • [24] Two Implant/Abutment Joint Designs: A Comparative Finite Element Analysis
    Khraisat, Ameen
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (02) : E83 - E87
  • [25] Biomechanics of 3-implant-supported and 4-implant-supported mandibular screw-retained prostheses: A 3D finite element analysis study
    Elsayyad, Ahmed A.
    Abbas, Nadia A.
    AbdelNabi, Nouran M.
    Osman, Reham B.
    JOURNAL OF PROSTHETIC DENTISTRY, 2020, 124 (01): : 68.e1 - 68.e10
  • [26] 3D EVALUATION OF MICROGAPS AT THE IMPLANT-ABUTMENT CONNECTION BEFORE AND AFTER THERMODYNAMIC CYCLING
    Cetin, Tugce
    Asian, Y. Umut
    INTERNATIONAL JOURNAL OF PROSTHODONTICS, 2024, 37 (03) : e99 - e111
  • [27] Effects of Vertical Interarch Space and Abutment Height on Stress Distributions: A 3d Finite Element Analysis
    Naveau, Adrien
    Renault, Patrick
    Pierrisnard, Laurent
    EUROPEAN JOURNAL OF PROSTHODONTICS AND RESTORATIVE DENTISTRY, 2009, 17 (02): : 90 - 94
  • [28] A Comparative Evaluation of Stress Distribution at the Implant Abutment Interface between Platforms Switched Morse Taper and Pseudo Platform Switched Internal Hex Connection Implant - A 3D Finite Element Analysis
    Borse, Anuradha
    Tabassum, Rubina
    Shetty, Omkar
    ADVANCES IN HUMAN BIOLOGY, 2014, 4 (02) : 22 - 31
  • [29] Biomechanical Evaluation of Different Implant-Abutment Connections, Retention Systems, and Restorative Materials in the Implant-Supported Single Crowns Using 3D Finite Element Analysis
    Araujo Lemos, Cleidiel Aparecido
    Verri, Fellippo Ramos
    Noritomi, Pedro Yoshito
    de Souza Batista, Victor Eduardo
    Cruz, Ronaldo Silva
    de Luna Gomes, Jessica Marcela
    Justino de Oliveira Limirio, Joao Pedro
    Pellizzer, Eduardo Piza
    JOURNAL OF ORAL IMPLANTOLOGY, 2022, 48 (03) : 194 - 201
  • [30] The effect of 3 torque delivery systems on gold screw preload at the gold cylinder-abutment screw joint
    Tan, KB
    Nicholls, JI
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2002, 17 (02) : 175 - 183